Assessment of Binocular Vision Function With Neos™ (ROCKET) (ROCKET)

2026年10月30日 更新者:machineMD AG

Clinical Performance of Neos™ for the Assessment of Binocular Vision Function: A Prospective Multicentre Cross-Sectional Study (ROCKET)

ROCKET evaluates the clinical performance of an investigational version of neos™, a non-invasive device that uses a virtual-reality headset and eye tracking to assess binocular vision-how the two eyes work together.

The study compares neos™ results with standard clinical examinations for three functions: sensory fusion, the ability to combine images from both eyes into a single image; stereopsis, the perception of depth from differences between the images seen by each eye; and suppression, reduced awareness of visual input from one eye during binocular viewing.

Approximately 184 patients aged 3 years and older who require a clinical binocular vision assessment will participate at ophthalmology centres in Switzerland. Each participant will undergo standard clinical testing and a neos™ examination during a single study visit. During the neos™ examination, visual stimuli are presented through the headset while the device records eye movements and pupil responses.

The main objective is to determine how accurately neos™ identifies the presence or absence of sensory fusion, stereopsis and central suppression compared with the clinical reference assessments. Additional analyses will compare fusion-related measurements, characterize stereopsis and suppression, and explore relationships between neos™ measurements and clinical findings.

調査の概要

状態

まだ募集していません

条件

介入・治療

詳細な説明

The investigation uses a prospective, multicentre, cross-sectional, paired within-participant design. Each participant receives both assessments on the same day and serves as their own comparator, reducing the influence of differences between participants. There is no random allocation. Although the investigation is not formally blinded, clinicians performing the reference examination do not have access to automated neos™ results when determining the clinical classifications. Investigational results do not influence treatment decisions.

The investigation comprises two sequential phases. An initial tuning phase includes 10 selectively recruited participants: five with preserved fusion and stereopsis without suppression, and five with absent fusion and stereopsis and suppression present. These data are used to refine algorithm parameters and thresholds. The algorithm configuration is then locked and remains unchanged throughout the validation phase. Tuning-phase participants are excluded from all analysis sets.

During validation, eligible patients are recruited consecutively when study personnel and procedures are available. Recruitment includes patients with and without suspected or confirmed binocular dysfunction, reflecting the population undergoing clinical binocularity assessment. Approximately 174 participants are expected to enter validation to obtain around 157 evaluable participants. The final enrolment depends on the observed distribution of clinical reference classifications and the proportion of evaluable examinations.

The investigational configuration uses neos™ software version 4.0.0-rocket, incorporating additional binocular visual stimuli and algorithms. The examination typically takes less than 15 minutes. It includes gaze holding and ocular alignment, stimuli for assessing suppression, fusion and stereopsis, pupillary function, saccades, smooth pursuit and visual-field assessment.

The clinical reference examination is performed by an appropriately trained healthcare professional using established methods selected according to clinical findings and participant cooperation. These may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests. Not every test is required for every participant; suitable reference information must be obtained for each primary classification. When specified test results are unavailable, unusable or omitted for clinical reasons, the protocol allows the orthoptist's clinical diagnosis to determine the reference classification.

Prespecified rules define the reference outcomes. Sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Stereopsis is present when stereoscopic disparity of 600 arcseconds or finer is detected. Central suppression is classified using a hierarchical rule: it is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is classified as absent.

Diagnostic performance is assessed separately for each primary endpoint using a two-by-two comparison of neos™ and clinical reference classifications. Sensitivity and specificity are reported with two-sided 95% exact Clopper-Pearson confidence intervals. An endpoint meets its prespecified success criterion when the lower confidence bounds for both sensitivity and specificity exceed 65%. There is no composite success criterion across the three endpoints and no multiplicity adjustment. Agreement on the affected eye is summarized descriptively among participants classified as suppression-present by both methods.

