Development of a Clinical Decision Support System for Cataract Surgery Planning Using Swept Source Optical Biometry (CAT-OCT)

2026년 11월 9일 업데이트: Clínica de Oftalmología de Cali S.A

This single-center, observational, analytical, ambispective study aims to develop and internally validate a clinical decision support system for cataract surgery planning using swept-source optical coherence tomography (SS-OCT) integrated into the ARGOS optical biometer.

The study will evaluate the ability of preoperative crystalline lens tomographic patterns (Types I-III) to predict surgical complexity and the surgical approach required during cataract extraction. Additionally, it will assess whether SS-OCT-derived information provides incremental predictive value beyond conventional clinical assessment using the Lens Opacities Classification System III (LOCS III).

Clinical, biometric, and surgical data will be analyzed to determine the predictive performance, reproducibility, and technical feasibility of the proposed classification system. The ultimate goal is to establish an objective, noninvasive, and reproducible tool to support personalized surgical planning and preoperative risk stratification in cataract surgery.

연구 개요

상태

완전한

정황

개입 / 치료

상세 설명

Background and Rationale

Cataract remains a leading cause of visual impairment and preventable blindness worldwide. Although advances in surgical techniques and technology have improved cataract surgery outcomes, predicting surgical complexity and selecting the most appropriate surgical approach remain important clinical challenges. Currently, preoperative assessment relies primarily on clinical examination and subjective grading systems, such as the Lens Opacities Classification System III (LOCS III), which may exhibit interobserver variability.

Swept-source optical coherence tomography (SS-OCT), integrated into optical biometers such as ARGOS (Alcon Laboratories, Inc.), provides noninvasive, high-resolution images of the crystalline lens and biometric measurements. These images may reveal morphological characteristics associated with lens opacity, nuclear density, and surgical complexity.

This study proposes the development and internal validation of a clinical decision support system based on the identification of specific crystalline lens tomographic patterns to improve preoperative risk stratification and support cataract surgery planning.

Study Design and Population

This is a single-center, observational, analytical, ambispective study conducted at Clínica de Oftalmología de Cali, Colombia. The study will include patients older than 40 years undergoing crystalline lens extraction who have undergone preoperative optical biometry using the ARGOS device, including SS-OCT imaging of the crystalline lens.

The ambispective design will incorporate retrospectively identified surgical cases and prospectively included patients. Surgical management will follow routine clinical practice, without experimental treatment assignment or modification of the surgical technique by the investigators.

Patients undergoing combined cataract surgery with glaucoma, corneal, or retinal procedures, those with a history of significant ocular trauma affecting anterior segment anatomy, and patients with ocular conditions preventing reliable tomographic image acquisition will be excluded.

A reference group of patients with clear crystalline lenses who underwent ARGOS biometry for indications other than cataract will be used exclusively to characterize the Type I tomographic pattern. These individuals will not be included in the statistical analysis.

Preoperative Assessment and Tomographic Classification

Preoperative evaluation will include demographic characteristics, visual acuity, LOCS III cataract grading, relevant ocular comorbidities, and biometric parameters obtained using ARGOS.

Crystalline lens SS-OCT images will be classified according to three predefined tomographic patterns (Types I, II, and III), based on lens reflectivity, intralenticular signal distribution, and the morphological characteristics of the anterior and posterior lens peaks.

Images will be independently assessed by trained observers blinded to the surgical technique and intraoperative outcomes. Disagreements will be resolved through consensus or third-observer adjudication. Interobserver and intraobserver reproducibility will be evaluated.

Study Outcomes

The primary outcome will be the need for a complex surgical approach during cataract extraction, defined using prespecified objective criteria derived from surgical records and, when available, surgical recordings.

Secondary outcomes will include the specific surgical technique performed, conversion from the initially planned approach, phacoemulsification energy parameters, intraoperative complications, and technical feasibility of SS-OCT image acquisition and interpretation.

Statistical Analysis and Internal Validation

The predictive performance of the proposed tomographic classification will be assessed using sensitivity, specificity, positive and negative predictive values, likelihood ratios, and overall classification accuracy, with 95% confidence intervals.

Multivariable logistic regression will be used to evaluate the association between SS-OCT patterns and surgical complexity. A conventional clinical prediction model based on LOCS III and prespecified clinical variables will be compared with an expanded model incorporating the SS-OCT tomographic classification.

Model discrimination will be evaluated using the area under the receiver operating characteristic curve (AUC-ROC), while calibration will be assessed by comparing predicted and observed outcomes. Internal validation will be performed using bootstrap resampling.

Expected Clinical Relevance

The study seeks to determine whether SS-OCT-derived crystalline lens patterns can provide clinically useful information beyond conventional cataract grading. If validated, this classification could support more objective preoperative assessment, improve identification of surgically complex cases, and facilitate individualized cataract surgery planning using information already available from routine optical biometry.

The proposed system is intended to complement, rather than replace, clinical judgment. Further external validation will be required before its implementation in routine clinical practice.

연구 유형

관찰

등록 (실제)

499

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 장소

    • Valle del Cauca Department
      • Cali, Valle del Cauca Department, 콜롬비아, 760036
        • Clínica de Oftalmología de Cali

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

40년 이상 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

예

연구 대상 성별

모두

샘플링 방법

비확률 샘플

연구 인구

For the establishment of the reference group with clear crystalline lenses, patients who underwent optical biometry using the ARGOS device for indications other than cataract, such as myopia monitoring, will be included to characterize the Type I pattern. These patients were not included in the statistical analysis of the study.

The study population will consist of patients older than 40 years with an indication for crystalline lens extraction and preoperative biometry performed using ARGOS, including acquisition of SS-OCT images of the crystalline lens. Patients in whom both the preoperative tomographic pattern and the surgical outcome can be adequately determined will be eligible.

설명

Inclusion Criteria:

  • Patients older than 40 years undergoing crystalline lens extraction surgery.
  • Preoperative optical biometry performed using the ARGOS biometer, including swept-source optical coherence tomography (SS-OCT) imaging of the crystalline lens.
  • Availability of sufficient preoperative tomographic information and surgical records to determine both the lenticular tomographic pattern and the surgical outcome.

Exclusion Criteria:

  • Patients undergoing combined cataract surgery with glaucoma, corneal, or retinal procedures.
  • Patients with a history of ocular trauma capable of significantly altering anterior segment anatomy.
  • Patients with ocular conditions that prevent the acquisition of reliable tomographic images using swept-source optical coherence tomography (SS-OCT).

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 관찰 모델: COHORT
  • 시간 관점: PROSPECTIVE

코호트 및 개입

그룹/코호트
개입 / 치료
Cataract Surgery Patients

Patients older than 40 years undergoing cataract surgery with preoperative swept-source optical coherence tomography (SS-OCT) imaging obtained using the ARGOS biometer. Lenticular tomographic patterns will be classified and evaluated for their ability to predict surgical complexity and the surgical approach required.

Preoperative assessment of crystalline lens tomographic patterns using swept-source optical coherence tomography (SS-OCT) integrated into the ARGOS optical biometer. Images will be classified according to predefined lenticular patterns (Types I-III) to evaluate their predictive value for surgical complexity and surgical approach during cataract extraction.

다른 이름들:
  • ARGOS Biometer
  • ARGOS Swept-Source OCT
  • SS-OCT Optical Biometry

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Predictive Performance of SS-OCT Lenticular Patterns for Surgical Complexity in Cataract Surgery
기간: 6 months

Assessment of the ability of preoperative crystalline lens tomographic patterns (Types I-III), obtained using swept-source optical coherence tomography (SS-OCT) with the ARGOS biometer, to predict the need for a complex surgical approach during cataract extraction. Surgical complexity will be classified as conventional or complex according to predefined objective criteria derived from operative records. Predictive performance will be evaluated using sensitivity, specificity, positive and negative predictive values, likelihood ratios, and overall classification accuracy, with 95% confidence intervals.

6 months

2차 결과 측정

결과 측정
측정값 설명
기간
Incremental Predictive Value of SS-OCT Compared With LOCS III
기간: 6 months

Evaluation of the additional predictive value of SS-OCT lenticular patterns beyond conventional clinical assessment using LOCS III. Predictive models will be compared in terms of discrimination, measured by the area under the receiver operating characteristic curve (AUC-ROC), and calibration.

6 months
Distribution of Surgical Techniques Across SS-OCT Lenticular Pattern Categories
기간: 6 months

To determine the percentage of eyes undergoing each surgical technique within the preoperative lenticular pattern categories (I, II, and III), as classified using swept-source optical coherence tomography (SS-OCT) images obtained with the ARGOS optical biometer.

Surgical techniques will be identified from operative records and categorized as phacoemulsification (divide and conquer, stop and chop, vertical chop, or horizontal chop), extracapsular cataract extraction, or manual small-incision cataract surgery (MSICS).

The frequency distribution of surgical techniques will be compared across SS-OCT lenticular pattern categories using appropriate categorical statistical methods.

6 months
Cumulative Dissipated Energy During Phacoemulsification Across SS-OCT Lenticular Pattern Categories
기간: 6 months

To quantify cumulative dissipated energy (CDE) during cataract surgery performed using phacoemulsification and describe its distribution across preoperative SS-OCT lenticular pattern categories (I, II, and III).

CDE will be obtained from the phacoemulsification system records and expressed in the units reported by the device.

Differences in CDE across lenticular pattern categories will be evaluated using statistical methods appropriate to the observed distribution of the data.

6 months
Interobserver and Intraobserver Reproducibility of SS-OCT Lenticular Pattern Classification
기간: 6 months

Assessment of interobserver and intraobserver agreement in the classification of SS-OCT lenticular patterns (Types I-III), using percentage agreement and Cohen's kappa coefficient with 95% confidence intervals.

6 months
Incidence of Intraoperative Complications
기간: 6 months

Assessment of the occurrence of intraoperative complications during cataract surgery and their association with preoperative SS-OCT lenticular patterns. Complication frequencies will be reported according to tomographic pattern classification.

6 months

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

스폰서

간행물 및 유용한 링크

연구에 대한 정보 입력을 담당하는 사람이 자발적으로 이러한 간행물을 제공합니다. 이것은 연구와 관련된 모든 것에 관한 것일 수 있습니다.

일반 간행물

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (실제)

2026년 5월 1일

기본 완료 (실제)

2026년 10월 19일

연구 완료 (실제)

2026년 10월 19일

연구 등록 날짜

최초 제출

2026년 10월 20일

QC 기준을 충족하는 최초 제출

2026년 11월 9일

처음 게시됨 (실제)

2026년 11월 16일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 11월 16일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 11월 9일

마지막으로 확인됨

2026년 11월 1일

추가 정보

이 연구와 관련된 용어

추가 관련 MeSH 약관

기타 연구 ID 번호

  • 2020001-01
  • 2020001-01 (기타 식별자: Self-financed)

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

예

IPD 공유 기간

In 6 months

IPD 공유 액세스 기준

Restricted access by request

IPD 공유 지원 정보 유형

  • 연구_프로토콜
  • 수액
  • CSR

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

예

미국에서 제조되어 미국에서 수출되는 제품

예