Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)
A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients
A tanulmány áttekintése
Állapot
Körülmények
Részletes leírás
Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:
Device Performance:
- Functionality of HGMS
- Alerts/Alarms Specificity and Sensitivity
- System Workflow
Safety:
- Descriptive statistics will be used to characterize safety events
- Moderate and severe anticipated device and procedure related adverse events
- All serious adverse events and unanticipated adverse device effects related events
Tanulmány típusa
Beiratkozás (Tényleges)
Kapcsolatok és helyek
Tanulmányi helyek
-
-
Missouri
-
Kansas City, Missouri, Egyesült Államok, 64111
- St Luke's Hospital Mid America Heart Institute
-
-
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Mintavételi módszer
Tanulmányi populáció
Leírás
Inclusion Criteria:
- Subject is at least 18 years old
- Subject is admitted to the ICU
- Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours
a. Including patients with no previous diagnosis of Diabetes Mellitus
- Subject has anticipated life expectancy greater than 96 hours
- Subject has recent platelet count greater than 30,000 per micro-liter
Exclusion Criteria:
- Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
- Subject is currently participating in another investigational drug or device study
- Subject is pregnant, as determined by hospital admission
- Subject is receiving treatment that includes Hydroxyurea.
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Device Performance: Accuracy of HGMS
Időkeret: up to 72 hours
|
Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values]. The portable i-STAT handheld makes patient-side testing easy:
|
up to 72 hours
|
Egyéb eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
Időkeret: up to 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels.
The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
Időkeret: 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels.
The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
Időkeret: Up to 72 hours
|
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL.
The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
Időkeret: Up to 72 hours
|
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL .
The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
Együttműködők és nyomozók
Szponzor
Nyomozók
- Kutatásvezető: Mikhail Kosiborod, M.D, St Luke's Hospital - Mid America Heart Institute
Publikációk és hasznos linkek
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete
Elsődleges befejezés (Tényleges)
A tanulmány befejezése (Tényleges)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Becslés)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Becslés)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
További vonatkozó MeSH feltételek
Egyéb vizsgálati azonosító számok
- CEP244
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .