Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)

5. května 2015 aktualizováno: Medtronic Diabetes

A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients

The purpose of this study is to assess safety and device performance of the Medtronic Hospital Glucose Management System (HGMS) for up to 72-hours.

Přehled studie

Postavení

Dokončeno

Podmínky

Detailní popis

Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:

  1. Device Performance:

    • Functionality of HGMS
    • Alerts/Alarms Specificity and Sensitivity
    • System Workflow
  2. Safety:

    • Descriptive statistics will be used to characterize safety events
    • Moderate and severe anticipated device and procedure related adverse events
    • All serious adverse events and unanticipated adverse device effects related events

Typ studie

Pozorovací

Zápis (Aktuální)

19

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní místa

    • Missouri
      • Kansas City, Missouri, Spojené státy, 64111
        • St Luke's Hospital Mid America Heart Institute

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

18 let a starší (Dospělý, Starší dospělý)

Přijímá zdravé dobrovolníky

Ne

Pohlaví způsobilá ke studiu

Všechno

Metoda odběru vzorků

Vzorek nepravděpodobnosti

Studijní populace

Adult patients admitted to the Intensive Care Unit (ICU) and on intravenous (IV) insulin therapy for the treatment of transient or diabetes-related hyperglycemia

Popis

Inclusion Criteria:

  1. Subject is at least 18 years old
  2. Subject is admitted to the ICU
  3. Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
  4. Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours

    a. Including patients with no previous diagnosis of Diabetes Mellitus

  5. Subject has anticipated life expectancy greater than 96 hours
  6. Subject has recent platelet count greater than 30,000 per micro-liter

Exclusion Criteria:

  1. Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
  2. Subject is currently participating in another investigational drug or device study
  3. Subject is pregnant, as determined by hospital admission
  4. Subject is receiving treatment that includes Hydroxyurea.

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Device Performance: Accuracy of HGMS
Časové okno: up to 72 hours

Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values].

The portable i-STAT handheld makes patient-side testing easy:

  • Requires no special sample preparation or user calibration; maintenance is minimal
  • Weighs 18 ounces, making it portable
  • Patient-side testing is as easy as entering the operator and patient information into the handheld, inserting one of the several filled test cartridges, and then viewing test results:
  • The system prompts users step by step through the testing process
  • Operator and patient information can be entered via barcode scanner
  • Operator lockout prevents unauthorized users from performing or viewing test results
  • Test results are uploaded automatically when the i-STAT handheld is placed in a downloader
up to 72 hours

Další výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
Časové okno: up to 72 hours
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels. The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
up to 72 hours
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
Časové okno: 72 hours
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels. The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
72 hours
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
Časové okno: Up to 72 hours
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL. The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants. NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
Up to 72 hours
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
Časové okno: Up to 72 hours
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL . The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants. NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
Up to 72 hours

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Vyšetřovatelé

  • Vrchní vyšetřovatel: Mikhail Kosiborod, M.D, St Luke's Hospital - Mid America Heart Institute

Publikace a užitečné odkazy

Osoba odpovědná za zadávání informací o studiu tyto publikace poskytuje dobrovolně. Mohou se týkat čehokoli, co souvisí se studiem.

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia

1. března 2012

Primární dokončení (Aktuální)

1. srpna 2012

Dokončení studie (Aktuální)

1. září 2012

Termíny zápisu do studia

První předloženo

21. prosince 2012

První předloženo, které splnilo kritéria kontroly kvality

4. ledna 2013

První zveřejněno (Odhad)

9. ledna 2013

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Odhad)

7. května 2015

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

5. května 2015

Naposledy ověřeno

1. května 2015

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • CEP244

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .

Předplatit