Validity of activity monitors in health and chronic disease: a systematic review

Hans Van Remoortel, Santiago Giavedoni, Yogini Raste, Chris Burtin, Zafeiris Louvaris, Elena Gimeno-Santos, Daniel Langer, Alastair Glendenning, Nicholas S Hopkinson, Ioannis Vogiatzis, Barry T Peterson, Frederick Wilson, Bridget Mann, Roberto Rabinovich, Milo A Puhan, Thierry Troosters, PROactive consortium, Caterina Brindicci, Tim Higenbottam, Thierry Troosters, Fabienne Dobbels, Marc Decramer, Margaret X Tabberer, Roberto A Rabinovich, William McNee, Ioannis Vogiatzis, Michael Polkey, Nick Hopkinson, Judith Garcia-Aymerich, Milo Puhan, Anja Frei, Thys van der Molen, Corina De Jong, Pim de Boer, Ian Jarrod, Paul McBride, Nadia Kamel, Katja Rudell, Frederick J Wilson, Nathalie Ivanoff, Karoly Kulich, Alistair Glendenning, Niklas X Karlsson, Solange Corriol-Rohou, Enkeleida Nikai, Damijen Erzen, Hans Van Remoortel, Santiago Giavedoni, Yogini Raste, Chris Burtin, Zafeiris Louvaris, Elena Gimeno-Santos, Daniel Langer, Alastair Glendenning, Nicholas S Hopkinson, Ioannis Vogiatzis, Barry T Peterson, Frederick Wilson, Bridget Mann, Roberto Rabinovich, Milo A Puhan, Thierry Troosters, PROactive consortium, Caterina Brindicci, Tim Higenbottam, Thierry Troosters, Fabienne Dobbels, Marc Decramer, Margaret X Tabberer, Roberto A Rabinovich, William McNee, Ioannis Vogiatzis, Michael Polkey, Nick Hopkinson, Judith Garcia-Aymerich, Milo Puhan, Anja Frei, Thys van der Molen, Corina De Jong, Pim de Boer, Ian Jarrod, Paul McBride, Nadia Kamel, Katja Rudell, Frederick J Wilson, Nathalie Ivanoff, Karoly Kulich, Alistair Glendenning, Niklas X Karlsson, Solange Corriol-Rohou, Enkeleida Nikai, Damijen Erzen

Abstract

The assessment of physical activity in healthy populations and in those with chronic diseases is challenging. The aim of this systematic review was to identify whether available activity monitors (AM) have been appropriately validated for use in assessing physical activity in these groups. Following a systematic literature search we found 134 papers meeting the inclusion criteria; 40 conducted in a field setting (validation against doubly labelled water), 86 in a laboratory setting (validation against a metabolic cart, metabolic chamber) and 8 in a field and laboratory setting. Correlation coefficients between AM outcomes and energy expenditure (EE) by the criterion method (doubly labelled water and metabolic cart/chamber) and percentage mean differences between EE estimation from the monitor and EE measurement by the criterion method were extracted. Random-effects meta-analyses were performed to pool the results across studies where possible. Types of devices were compared using meta-regression analyses. Most validation studies had been performed in healthy adults (n=118), with few carried out in patients with chronic diseases (n=16). For total EE, correlation coefficients were statistically significantly lower in uniaxial compared to multisensor devices. For active EE, correlations were slightly but not significantly lower in uniaxial compared to triaxial and multisensor devices. Uniaxial devices tended to underestimate TEE (-12.07 (95%CI; -18.28 to -5.85) %) compared to triaxial (-6.85 (95%CI; -18.20 to 4.49) %, p=0.37) and were statistically significantly less accurate than multisensor devices (-3.64 (95%CI; -8.97 to 1.70) %, p<0.001). TEE was underestimated during slow walking speeds in 69% of the lab validation studies compared to 37%, 30% and 37% of the studies during intermediate, fast walking speed and running, respectively. The high level of heterogeneity in the validation studies is only partly explained by the type of activity monitor and the activity monitor outcome. Triaxial and multisensor devices tend to be more valid monitors. Since activity monitors are less accurate at slow walking speeds and information about validated activity monitors in chronic disease populations is lacking, proper validation studies in these populations are needed prior to their inclusion in clinical trials.

Figures

Figure 1
Figure 1
Flow chart describing the identification and inclusion of relevant studies.
Figure 2
Figure 2
Study-specific correlation coefficients (r) and Fisher z-scores (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from doubly labelled water (TEEDLW). Each dot represents the z-score of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis. CV; coefficient of variation for TEEDLW.
Figure 3
Figure 3
Study-specific % mean difference (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from doubly labelled water (TEEDLW). Each dot represents the mean difference of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis. CV; coefficient of variation for TEEDLW. *Leenders et al. 2006 (Actigraph Model 7164); TEEAM estimated with most frequently used Freedson and Hendelman equation (walking outdoors), (not reported) data of % mean difference (±SD) between TEEAM - TEEDLW with other previously published equations can be found in the original paper [14].
Figure 4
Figure 4
Study-specific correlation coefficients and Fisher z-scores (diamond) between active energy expenditure estimate from the activity monitor (AEEAM) and active energy expenditure measure from doubly labelled water (AEEDLW). Each dot represents the z-score of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis. CV; coefficient of variation for AEEDLW.
Figure 5
Figure 5
Study-specific % mean difference (diamond) between active energy expenditure estimate from the activity monitor (AEEAM) and total energy expenditure measure from doubly labelled water (AEEDLW). Each dot represents the mean difference of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis. CV; coefficient of variation for AEEDLW. *Assah et al. 2009 (Actigraph Model 7164); AEEAM estimated with most frequently used Freedson and Hendelman equation, (not reported) data of % mean difference between AEEAM - AEEDLW with other data derived and previously published equations can be found in the original paper [48].
Figure 6
Figure 6
Study-specific correlation coefficients (r) and Fisher z-scores (diamond) between activity monitor outcomes and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols. Each dot represents the z-score of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis.
Figure 7
Figure 7
Study-specific correlation coefficients and Fisher z-scores (diamond) between activity monitor outcomes and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on walking activities. Each dot represents the z-score of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis.
Figure 8
Figure 8
Study-specific correlation coefficients and Fisher z-scores (diamond) between activity monitor outcomes and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on activities of daily living activities involving the upper and lower limbs. Each dot represents the z-score of the respective study together with a 95% confidence interval (CI) and the size of the box represents the weight of the study in the meta-analysis. Weights are from random effects analysis.
Figure 9
Figure 9
Study-specific % mean difference (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on slow walking speed. Each dot represents the % mean difference of the respective study.
Figure 10
Figure 10
Study-specific % mean difference (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on intermediate walking speed. Each dot represents the % mean difference of the respective study.
Figure 11
Figure 11
Study-specific % mean difference (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on fast walking speed. Each dot represents the % mean difference of the respective study.
Figure 12
Figure 12
Study-specific % mean difference (diamond) between total energy expenditure estimate from the activity monitor (TEEAM) and total energy expenditure measure from indirect calorimetry (TEEIC) during laboratory protocols based on running speed. Each dot represents the % mean difference of the respective study.
Figure 13
Figure 13
Accuracy of steps at different walking speeds. The dots are reflecting walking speed: slow walking (<3.2 km/hr (□)), intermediate walking (3.2-6.4 km/hr (■)) and fast walking (6.5-8 km/hr (▲)).

References

    1. Warburton DE, Nicol CW, Bredin SS. Health benefits of physical activity: the evidence. CMAJ. 2006;174:801–809. doi: 10.1503/cmaj.051351.
    1. Blair SN, Cheng Y, Holder JS. Is physical activity or physical fitness more important in defining health benefits? Med Sci Sports Exerc. 2001;33:S379–S399. doi: 10.1097/00005768-200106001-00007. discussion S419-320.
    1. Garcia-Aymerich J, Lange P, Benet M, Schnohr P, Anto JM. Regular physical activity reduces hospital admission and mortality in chronic obstructive pulmonary disease: a population based cohort study. Thorax. 2006;61:772–778. doi: 10.1136/thx.2006.060145.
    1. Caspersen CJ, Powell KE, Christenson GM. Physical activity, exercise, and physical fitness: definitions and distinctions for health-related research. Public Health Rep. 1985;100:126–131.
    1. Steele BG, Belza B, Cain K, Warms C, Coppersmith J, Howard J. Bodies in motion: monitoring daily activity and exercise with motion sensors in people with chronic pulmonary disease. J Rehabil Res Dev. 2003;40:45–58. doi: 10.1682/JRRD.2003.10.0045.
    1. Garcia-Aymerich J, Serra I, Gomez FP, Farrero E, Balcells E, Rodriguez DA, de Batlle J, Gimeno E, Donaire-Gonzalez D, Orozco-Levi M. et al.Physical activity and clinical and functional status in COPD. Chest. 2009;136:62–70. doi: 10.1378/chest.08-2532.
    1. Bassett DR, Ainsworth BE, Swartz AM, Strath SJ, O’Brien WL, King GA. Validity of four motion sensors in measuring moderate intensity physical activity. Med Sci Sports Exerc. 2000;32:S471–S480. doi: 10.1097/00005768-200009001-00006.
    1. Le Masurier GC, Tudor-Locke C. Comparison of pedometer and accelerometer accuracy under controlled conditions. Med Sci Sports Exerc. 2003;35:867–871. doi: 10.1249/01.MSS.0000064996.63632.10.
    1. Plasqui G, Westerterp KR. Physical activity assessment with accelerometers: an evaluation against doubly labeled water. Obesity (Silver Spring) 2007;15:2371–2379. doi: 10.1038/oby.2007.281.
    1. Cyarto EV, Myers AM, Tudor-Locke C. Pedometer accuracy in nursing home and community-dwelling older adults. Med Sci Sports Exerc. 2004;36:205–209. doi: 10.1249/01.MSS.0000113476.62469.98.
    1. Giacomozzi C, Caselli A, Macellari V, Giurato L, Lardieri L, Uccioli L. Walking strategy in diabetic patients with peripheral neuropathy. Diabetes Care. 2002;25:1451–1457. doi: 10.2337/diacare.25.8.1451.
    1. Guazzi M, Dickstein K, Vicenzi M, Arena R. Six-minute walk test and cardiopulmonary exercise testing in patients with chronic heart failure: a comparative analysis on clinical and prognostic insights. Circ Heart Fail. 2009;2:549–555. doi: 10.1161/CIRCHEARTFAILURE.109.881326.
    1. Corey DM, Dunlap WP. Averaging Correlations: expected Values and Bias in Combined Pearson rs and Fisher’s z Transformations. J Gen Psychol. 1998;125:245–261. doi: 10.1080/00221309809595548.
    1. Leenders NY, Sherman WM, Nagaraja HN. Energy expenditure estimated by accelerometry and doubly labeled water: do they agree? Med Sci Sports Exerc. 2006;38:2165–2172. doi: 10.1249/01.mss.0000235883.94357.95.
    1. Adams SA, Matthews CE, Ebbeling CB, Moore CG, Cunningham JE, Fulton J, Hebert JR. The effect of social desirability and social approval on self-reports of physical activity. Am J Epidemiol. 2005;161:389–398. doi: 10.1093/aje/kwi054.
    1. Leenders NY, Sherman WM, Nagaraja HN, Kien CL. Evaluation of methods to assess physical activity in free-living conditions. Med Sci Sports Exerc. 2001;33:1233–1240.
    1. Dalton AF, Ni Scanaill C, Carew S, Lyons D, OL G. A clinical evaluation of a remote mobility monitoring system based on SMS messaging. Conf Proc IEEE Eng Med Biol Soc. 2007;2007:2327–2330.
    1. Kochersberger G, McConnell E, Kuchibhatla MN, Pieper C. The reliability, validity, and stability of a measure of physical activity in the elderly. Arch Phys Med Rehabil. 1996;77:793–795. doi: 10.1016/S0003-9993(96)90258-0.
    1. Masse LC, Fulton JE, Watson KL, Heesch KC, Kohl HW, Blair SN, Tortolero SR. Detecting bouts of physical activity in a field setting. Res Q Exerc Sport. 1999;70:212–219.
    1. McClain JJ, Craig CL, Sisson SB, Tudor-Locke C. Comparison of Lifecorder EX and ActiGraph accelerometers under free-living conditions. Appl Physiol Nutr Metab. 2007;32:753–761. doi: 10.1139/H07-060.
    1. Paul DR, Kramer M, Moshfegh AJ, Baer DJ, Rumpler WV. Comparison of two different physical activity monitors. BMC Med Res Methodol. 2007;7:26. doi: 10.1186/1471-2288-7-26.
    1. Lof M, Hannestad U, Forsum E. Comparison of commonly used procedures, including the doubly-labelled water technique, in the estimation of total energy expenditure of women with special reference to the significance of body fatness. Br J Nutr. 2003;90:961–968. doi: 10.1079/BJN2003975.
    1. Masse LC, Fulton JE, Watson KL, Mahar MT, Meyers MC, Wong WW. Influence of body composition on physical activity validation studies using doubly labeled water. J Appl Physiol. 2004;96:1357–1364. doi: 10.1152/japplphysiol.00901.2003.
    1. Patrik Johansson H, Rossander-Hulthen L, Slinde F, Ekblom B. Accelerometry combined with heart rate telemetry in the assessment of total energy expenditure. Br J Nutr. 2006;95:631–639. doi: 10.1079/BJN20051527.
    1. Brooks AG, Gunn SM, Withers RT, Gore CJ, Plummer JL. Predicting walking METs and energy expenditure from speed or accelerometry. Med Sci Sports Exerc. 2005;37:1216–1223. doi: 10.1249/01.mss.0000170074.19649.0e.
    1. Crouter SE, Churilla JR, Bassett DR. Estimating energy expenditure using accelerometers. Eur J Appl Physiol. 2006;98:601–612. doi: 10.1007/s00421-006-0307-5.
    1. Focht BC. Initial Validation of the CSA Activity Monitor During Rehabilitative Exercise Among Older Adults With Chronic Disease. Journal of Aging and Physical Activity. 2003;11:293–304.
    1. Freedson PS, Melanson E, Sirard J. Calibration of the Computer Science and Applications, Inc. accelerometer. Med Sci Sports Exerc. 1998;30:777–781. doi: 10.1097/00005768-199805000-00021.
    1. Fudge BW, Wilson J, Easton C, Irwin L, Clark J, Haddow O, Kayser B, Pitsiladis YP. Estimation of oxygen uptake during fast running using accelerometry and heart rate. Med Sci Sports Exerc. 2007;39:192–198. doi: 10.1249/01.mss.0000235884.71487.21.
    1. Heil DP. Body size as a determinant of activity moniyor output during overground walking. JEP online. 2003;6:1–11.
    1. Hendelman D, Miller K, Baggett C, Debold E, Freedson P. Validity of accelerometry for the assessment of moderate intensity physical activity in the field. Med Sci Sports Exerc. 2000;32:S442–S449. doi: 10.1097/00005768-200009001-00002.
    1. King GA, Torres N, Potter C, Brooks TJ, Coleman KJ. Comparison of activity monitors to estimate energy cost of treadmill exercise. Med Sci Sports Exerc. 2004;36:1244–1251. doi: 10.1249/01.MSS.0000132379.09364.F8.
    1. Leenders NY, Nelson TE, Sherman WM. Ability of different physical activity monitors to detect movement during treadmill walking. Int J Sports Med. 2003;24:43–50. doi: 10.1055/s-2003-37196.
    1. Melanson EL, Freedson PS. Validity of the Computer Science and Applications, Inc. (CSA) activity monitor. Med Sci Sports Exerc. 1995;27:934–940.
    1. Nichols JF, Patterson P, Early T. A validation of a physical activity monitor for young and older adults. Can J Sport Sci. 1992;17:299–303.
    1. Patterson SM, Krantz DS, Montgomery LC, Deuster PA, Hedges SM, Nebel LE. Automated physical activity monitoring: validation and comparison with physiological and self-report measures. Psychophysiology. 1993;30:296–305. doi: 10.1111/j.1469-8986.1993.tb03356.x.
    1. Slootmaker SM, Chin APMJ, Schuit AJ, van Mechelen W, Koppes LL. Concurrent validity of the PAM accelerometer relative to the MTI Actigraph using oxygen consumption as a reference. Scand J Med Sci Sports. 2009;19:36–43.
    1. Stone MR, Esliger DW, Tremblay MS. Comparative validity assessment of five activity monitors: does being a child matter? Pediatr Exerc Sci. 2007;19:291–309.
    1. Strath SJ, Bassett DR, Thompson DL, Swartz AM. Validity of the simultaneous heart rate-motion sensor technique for measuring energy expenditure. Med Sci Sports Exerc. 2002;34:888–894. doi: 10.1097/00005768-200205000-00025.
    1. Swartz AM, Strath SJ, Bassett DR, O’Brien WL, King GA, Ainsworth BE. Estimation of energy expenditure using CSA accelerometers at hip and wrist sites. Med Sci Sports Exerc. 2000;32:S450–S456. doi: 10.1097/00005768-200009001-00003.
    1. Welk GJ, Blair SN, Wood K, Jones S, Thompson RW. A comparative evaluation of three accelerometry-based physical activity monitors. Med Sci Sports Exerc. 2000;32:S489–S497. doi: 10.1097/00005768-200009001-00008.
    1. Crouter SE, Clowers KG, Bassett DR. A novel method for using accelerometer data to predict energy expenditure. J Appl Physiol. 2006;100:1324–1331.
    1. Nichols JF, Morgan CG, Chabot LE, Sallis JF, Calfas KJ. Assessment of physical activity with the Computer Science and Applications, Inc., accelerometer: laboratory versus field validation. Res Q Exerc Sport. 2000;71:36–43.
    1. Esliger DW, Probert A, Gorber SC, Bryan S, Laviolette M, Tremblay MS. Validity of the Actical accelerometer step-count function. Med Sci Sports Exerc. 2007;39:1200–1204. doi: 10.1249/mss.0b013e3804ec4e9.
    1. Storti KL, Pettee KK, Brach JS, Talkowski JB, Richardson CR, Kriska AM. Gait speed and step-count monitor accuracy in community-dwelling older adults. Med Sci Sports Exerc. 2008;40:59–64.
    1. Berntsen S, Hageberg R, Aandstad A, Mowinckel P, Anderssen SA, Carlsen KH, Andersen LB. Validity of physical activity monitors in adults participating in free-living activities. Br J Sports Med. 2010;44:657–664. doi: 10.1136/bjsm.2008.048868.
    1. Miller NE, Strath SJ, Swartz AM, Cashin SE. Estimating absolute and relative physical activity intensity across age via accelerometry in adults. J Aging Phys Act. 2010;18:158–170.
    1. Assah FK, Ekelund U, Brage S, Corder K, Wright A, Mbanya JC, Wareham NJ. Predicting physical activity energy expenditure using accelerometry in adults from sub-Sahara Africa. Obesity (Silver Spring) 2009;17:1588–1595. doi: 10.1038/oby.2009.39.
    1. Rothney MP, Brychta RJ, Meade NN, Chen KY, Buchowski MS. Validation of the ActiGraph two-regression model for predicting energy expenditure. Med Sci Sports Exerc. 2010;42:1785–1792. doi: 10.1249/MSS.0b013e3181d5a984.
    1. Colbert LH, Matthews CE, Havighurst TC, Kim K, Schoeller DA. Comparative validity of physical activity measures in older adults. Med Sci Sports Exerc. 2011;43:867–876.
    1. Abel MG, Hannon JC, Sell K, Lillie T, Conlin G, Anderson D. Validation of the Kenz Lifecorder EX and ActiGraph GT1M accelerometers for walking and running in adults. Appl Physiol Nutr Metab. 2008;33:1155–1164. doi: 10.1139/h08-103.
    1. Edwards AG, Hill JO, Byrnes WC, Browning RC. Accuracy of optimized branched algorithms to assess activity-specific physical activity energy expenditure. Med Sci Sports Exerc. 2010;42:672–682.
    1. Kozey SL, Lyden K, Howe CA, Staudenmayer JW, Freedson PS. Accelerometer output and MET values of common physical activities. Med Sci Sports Exerc. 2010;42:1776–1784. doi: 10.1249/MSS.0b013e3181d479f2.
    1. Lyden K, Kozey SL, Staudenmeyer JW, Freedson PS. A comprehensive evaluation of commonly used accelerometer energy expenditure and MET prediction equations. Eur J Appl Physiol. 2011;111:187–201. doi: 10.1007/s00421-010-1639-8.
    1. Esliger DW, Rowlands AV, Hurst TL, Catt M, Murray P, Eston RG. Validation of the GENEA Accelerometer. Med Sci Sports Exerc. 2010;43:1085–1093.
    1. Kuffel EE, Crouter SE, Haas JD, Frongillo EA, Bassett DR. Validity of estimating minute-by-minute energy expenditure of continuous walking bouts by accelerometry. Int J Behav Nutr Phys Act. 2011;8:92. doi: 10.1186/1479-5868-8-92.
    1. Harrington DM, Welk GJ, Donnelly AE. Validation of MET estimates and step measurement using the ActivPAL physical activity logger. J Sports Sci. 2011;29:627–633. doi: 10.1080/02640414.2010.549499.
    1. Fogelholm M, Hiilloskorpi H, Laukkanen R, Oja P, Van Marken Lichtenbelt W, Westerterp K. Assessment of energy expenditure in overweight women. Med Sci Sports Exerc. 1998;30:1191–1197. doi: 10.1097/00005768-199808000-00002.
    1. Fuller Z, Horgan G, O’Reilly LM, Ritz P, Milne E, Stubbs RJ. Comparing different measures of energy expenditure in human subjects resident in a metabolic facility. Eur J Clin Nutr. 2008;62:560–569. doi: 10.1038/sj.ejcn.1602739.
    1. Gardner AW, Womack CJ, Sieminski DJ, Montgomery PS, Killewich LA, Fonong T. Relationship between free-living daily physical activity and ambulatory measures in older claudicants. Angiology. 1998;49:327–337. doi: 10.1177/000331979804900501.
    1. Starling RD, Matthews DE, Ades PA, Poehlman ET. Assessment of physical activity in older individuals: a doubly labeled water study. J Appl Physiol. 1999;86:2090–2096.
    1. Choquette S, Chuin A, Lalancette DA, Brochu M, Dionne IJ. Predicting energy expenditure in elders with the metabolic cost of activities. Med Sci Sports Exerc. 2009;41:1915–1920. doi: 10.1249/MSS.0b013e3181a6164a.
    1. Fehling PC, Smith DL, Warner SE, Dalsky GP. Comparison of accelerometers with oxygen consumption in older adults during exercise. Med Sci Sports Exerc. 1999;31:171–175. doi: 10.1097/00005768-199901000-00026.
    1. Hageman PA. Comparison of two Physical Activity Monitors During a 1-Mile Walking field test. JEP online. 2004;7:102–110.
    1. Pambianco G, Wing RR, Robertson R. Accuracy and reliability of the Caltrac accelerometer for estimating energy expenditure. Med Sci Sports Exerc. 1990;22:858–862.
    1. Rafamantanantsoa HH, Ebine N, Yoshioka M, Higuchi H, Yoshitake Y, Tanaka H, Saitoh S, Jones PJ. Validation of three alternative methods to measure total energy expenditure against the doubly labeled water method for older Japanese men. J Nutr Sci Vitaminol (Tokyo) 2002;48:517–523. doi: 10.3177/jnsv.48.517.
    1. Yamada Y, Yokoyama K, Noriyasu R, Osaki T, Adachi T, Itoi A, Naito Y, Morimoto T, Kimura M, Oda S. Light-intensity activities are important for estimating physical activity energy expenditure using uniaxial and triaxial accelerometers. Eur J Appl Physiol. 2009;105:141–152. doi: 10.1007/s00421-008-0883-7.
    1. Kumahara H, Schutz Y, Ayabe M, Yoshioka M, Yoshitake Y, Shindo M, Ishii K, Tanaka H. The use of uniaxial accelerometry for the assessment of physical-activity-related energy expenditure: a validation study against whole-body indirect calorimetry. Br J Nutr. 2004;91:235–243. doi: 10.1079/BJN20031033.
    1. Schneider PL, Crouter SE, Lukajic O, Bassett DR. Accuracy and reliability of 10 pedometers for measuring steps over a 400-m walk. Med Sci Sports Exerc. 2003;35:1779–1784. doi: 10.1249/01.MSS.0000089342.96098.C4.
    1. Ohkawara K, Ishikawa-Takata K, Park JH, Tabata I, Tanaka S. How much locomotive activity is needed for an active physical activity level: analysis of total step counts. BMC Res Notes. 2011;4:512. doi: 10.1186/1756-0500-4-512.
    1. Yokoyama Y, Kawamura T, Tamakoshi A, Noda A, Hirai M, Saito H, Ohno Y. Comparison of accelerometry and oxymetry for measuring daily physical activity. Circ J. 2002;66:751–754. doi: 10.1253/circj.66.751.
    1. Grant PM, Dall PM, Mitchell SL, Granat MH. Activity-monitor accuracy in measuring step number and cadence in community-dwelling older adults. J Aging Phys Act. 2008;16:201–214.
    1. Ryan CG, Grant PM, Tigbe WW, Granat MH. The validity and reliability of a novel activity monitor as a measure of walking. Br J Sports Med. 2006;40:779–784. doi: 10.1136/bjsm.2006.027276.
    1. Maddocks M, Petrou A, Skipper L, Wilcock A. Validity of three accelerometers during treadmill walking and motor vehicle travel. Br J Sports Med. 2010;44:606–608. doi: 10.1136/bjsm.2008.051128.
    1. Chen KY, Acra SA, Majchrzak K, Donahue CL, Baker L, Clemens L, Sun M, Buchowski MS. Predicting energy expenditure of physical activity using hip- and wrist-worn accelerometers. Diabetes Technol Ther. 2003;5:1023–1033. doi: 10.1089/152091503322641088.
    1. Welk GJ, Almeida J, Morss G. Laboratory calibration and validation of the Biotrainer and Actitrac activity monitors. Med Sci Sports Exerc. 2003;35:1057–1064. doi: 10.1249/01.MSS.0000069525.56078.22.
    1. Kane NA, Simmons MC, John D, Thompson DL, Bassett DR. Validity of the Nike+ device during walking and running. Int J Sports Med. 2010;31:101–105. doi: 10.1055/s-0029-1242810.
    1. Brugniaux JV, Niva A, Pulkkinen I, Laukkanen RM, Richalet JP, Pichon AP. Polar Activity Watch 200: a new device to accurately assess energy expenditure. Br J Sports Med. 2010;44:245–249. doi: 10.1136/bjsm.2007.045575.
    1. Connolly CP, Coe DP, Kendrick JM, Bassett DR, Thompson DL. Accuracy of physical activity monitors in pregnant women. Med Sci Sports Exerc. 2010;43:1100–1105.
    1. Jacobi D, Perrin AE, Grosman N, Dore MF, Normand S, Oppert JM, Simon C. Physical activity-related energy expenditure with the RT3 and TriTrac accelerometers in overweight adults. Obesity (Silver Spring) 2007;15:950–956. doi: 10.1038/oby.2007.605.
    1. Maddison R, Jiang Y, Hoorn SV, Mhurchu CN, Lawes CM, Rodgers A, Rush E. Estimating energy expenditure with the RT3 triaxial accelerometer. Res Q Exerc Sport. 2009;80:249–256.
    1. DeVoe D, Gotshall R, McArthur T. Comparison of the RT3 Research Tracker and Tritrac R3D accelerometers. Percept Mot Skills. 2003;97:510–518.
    1. Howe CA, Staudenmayer JW, Freedson PS. Accelerometer prediction of energy expenditure: vector magnitude versus vertical axis. Med Sci Sports Exerc. 2009;41:2199–2206. doi: 10.1249/MSS.0b013e3181aa3a0e.
    1. Campbell KL, Crocker PR, McKenzie DC. Field evaluation of energy expenditure in women using Tritrac accelerometers. Med Sci Sports Exerc. 2002;34:1667–1674. doi: 10.1097/00005768-200210000-00020.
    1. Chen KY, Sun M. Improving energy expenditure estimation by using a triaxial accelerometer. J Appl Physiol. 1997;83:2112–2122.
    1. Jakicic JM, Winters C, Lagally K, Ho J, Robertson RJ, Wing RR. The accuracy of the TriTrac-R3D accelerometer to estimate energy expenditure. Med Sci Sports Exerc. 1999;31:747–754. doi: 10.1097/00005768-199905000-00020.
    1. Plasqui G, Joosen AM, Kester AD, Goris AH, Westerterp KR. Measuring free-living energy expenditure and physical activity with triaxial accelerometry. Obes Res. 2005;13:1363–1369. doi: 10.1038/oby.2005.165.
    1. Bouten CV, Verboeket-van de Venne WP, Westerterp KR, Verduin M, Janssen JD. Daily physical activity assessment: comparison between movement registration and doubly labeled water. J Appl Physiol. 1996;81:1019–1026.
    1. Meijer EP, Goris AH, Wouters L, Westerterp KR. Physical inactivity as a determinant of the physical activity level in the elderly. Int J Obes Relat Metab Disord. 2001;25:935–939. doi: 10.1038/sj.ijo.0801644.
    1. Bonomi AG, Plasqui G, Goris AH, Westerterp KR. Estimation of free-living energy expenditure using a novel activity monitor designed to minimize obtrusiveness. Obesity (Silver Spring) 2010;18:1845–1851. doi: 10.1038/oby.2010.34.
    1. Goris AH, Meijer EP, Kester A, Westerterp KR. Use of a triaxial accelerometer to validate reported food intakes. Am J Clin Nutr. 2001;73:549–553.
    1. Meijer GA, Westerterp KR, van Hulsel AM, ten Hoor F. Physical activity and energy expenditure in lean and obese adult human subjects. Eur J Appl Physiol Occup Physiol. 1992;65:525–528. doi: 10.1007/BF00602359.
    1. Pietilainen KH, Kaprio J, Borg P, Plasqui G, Yki-Jarvinen H, Kujala UM, Rose RJ, Westerterp KR, Rissanen A. Physical inactivity and obesity: a vicious circle. Obesity (Silver Spring) 2008;16:409–414. doi: 10.1038/oby.2007.72.
    1. Verbunt JA, Westerterp KR, van der Heijden GJ, Seelen HA, Vlaeyen JW, Knottnerus JA. Physical activity in daily life in patients with chronic low back pain. Arch Phys Med Rehabil. 2001;82:726–730. doi: 10.1053/apmr.2001.23182.
    1. Bouten CV, Koekkoek KT, Verduin M, Kodde R, Janssen JD. A triaxial accelerometer and portable data processing unit for the assessment of daily physical activity. IEEE Trans Biomed Eng. 1997;44:136–147. doi: 10.1109/10.554760.
    1. Bouten CV, Westerterp KR, Verduin M, Janssen JD. Assessment of energy expenditure for physical activity using a triaxial accelerometer. Med Sci Sports Exerc. 1994;26:1516–1523.
    1. Levine JA, Baukol PA, Westerterp KR. Validation of the Tracmor triaxial accelerometer system for walking. Med Sci Sports Exerc. 2001;33:1593–1597. doi: 10.1097/00005768-200109000-00024.
    1. Bonomi AG, Plasqui G, Goris AH, Westerterp KR. Estimation of free-living energy expenditure using a novel activity monitor designed to minimize obtrusiveness. Obesity (Silver Spring) 2010;18:1845–1851. doi: 10.1038/oby.2010.34.
    1. Brandes M, Rosenbaum D. Correlations between the step activity monitor and the DynaPort ADL-monitor. Clin Biomech (Bristol, Avon) 2004;19:91–94. doi: 10.1016/j.clinbiomech.2003.08.001.
    1. Langer D, Gosselink R, Sena R, Burtin C, Decramer M, Troosters T. Validation of two activity monitors in patients with COPD. Thorax. 2009;64:641–642. doi: 10.1136/thx.2008.112102.
    1. van Hees VT, van Lummel RC, Westerterp KR. Estimating activity-related energy expenditure under sedentary conditions using a tri-axial seismic accelerometer. Obesity (Silver Spring) 2009;17:1287–1292.
    1. Carter J, Wilkinson D, Blacker S, Rayson M, Bilzon J, Izard R, Coward A, Wright A, Nevill A, Rennie K. et investigation of a novel three-dimensional activity monitor to predict free-living energy expenditure. J Sports Sci. 2008;26:553–561. doi: 10.1080/02640410701708979.
    1. Horner FE, Rayson MP, Bilzon JL. Reliability and validity of the 3DNX accelerometer during mechanical and human treadmill exercise testing. Int J Obes (Lond) 2011;35(Suppl 1):S88–S97.
    1. Hikihara Y, Tanaka S, Ohkawara K, Ishikawa-Takata K, Tabata I. Validation and Comparison of Three Accelerometers for Measuring Physical Activity Intensity During Nonlocomotive Activities and Locomotive Movements. J Phys Act Health. 2011; : .
    1. Dugas LR, Carstens MA, Ebersole K, Schoeller DA, Durazo-Arvizu RA, Lambert EV, Luke A. Energy expenditure in young adult urban informal settlement dwellers in South Africa. Eur J Clin Nutr. 2009;63:805–807. doi: 10.1038/ejcn.2008.75.
    1. Heil DP. Predicting activity energy expenditure using the Actical activity monitor. Res Q Exerc Sport. 2006;77:64–80. doi: 10.5641/027013606X13080769703920.
    1. Crouter SE, Schneider PL, Karabulut M, Bassett DR. Validity of 10 electronic pedometers for measuring steps, distance, and energy cost. Med Sci Sports Exerc. 2003;35:1455–1460. doi: 10.1249/01.MSS.0000078932.61440.A2.
    1. Spierer DK, Hagins M, Rundle A, Pappas E. A comparison of energy expenditure estimates from the Actiheart and Actical physical activity monitors during low intensity activities, walking, and jogging. Eur J Appl Physiol. 2011;111:659–667. doi: 10.1007/s00421-010-1672-7.
    1. Crouter SE, Dellavalle DM, Horton M, Haas JD, Frongillo EA, Bassett DR. Validity of the Actical for estimating free-living physical activity. Eur J Appl Physiol. 2011;111:1381–1389. doi: 10.1007/s00421-010-1758-2.
    1. Heiermann S, Khalaj Hedayati K, Muller MJ, Dittmar M. Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry. Gerontology. 2010;57:473–479.
    1. Duncan GE, Lester J, Migotsky S, Goh J, Higgins L, Borriello G. Accuracy of a novel multi-sensor board for measuring physical activity and energy expenditure. Eur J Appl Physiol. 2011;111:2025–2032. doi: 10.1007/s00421-011-1834-2.
    1. Hooker SP, Feeney A, Hutto B, Pfeiffer KA, McIver K, Heil DP, Vena JE, Lamonte MJ, Blair SN. Validation of the actical activity monitor in middle-aged and older adults. J Phys Act Health. 2011;8:372–381.
    1. Atallah L, Leong JJ, Lo B, Yang GZ. Energy expenditure prediction using a miniaturized ear-worn sensor. Med Sci Sports Exerc. 2011;43:1369–1377. doi: 10.1249/MSS.0b013e3182093014.
    1. Manohar C, McCrady S, Pavlidis IT, Levine JA. An accelerometer-based earpiece to monitor and quantify physical activity. J Phys Act Health. 2009;6:781–789.
    1. van Hees VT, Renstrom F, Wright A, Gradmark A, Catt M, Chen KY, Lof M, Bluck L, Pomeroy J, Wareham NJ. et al.Estimation of daily energy expenditure in pregnant and non-pregnant women using a wrist-worn tri-axial accelerometer. PLoS One. 2011;6:e22922. doi: 10.1371/journal.pone.0022922.
    1. Ohkawara K, Oshima Y, Hikihara Y, Ishikawa-Takata K, Tabata I, Tanaka S. Real-time estimation of daily physical activity intensity by a triaxial accelerometer and a gravity-removal classification algorithm. Br J Nutr. 2011;105:1681–1691. doi: 10.1017/S0007114510005441.
    1. Annegarn J, Spruit MA, Uszko-Lencer NH, Vanbelle S, Savelberg HH, Schols AM, Wouters EF, Meijer K. Objective physical activity assessment in patients with chronic organ failure: a validation study of a new single-unit activity monitor. Arch Phys Med Rehabil. 2011;92:1852–1857. doi: 10.1016/j.apmr.2011.06.021. e1851.
    1. Harris AM, McAlpine DE, Shirbhate R, Manohar CU, Levine JA. Measurement of daily activity in restrictive type anorexia nervosa. Int J Eat Disord. 2008;41:280–283. doi: 10.1002/eat.20486.
    1. Arvidsson D, Slinde F, Nordenson A, Larsson S, Hulthen L. Validity of the ActiReg system in assessing energy requirement in chronic obstructive pulmonary disease patients. Clin Nutr. 2006;25:68–74. doi: 10.1016/j.clnu.2005.09.001.
    1. Hustvedt BE, Christophersen A, Johnsen LR, Tomten H, McNeill G, Haggarty P, Lovo A. Description and validation of the ActiReg: a novel instrument to measure physical activity and energy expenditure. Br J Nutr. 2004;92:1001–1008. doi: 10.1079/BJN20041272.
    1. Bussmann JB, Hartgerink I, van der Woude LH, Stam HJ. Measuring physical strain during ambulation with accelerometry. Med Sci Sports Exerc. 2000;32:1462–1471. doi: 10.1097/00005768-200008000-00015.
    1. Mignault D, St-Onge M, Karelis AD, Allison DB, Rabasa-Lhoret R. Evaluation of the Portable HealthWear Armband: a device to measure total daily energy expenditure in free-living type 2 diabetic individuals. Diabetes Care. 2005;28:225–227.
    1. St-Onge M, Mignault D, Allison DB, Rabasa-Lhoret R. Evaluation of a portable device to measure daily energy expenditure in free-living adults. Am J Clin Nutr. 2007;85:742–749.
    1. Johannsen DL, Calabro MA, Stewart J, Franke W, Rood JC, Welk GJ. Accuracy of armband monitors for measuring daily energy expenditure in healthy adults. Med Sci Sports Exerc. 2010;42:2134–2140. doi: 10.1249/MSS.0b013e3181e0b3ff.
    1. Koehler K, Braun H, de Marees M, Fusch G, Fusch C, Schaenzer W. Assessing energy expenditure in male endurance athletes: validity of the SenseWear Armband. Med Sci Sports Exerc. 2011;43:1328–1333. doi: 10.1249/MSS.0b013e31820750f5.
    1. Mackey DC, Manini TM, Schoeller DA, Koster A, Glynn NW, Goodpaster BH, Satterfield S, Newman AB, Harris TB, Cummings SR. Validation of an armband to measure daily energy expenditure in older adults. J Gerontol A Biol Sci Med Sci. 2011;66:1108–1113.
    1. Cole PJ, LeMura LM, Klinger TA, Strohecker K, McConnell TR. Measuring energy expenditure in cardiac patients using the Body Media Armband versus indirect calorimetry. A validation study. J Sports Med Phys Fitness. 2004;44:262–271.
    1. Fruin ML, Rankin JW. Validity of a multi-sensor armband in estimating rest and exercise energy expenditure. Med Sci Sports Exerc. 2004;36:1063–1069. doi: 10.1249/01.MSS.0000128144.91337.38.
    1. Jakicic JM, Marcus M, Gallagher KI, Randall C, Thomas E, Goss FL, Robertson RJ. Evaluation of the SenseWear Pro Armband to assess energy expenditure during exercise. Med Sci Sports Exerc. 2004;36:897–904.
