Feasibility Study to Evaluate Safety and Device Performance of the Hospital Glucose Management System (HGMS) (MAHI)
A Multi-Phased, Multi-Center Study to Evaluate Safety and Device Performance of the Medtronic Hospital Glucose Management System (HGMS) in Critically Ill Adult Patients
Panoramica dello studio
Stato
Condizioni
Descrizione dettagliata
Assess safety and device performance of the Medtronic HGMS in the critically-ill setting and analyze the following:
Device Performance:
- Functionality of HGMS
- Alerts/Alarms Specificity and Sensitivity
- System Workflow
Safety:
- Descriptive statistics will be used to characterize safety events
- Moderate and severe anticipated device and procedure related adverse events
- All serious adverse events and unanticipated adverse device effects related events
Tipo di studio
Iscrizione (Effettivo)
Contatti e Sedi
Luoghi di studio
-
-
Missouri
-
Kansas City, Missouri, Stati Uniti, 64111
- St Luke's Hospital Mid America Heart Institute
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
- Subject is at least 18 years old
- Subject is admitted to the ICU
- Subject has a treatment regimen that includes a glucose target range of ≥140 mg/dl
Subject has a treatment regimen that includes an intended continuous intravenous insulin for at least 24 hours
a. Including patients with no previous diagnosis of Diabetes Mellitus
- Subject has anticipated life expectancy greater than 96 hours
- Subject has recent platelet count greater than 30,000 per micro-liter
Exclusion Criteria:
- Subject currently has a suspected or diagnosed medical condition that, in the opinion of the Investigator, warrants exclusion from the study or prevents the subject from completing the study
- Subject is currently participating in another investigational drug or device study
- Subject is pregnant, as determined by hospital admission
- Subject is receiving treatment that includes Hydroxyurea.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Device Performance: Accuracy of HGMS
Lasso di tempo: up to 72 hours
|
Mean Absolute Relative Difference (MARD), calculated as the absolute difference of [(sensor glucose values - iSTAT glucose values) / iSTAT glucose values]. The portable i-STAT handheld makes patient-side testing easy:
|
up to 72 hours
|
Altre misure di risultato
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo Events Correctly Detected
Lasso di tempo: up to 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes below 70 mg/dL setting levels.
The % of hypo events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper Events Correctly Detected
Lasso di tempo: 72 hours
|
The device alarmed within 30 minutes before or after the reference blood glucose value (i-STAT) goes above 250 mg/dL setting levels.
The % of hyper events correctly detected was calculated as: total number of correct events divided by total number of events from all 19 participants.
|
72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hypo False Alert
Lasso di tempo: Up to 72 hours
|
Definition of % of hypo false alert: within 30 minutes before or after the sensor alarmed at 70 mg/dL, there is no reference blood glucose value (i-STAT) that goes below 70 mg/dL.
The % of hypo false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hypo False Alert and % Hypo Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
|
Functionality of HGMS: Alerts and Alarms - % of Hyper False Alert
Lasso di tempo: Up to 72 hours
|
Definition of % of Hyper False Alert: within 30 minutes before or after the sensor alarmed at 250 mg/dL, there is no reference blood glucose value (i-STAT) that goes above 250mg/dL .
The % of hyper false alert was calculated as: total number of false events divided by total number of events from all 19 participants.
NOTE: % of Hyper False Alert and % Hyper Event Correctly Detected do not necessarily add up to 100% because the denomintors are not the same.
|
Up to 72 hours
|
Collaboratori e investigatori
Sponsor
Investigatori
- Investigatore principale: Mikhail Kosiborod, M.D, St Luke's Hospital - Mid America Heart Institute
Pubblicazioni e link utili
Studiare le date dei record
Studia le date principali
Inizio studio
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Stima)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Stima)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- CEP244
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .