Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)
2015年5月5日 更新者:Medtronic Diabetes
A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients
The purpose of this study is to assess safety and device performance of the Medtronic Hospital Glucose Management System (HGMS) for up to 72-hours.
調査の概要
状態
完了
条件
詳細な説明
Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:
Device Performance:
- Functionality of HGMS
- Alerts/Alarms Specificity and Sensitivity
- System Workflow
Safety:
- Descriptive statistics will be used to characterize safety events
- Moderate and severe anticipated device and procedure related adverse events
- All serious adverse events and unanticipated adverse device effects related events
研究の種類
観察的
入学 (実際)
19
連絡先と場所
このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。
研究場所
-
-
Missouri
-
Kansas City、Missouri、アメリカ、64111
- St Luke's Hospital Mid America Heart Institute
-
-
参加基準
研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。
適格基準
就学可能な年齢
18年歳以上 (大人、高齢者)
健康ボランティアの受け入れ
いいえ
受講資格のある性別
全て
サンプリング方法
非確率サンプル
調査対象母集団
Adult patients admitted to the Intensive Care Unit (ICU) and on intravenous (IV) insulin therapy for the treatment of transient or diabetes-related hyperglycemia
説明
Inclusion Criteria:
- Subject is at least 18 years old
- Subject is admitted to the ICU
- Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours
a. Including patients with no previous diagnosis of Diabetes Mellitus
- Subject has anticipated life expectancy greater than 96 hours
- Subject has recent platelet count greater than 30,000 per micro-liter
Exclusion Criteria:
- Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
- Subject is currently participating in another investigational drug or device study
- Subject is pregnant, as determined by hospital admission
- Subject is receiving treatment that includes Hydroxyurea.
研究計画
このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。
研究はどのように設計されていますか?
デザインの詳細
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Device Performance: Accuracy of HGMS
時間枠:up to 72 hours
|
Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values]. The portable i-STAT handheld makes patient-side testing easy:
|
up to 72 hours
|
その他の成果指標
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
時間枠:up to 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels.
The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
時間枠:72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels.
The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
時間枠:Up to 72 hours
|
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL.
The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
時間枠:Up to 72 hours
|
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL .
The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
協力者と研究者
ここでは、この調査に関係する人々や組織を見つけることができます。
スポンサー
捜査官
- 主任研究者:Mikhail Kosiborod, M.D、St Luke's Hospital - Mid America Heart Institute
出版物と役立つリンク
研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。
研究記録日
これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。
主要日程の研究
研究開始
2012年3月1日
一次修了 (実際)
2012年8月1日
研究の完了 (実際)
2012年9月1日
試験登録日
最初に提出
2012年12月21日
QC基準を満たした最初の提出物
2013年1月4日
最初の投稿 (見積もり)
2013年1月9日
学習記録の更新
投稿された最後の更新 (見積もり)
2015年5月7日
QC基準を満たした最後の更新が送信されました
2015年5月5日
最終確認日
2015年5月1日
詳しくは