Validation recruitment ends when all prespecified reference-class requirements are met: at least 41 evaluable participants without sensory fusion, 73 with stereopsis, 73 without stereopsis and 22 with suppression. An operational review at approximately 80 evaluable participants, or earlier if indicated, examines reference-class prevalence to update enrolment projections. It does not examine agreement between neos™ and the reference assessment. No formal interim efficacy or futility analysis is planned.

Secondary analyses compare neos™ fusion break-off angles with clinical prism-bar measurements using Bland-Altman analysis. Agreement in stereopsis severity categories is assessed using weighted Cohen's kappa. Suppression-related outputs and other eye-movement, pupil and visual-field measurements are characterized descriptively, with exploratory comparisons across clinically defined groups. No formal performance goals apply to these secondary analyses.

Primary analyses include participants with conclusive results from both methods for the endpoint concerned. Missing classifications are not imputed. Participants may contribute to one endpoint analysis even if another endpoint is not evaluable. Evaluability rates and reasons for missing or inconclusive results are reported. Safety assessments include validation-phase participants who begin the neos™ examination. No routine follow-up visit is planned; ongoing adverse events are followed as clinically appropriate until resolution or stabilization.

研究の種類

介入

入学 (推定)

184

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

3年歳以上 (子、大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • Patients scheduled to undergo, or having undergone, a clinical binocularity assessment at the investigational site on the same day.
  • Age ≥3 years at the time of consent.
  • Ability to understand and comply with the examination, in the investigator's opinion.
  • Written informed consent provided by the participant and/or legal representative.

Exclusion Criteria:

  • Best-corrected visual acuity worse than 0.2 decimal (0.7 logMAR) in either eye.
  • Refractive error beyond the range that can be corrected by the system, including lens inserts and the use of contact lenses.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:診断
  • 割り当て:なし
  • 介入モデル:単一グループの割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:neos™ Assessment and Clinical Reference Examination

All participants undergo the investigational neos™ assessment and a standard clinical binocular vision examination during the same visit. The clinical examination serves as the reference standard for comparing classifications of sensory fusion, stereopsis and central suppression, and for assessing agreement in fusion-related measurements.

Standard clinical binocular vision examination performed by a trained healthcare professional. Assessments may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests, selected according to clinical findings and participant cooperation. Suitable reference information is obtained for each primary outcome. The examination lasts approximately 30 minutes.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Sensitivity and Specificity of neos™ for Detecting Sensory Fusion
時間枠:During the single study visit (approximately 60 minutes).

Sensitivity and specificity of the neos™ sensory fusion classification (present/absent) compared with the clinical reference classification. Reference sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals.

During the single study visit (approximately 60 minutes).
Sensitivity and Specificity of neos™ for Detecting Stereopsis
時間枠:During the single study visit (approximately 60 minutes).

Sensitivity and specificity of the neos™ stereopsis classification (present/absent) compared with standard clinical stereotesting. For both methods, stereopsis is classified as present when stereoscopic disparity of 600 arcseconds or finer is detected. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals.

During the single study visit (approximately 60 minutes).
Sensitivity and Specificity of neos™ for Detecting Central Suppression
時間枠:During the single study visit (approximately 60 minutes).

Sensitivity and specificity of the neos™ central (foveal) suppression classification (present/absent) compared with the clinical reference classification. The reference uses a prespecified hierarchical rule: suppression is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is absent. Performance is reported with two-sided 95% exact Clopper-Pearson confidence intervals.

During the single study visit (approximately 60 minutes).

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

協力者

捜査官

  • 主任研究者:Muriel Dysli, MD PhD、Kantonsspital Aarau

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2027年1月1日

一次修了 (推定)

2027年6月1日

研究の完了 (推定)

2027年6月1日

試験登録日

最初に提出

2026年10月30日

QC基準を満たした最初の提出物

2026年10月30日

最初の投稿 (実際)

2026年11月5日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月5日

QC基準を満たした最後の更新が送信されました

2026年10月30日

最終確認日

2026年10月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • CI-002
  • 2026-D0075 (その他の識別子:Swissethics)

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

米国で製造され、米国から輸出された製品。

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。