Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)
A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients
Tutkimuksen yleiskatsaus
Tila
Ehdot
Yksityiskohtainen kuvaus
Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:
Device Performance:
- Functionality of HGMS
- Alerts/Alarms Specificity and Sensitivity
- System Workflow
Safety:
- Descriptive statistics will be used to characterize safety events
- Moderate and severe anticipated device and procedure related adverse events
- All serious adverse events and unanticipated adverse device effects related events
Opintotyyppi
Ilmoittautuminen (Todellinen)
Yhteystiedot ja paikat
Opiskelupaikat
-
-
Missouri
-
Kansas City, Missouri, Yhdysvallat, 64111
- St Luke's Hospital Mid America Heart Institute
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Näytteenottomenetelmä
Tutkimusväestö
Kuvaus
Inclusion Criteria:
- Subject is at least 18 years old
- Subject is admitted to the ICU
- Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours
a. Including patients with no previous diagnosis of Diabetes Mellitus
- Subject has anticipated life expectancy greater than 96 hours
- Subject has recent platelet count greater than 30,000 per micro-liter
Exclusion Criteria:
- Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
- Subject is currently participating in another investigational drug or device study
- Subject is pregnant, as determined by hospital admission
- Subject is receiving treatment that includes Hydroxyurea.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Device Performance: Accuracy of HGMS
Aikaikkuna: up to 72 hours
|
Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values]. The portable i-STAT handheld makes patient-side testing easy:
|
up to 72 hours
|
Muut tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
Aikaikkuna: up to 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels.
The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
Aikaikkuna: 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels.
The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
Aikaikkuna: Up to 72 hours
|
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL.
The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
Aikaikkuna: Up to 72 hours
|
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL .
The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
Yhteistyökumppanit ja tutkijat
Sponsori
Tutkijat
- Päätutkija: Mikhail Kosiborod, M.D, St Luke's Hospital - Mid America Heart Institute
Julkaisuja ja hyödyllisiä linkkejä
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus
Ensisijainen valmistuminen (Todellinen)
Opintojen valmistuminen (Todellinen)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Arvio)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Arvio)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- CEP244
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .