Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)
A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients
Visão geral do estudo
Status
Condições
Descrição detalhada
Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:
Device Performance:
- Functionality of HGMS
- Alerts/Alarms Specificity and Sensitivity
- System Workflow
Safety:
- Descriptive statistics will be used to characterize safety events
- Moderate and severe anticipated device and procedure related adverse events
- All serious adverse events and unanticipated adverse device effects related events
Tipo de estudo
Inscrição (Real)
Contactos e Locais
Locais de estudo
-
-
Missouri
-
Kansas City, Missouri, Estados Unidos, 64111
- St Luke's Hospital Mid America Heart Institute
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Método de amostragem
População do estudo
Descrição
Inclusion Criteria:
- Subject is at least 18 years old
- Subject is admitted to the ICU
- Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours
a. Including patients with no previous diagnosis of Diabetes Mellitus
- Subject has anticipated life expectancy greater than 96 hours
- Subject has recent platelet count greater than 30,000 per micro-liter
Exclusion Criteria:
- Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
- Subject is currently participating in another investigational drug or device study
- Subject is pregnant, as determined by hospital admission
- Subject is receiving treatment that includes Hydroxyurea.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Device Performance: Accuracy of HGMS
Prazo: up to 72 hours
|
Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values]. The portable i-STAT handheld makes patient-side testing easy:
|
up to 72 hours
|
Outras medidas de resultado
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
Prazo: up to 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels.
The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
Prazo: 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels.
The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
Prazo: Up to 72 hours
|
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL.
The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
Prazo: Up to 72 hours
|
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL .
The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Mikhail Kosiborod, M.D, St Luke's Hospital - Mid America Heart Institute
Publicações e links úteis
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo
Conclusão Primária (Real)
Conclusão do estudo (Real)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Estimativa)
Atualizações de registro de estudo
Última Atualização Postada (Estimativa)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- CEP244
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .