Deze pagina is automatisch vertaald en de nauwkeurigheid van de vertaling kan niet worden gegarandeerd. Raadpleeg de Engelse versie voor een brontekst.

A Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations (PropR)

27 februari 2012 bijgewerkt door: Saint Francis Care

Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations

The PropR study will evaluate sensing during ventricular fibrillation (VF) in both bipolar and extended bipolar configurations, in order to evaluate if both can be used interchangeably in caring for patients. In addition, follow up evaluation of R wave amplitude over time would allow us to determine whether one configuration is more likely to be associated with change. This understanding would be important in selecting the proper configuration at the time of implant.

Studie Overzicht

Gedetailleerde beschrijving

Hypothesis:

Sensing in bipolar and extended bipolar configurations are the same. However, sensing in bipolar over time is more likely to decrease when compared to extended bipolar.

Objectives:

  1. To compare bipolar sensing and extended bipolar sensing in VF.
  2. Compare sensing in bipolar and extended bipolar over time.

Methods:

  1. This study will be a randomized comparison of bipolar and extended bipolar sensing. As per usual standard of care, at the end of ICD implantation two sensing tests in VF will be performed. Patients will be randomized to sensing in VF through the device. Each patient will undergo testing in one randomized configuration followed by a second test in the other configuration. After testing at the time of implant, each patient will be seen at one month in follow up. Sensing of each configuration will be reevaluated at that time in sinus rhythm.
  2. Each patient will serve as their own control, as data for sensing in both configurations will be obtained in VF at the time of implant. During testing both signals types will be recorded simultaneously through the ICD. The primary endpoint will be duration of sensing (time to detection) in VF at the time of implant.
  3. The secondary endpoint will be sensing in sinus rhythm at one month.
  4. The patient population will include approximately 30-50 volunteers age >18, male and female, both ischemic and nonischemic cardiomyopathy, with primary and secondary implant indications.
  5. Patients will be recruited when being evaluated for defibrillator implant.
  6. Inclusion criteria: age >18 with ability to give informed consent, who are expected to live more than one year, with indication for defibrillator implant and who are not pacemaker dependent. Patients will have either ischemic or nonischemic cardiomyopathy.
  7. Exclusion criteria: age <18, inability to give informed consent, or life expectancy less than one year, pacemaker dependence, or use of preexisting lead for sensing, any condition which would preclude ICD testing at the end of the implant. Patients who are undergoing upgrade to an ICD from a pacemaker will also be excluded. In addition, patients with inherited arrhythmias or ion channel related arrhythmias will be excluded from the study.
  8. Withdrawal criteria: inability to implant device, implant lead at the RV apex, or make follow up appointment (living out of state), patient death at time of implant, elevated DFT (<10 J margin between DFT and maximum output of the device) requiring additional procedures, subcutaneous array etc.
  9. Duration of patient participation would be one month.
  10. Duration of study would be one year.

Procedures:

  1. Procedures to be followed would include screening of all eligible patients undergoing implant of a Medtronic defibrillator for either primary or secondary prevention. If patients meet inclusion criteria they would then undergo informed consent and if willing be enrolled in the study. Defibrillator implant and testing would be performed as per usual practice of the implanting physician. The RV lead will be placed at the RV apex as is usual for regular implantation of devices. However, if a lead is placed elsewhere, the patient will need to be withdrawn from the study.
  2. When VF induction is performed, sensing configuration would be tested in a randomized fashion, unblinded to the operator. The untested configuration would then be tested during second VF induction. Both tests would be performed at a sensitivity setting of 1.2mV, with a number of intervals to detection (NID) of 24/32. Final programming polarity will be left at the initially tested polarity (first DFT test), unless the operator sees a need to program differently for patient safety reasons.
  3. Medtronic defibrillators with both sensing configurations available will be used. Data collection will be performed by Dr. Tolat, Dr. Berns, Dr. Lippman, Dr. Dell'Orfano, and the study coordinator involved with the study. A data collection form and spreadsheet will be kept for use when collecting data for analysis. Follow up will be performed at one month post implant in the office.

Risk Analysis:

  1. Anticipated risks are no more than that anticipated with usual ICD implant. To date, usual implant includes induction of VF with analysis of sensing in VF and evaluation of defibrillation threshold by delivery of shock.
  2. There is no expected increased exposure to risk by performing this study.
  3. Potential benefits would include gathering additional information on sensing in VF in both configurations instead of performing the two sensing tests in one configuration. This may directly benefit the patient by allowing for reprogramming defibrillator without bringing the patient back to the hospital for further testing.

Data Analysis:

  1. Data collection will include patient information kept in a database with name kept confidential. Patients will only be identified by number in keeping with privacy laws and practice. In addition to usual patient characteristics, data variables to be collected in both configurations will include: R wave amplitude in sinus rhythm, time to detection in VF, total episode duration (time to Vfib detection and charge delivered), number of "drop outs" (undersensing) in VF, and follow up R wave amplitude in one month.
  2. Statistical Methods will compare bipolar sensing configuration to the extended bipolar configuration at baseline, sensing in VF (as measured by "drop out", time to detection), and follow up amplitude of the R wave at one month.
  3. The current expected sample size with this pilot study is 30 patients. This may require reevaluation depending on results at the time of interim analysis.

Studietype

Observationeel

Inschrijving (Werkelijk)

50

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

    • Connecticut
      • Hartford, Connecticut, Verenigde Staten, 06105
        • St Francis Hospital and Medical Center

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Bemonsteringsmethode

Niet-waarschijnlijkheidssteekproef

Studie Bevolking

The patient population will include approximately 30-50 volunteers age >18, male and female, both ischemic and nonischemic cardiomyopathy, with primary and secondary implant indications.

Beschrijving

Inclusion Criteria:

  • age > 18 with ability to give informed consent,
  • patients who are expected to live more than one year,
  • patients with indication for defibrillator implant,
  • patients who are not pacemaker dependent,
  • patients with either ischemic or nonischemic cardiomyopathy.

Exclusion Criteria:

  • age < 18,
  • inability to give informed consent,
  • life expectancy less than one year,
  • pacemaker dependence, or use of preexisting lead for sensing,
  • any condition which would preclude ICD testing at the end of the implant,
  • patients who are undergoing upgrade to an ICD from a pacemaker will also be excluded,
  • patients with inherited arrhythmias or ion channel related arrhythmias will be excluded from the study.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

Cohorten en interventies

Groep / Cohort
cardiomyopathy, with implant indications
30-50 volunteers age >18, male and female with ability to give informed consent, who are expected to live more than one year, with indication for defibrillator implant and who are not pacemaker dependent.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Tijdsspanne
Duration of sensing (time to detection) in VF at the time of implant.
Tijdsspanne: At implant
At implant

Secundaire uitkomstmaten

Uitkomstmaat
Tijdsspanne
Sensing in sinus rhythm at one month
Tijdsspanne: One month post implant
One month post implant

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Medewerkers

Onderzoekers

  • Hoofdonderzoeker: Aneesh Tolat, MD, Attending Cardiac Electrophysiologist, St Francis Hospital and Medical Center

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start

1 december 2009

Primaire voltooiing (Werkelijk)

1 juli 2011

Studie voltooiing (Werkelijk)

1 juli 2011

Studieregistratiedata

Eerst ingediend

14 december 2009

Eerst ingediend dat voldeed aan de QC-criteria

14 december 2009

Eerst geplaatst (Schatting)

15 december 2009

Updates van studierecords

Laatste update geplaatst (Schatting)

28 februari 2012

Laatste update ingediend die voldeed aan QC-criteria

27 februari 2012

Laatst geverifieerd

1 februari 2012

Meer informatie

Termen gerelateerd aan deze studie

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

3
Abonneren