    1. Patel SA, Benzo RP, Slivka WA, Sciurba FC. Activity monitoring and energy expenditure in COPD patients: a validation study. COPD. 2007;4:107–112. doi: 10.1080/15412550701246658.
    1. Dwyer TJ, Alison JA, McKeough ZJ, Elkins MR, Bye PT. Evaluation of the SenseWear activity monitor during exercise in cystic fibrosis and in health. Respir Med. 2009;103:1511–1517. doi: 10.1016/j.rmed.2009.04.013.
    1. Furlanetto KC, Bisca GW, Oldemberg N, Sant’anna TJ, Morakami FK, Camillo CA, Cavalheri V, Hernandes NA, Probst VS, Ramos EM. et al.Step counting and energy expenditure estimation in patients with chronic obstructive pulmonary disease and healthy elderly: accuracy of 2 motion sensors. Arch Phys Med Rehabil. 2010;91:261–267. doi: 10.1016/j.apmr.2009.10.024.
    1. Hill K, Dolmage TE, Woon L, Goldstein R, Brooks D. Measurement properties of the SenseWear armband in adults with chronic obstructive pulmonary disease. Thorax. 2010;65:486–491. doi: 10.1136/thx.2009.128702.
    1. Drenowatz C, Eisenmann JC. Validation of the SenseWear Armband at high intensity exercise. Eur J Appl Physiol. 2011;111:883–887. doi: 10.1007/s00421-010-1695-0.
    1. Cavalheri V, Donaria L, Ferreira T, Finatti M, Camillo CA, Cipulo Ramos EM, Pitta F. Energy expenditure during daily activities as measured by two motion sensors in patients with COPD. Respir Med. 2011;105:922–929. doi: 10.1016/j.rmed.2011.01.004.
    1. Berntsen S, Stafne SN, Morkved S. Physical activity monitor for recording energy expenditure in pregnancy. Acta Obstet Gynecol Scand. 2010;90:903–907.
    1. Brazeau AS, Karelis AD, Mignault D, Lacroix MJ, Prud’homme D, Rabasa-Lhoret R. Accuracy of the SenseWear Armband during ergocycling. Int J Sports Med. 2011;32:761–764. doi: 10.1055/s-0031-1279768.
    1. Munguia-Izquierdo D, Santalla A, Legaz-Arrese A. Evaluation of a wearable body monitoring device during treadmill walking and jogging in patients with fibromyalgia syndrome. Arch Phys Med Rehabil. 2012;93:115–122. doi: 10.1016/j.apmr.2011.08.021.
    1. Crouter SE, Churilla JR, Bassett DR. Accuracy of the Actiheart for the assessment of energy expenditure in adults. Eur J Clin Nutr. 2008;62:704–711. doi: 10.1038/sj.ejcn.1602766.
    1. Lof M. Physical activity pattern and activity energy expenditure in healthy pregnant and non-pregnant Swedish women. Eur J Clin Nutr. 2011;65:1295–1301. doi: 10.1038/ejcn.2011.129.
    1. Kwon S, Jamal M, Zamba GK, Stumbo P, Samuel I. Validation of a novel physical activity assessment device in morbidly obese females. J Obes. 2010;2010:1–8.
    1. Borrell B. Epidemiology: every bite you take. Nature. 2011;470:320–322. doi: 10.1038/470320a.
    1. Schenk AK, Witbrodt BC, Hoarty CA, Carlson RH, Goulding EH, Potter JF, Bonasera SJ. Cellular telephones measure activity and lifespace in community-dwelling adults: proof of principle. J Am Geriatr Soc. 2011;59:345–352. doi: 10.1111/j.1532-5415.2010.03267.x.
    1. Ades PA, Savage PD, Brochu M, Tischler MD, Lee NM, Poehlman ET. Resistance training increases total daily energy expenditure in disabled older women with coronary heart disease. J Appl Physiol. 2005;98:1280–1285. doi: 10.1152/japplphysiol.00360.2004.
    1. Gosselink R, Troosters T, Decramer M. Peripheral muscle weakness contributes to exercise limitation in COPD. Am J Respir Crit Care Med. 1996;153:976–980.
    1. Balashov K, Feldman DE, Savard S, Houde S, Frenette M, Ducharme A, Giannetti N, Michel C, Pilote L. Percent predicted value for the 6-minute walk test: using norm-referenced equations to characterize severity in persons with CHF. J Card Fail. 2008;14:75–81. doi: 10.1016/j.cardfail.2007.09.005.
    1. Pitta F, Troosters T, Probst VS, Spruit MA, Decramer M, Gosselink R. Quantifying physical activity in daily life with questionnaires and motion sensors in COPD. Eur Respir J. 2006;27:1040–1055. doi: 10.1183/09031936.06.00064105.
    1. Tudor-Locke C, Williams JE, Reis JP, Pluto D. Utility of pedometers for assessing physical activity: convergent validity. Sports Med. 2002;32:795–808. doi: 10.2165/00007256-200232120-00004.
    1. Baarends EM, Schols AM, Pannemans DL, Westerterp KR, Wouters EF. Total free living energy expenditure in patients with severe chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 1997;155:549–554.
    1. Vorrink SN, Kort HS, Troosters T, Lammers JW. Level of daily physical activity in individuals with COPD compared with healthy controls. Respir Res. 2011;12:33. doi: 10.1186/1465-9921-12-33.
    1. Watz H, Waschki B, Meyer T, Magnussen H. Physical activity in patients with COPD. Eur Respir J. 2009;33:262–272.
    1. Murphy SL. Review of physical activity measurement using accelerometers in older adults: considerations for research design and conduct. Prev Med. 2009;48:108–114. doi: 10.1016/j.ypmed.2008.12.001.
    1. Schmidt ME, Steindorf K. Statistical methods for the validation of questionnaires–discrepancy between theory and practice. Methods Inf Med. 2006;45:409–413.
    1. Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.

Source: PubMed

3
Abonneren