Assessment of Binocular Vision Function With Neos™ (ROCKET) (ROCKET)
Clinical Performance of Neos™ for the Assessment of Binocular Vision Function: A Prospective Multicentre Cross-Sectional Study (ROCKET)
ROCKET evaluates the clinical performance of an investigational version of neos™, a non-invasive device that uses a virtual-reality headset and eye tracking to assess binocular vision-how the two eyes work together.
The study compares neos™ results with standard clinical examinations for three functions: sensory fusion, the ability to combine images from both eyes into a single image; stereopsis, the perception of depth from differences between the images seen by each eye; and suppression, reduced awareness of visual input from one eye during binocular viewing.
Approximately 184 patients aged 3 years and older who require a clinical binocular vision assessment will participate at ophthalmology centres in Switzerland. Each participant will undergo standard clinical testing and a neos™ examination during a single study visit. During the neos™ examination, visual stimuli are presented through the headset while the device records eye movements and pupil responses.
The main objective is to determine how accurately neos™ identifies the presence or absence of sensory fusion, stereopsis and central suppression compared with the clinical reference assessments. Additional analyses will compare fusion-related measurements, characterize stereopsis and suppression, and explore relationships between neos™ measurements and clinical findings.
調査の概要
状態
状態
条件
条件
介入・治療
介入・治療
詳細な説明
The investigation uses a prospective, multicentre, cross-sectional, paired within-participant design. Each participant receives both assessments on the same day and serves as their own comparator, reducing the influence of differences between participants. There is no random allocation. Although the investigation is not formally blinded, clinicians performing the reference examination do not have access to automated neos™ results when determining the clinical classifications. Investigational results do not influence treatment decisions.
The investigation comprises two sequential phases. An initial tuning phase includes 10 selectively recruited participants: five with preserved fusion and stereopsis without suppression, and five with absent fusion and stereopsis and suppression present. These data are used to refine algorithm parameters and thresholds. The algorithm configuration is then locked and remains unchanged throughout the validation phase. Tuning-phase participants are excluded from all analysis sets.
During validation, eligible patients are recruited consecutively when study personnel and procedures are available. Recruitment includes patients with and without suspected or confirmed binocular dysfunction, reflecting the population undergoing clinical binocularity assessment. Approximately 174 participants are expected to enter validation to obtain around 157 evaluable participants. The final enrolment depends on the observed distribution of clinical reference classifications and the proportion of evaluable examinations.
The investigational configuration uses neos™ software version 4.0.0-rocket, incorporating additional binocular visual stimuli and algorithms. The examination typically takes less than 15 minutes. It includes gaze holding and ocular alignment, stimuli for assessing suppression, fusion and stereopsis, pupillary function, saccades, smooth pursuit and visual-field assessment.
The clinical reference examination is performed by an appropriately trained healthcare professional using established methods selected according to clinical findings and participant cooperation. These may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests. Not every test is required for every participant; suitable reference information must be obtained for each primary classification. When specified test results are unavailable, unusable or omitted for clinical reasons, the protocol allows the orthoptist's clinical diagnosis to determine the reference classification.
Prespecified rules define the reference outcomes. Sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Stereopsis is present when stereoscopic disparity of 600 arcseconds or finer is detected. Central suppression is classified using a hierarchical rule: it is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is classified as absent.
Diagnostic performance is assessed separately for each primary endpoint using a two-by-two comparison of neos™ and clinical reference classifications. Sensitivity and specificity are reported with two-sided 95% exact Clopper-Pearson confidence intervals. An endpoint meets its prespecified success criterion when the lower confidence bounds for both sensitivity and specificity exceed 65%. There is no composite success criterion across the three endpoints and no multiplicity adjustment. Agreement on the affected eye is summarized descriptively among participants classified as suppression-present by both methods.
Validation recruitment ends when all prespecified reference-class requirements are met: at least 41 evaluable participants without sensory fusion, 73 with stereopsis, 73 without stereopsis and 22 with suppression. An operational review at approximately 80 evaluable participants, or earlier if indicated, examines reference-class prevalence to update enrolment projections. It does not examine agreement between neos™ and the reference assessment. No formal interim efficacy or futility analysis is planned.
Secondary analyses compare neos™ fusion break-off angles with clinical prism-bar measurements using Bland-Altman analysis. Agreement in stereopsis severity categories is assessed using weighted Cohen's kappa. Suppression-related outputs and other eye-movement, pupil and visual-field measurements are characterized descriptively, with exploratory comparisons across clinically defined groups. No formal performance goals apply to these secondary analyses.
Primary analyses include participants with conclusive results from both methods for the endpoint concerned. Missing classifications are not imputed. Participants may contribute to one endpoint analysis even if another endpoint is not evaluable. Evaluability rates and reasons for missing or inconclusive results are reported. Safety assessments include validation-phase participants who begin the neos™ examination. No routine follow-up visit is planned; ongoing adverse events are followed as clinically appropriate until resolution or stabilization.
研究の種類
研究の種類
入学 (推定)
入学
段階
段階
- 適用できない
連絡先と場所
研究連絡先
研究連絡先
- 名前:Pia Massatsch, PhD
- 電話番号:800-555-5555
- メール:clinical@machinemd.com
参加基準
適格基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Patients scheduled to undergo, or having undergone, a clinical binocularity assessment at the investigational site on the same day.
- Age ≥3 years at the time of consent.
- Ability to understand and comply with the examination, in the investigator's opinion.
- Written informed consent provided by the participant and/or legal representative.
Exclusion Criteria:
- Best-corrected visual acuity worse than 0.2 decimal (0.7 logMAR) in either eye.
- Refractive error beyond the range that can be corrected by the system, including lens inserts and the use of contact lenses.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
アーム数
武器と介入
参加者グループ / アーム参加者グループ / アーム |
介入・治療介入・治療 |
|---|---|
|
実験的:neos™ Assessment and Clinical Reference Examination
All participants undergo the investigational neos™ assessment and a standard clinical binocular vision examination during the same visit. The clinical examination serves as the reference standard for comparing classifications of sensory fusion, stereopsis and central suppression, and for assessing agreement in fusion-related measurements. |
Standard clinical binocular vision examination performed by a trained healthcare professional. Assessments may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests, selected according to clinical findings and participant cooperation. Suitable reference information is obtained for each primary outcome. The examination lasts approximately 30 minutes. |
この研究は何を測定していますか?
主要な結果の測定
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Sensitivity and Specificity of neos™ for Detecting Sensory Fusion
時間枠:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ sensory fusion classification (present/absent) compared with the clinical reference classification. Reference sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
|
Sensitivity and Specificity of neos™ for Detecting Stereopsis
時間枠:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ stereopsis classification (present/absent) compared with standard clinical stereotesting. For both methods, stereopsis is classified as present when stereoscopic disparity of 600 arcseconds or finer is detected. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
|
Sensitivity and Specificity of neos™ for Detecting Central Suppression
時間枠:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ central (foveal) suppression classification (present/absent) compared with the clinical reference classification. The reference uses a prespecified hierarchical rule: suppression is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is absent. Performance is reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
協力者と研究者
協力者
協力者
捜査官
捜査官
- 主任研究者:Muriel Dysli, MD PhD、Kantonsspital Aarau
研究記録日
主要日程の研究
研究開始 (推定)
研究開始
一次修了 (推定)
一次修了
研究の完了 (推定)
研究の完了
試験登録日
最初に提出
最初に提出
QC基準を満たした最初の提出物
QC基準を満たした最初の提出物
最初の投稿 (実際)
最初の投稿
学習記録の更新
投稿された最後の更新 (実際)
投稿された最後の更新
QC基準を満たした最後の更新が送信されました
QC基準を満たした最後の更新が送信されました
最終確認日
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
その他の研究ID番号
- CI-002
- 2026-D0075 (その他の識別子:Swissethics)
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。