Preserved periprosthetic bone stock at 5 years post-operatively with uncemented short hip stem in both collared and collarless version

Ola Belfrage, Erik Weber, Martin Sundberg, Gunnar Flivik, Ola Belfrage, Erik Weber, Martin Sundberg, Gunnar Flivik

Abstract

Introduction: Previous bone density studies have generally shown bone resorption around both cemented and uncemented total hip arthroplasty (THA) stems. This is presumed to be due to stress shielding. Short stems have been introduced partly to preserve bone in the proximal femur by a more physiological loading of the bone. The purpose of this study was to evaluate bone remodeling around a short, fully hydroxyapatite-coated titanium stem that comes in a collared and collarless version.

Patients and methods: A prospective cohort of 50 patients included in a study evaluating the Furlong Evolution stem has been followed for 5 years. Examination was done with dual energy X-ray absorptiometry (DXA) postoperatively, at 1, 2 and 5 years. Clinical outcome was followed with radiography and both general and hip specific outcome measures.

Results: The two versions of the stem behaved similarly regarding bone remodeling. After an initial decrease up to 1 year, bone mineral density (BMD) increased in all Gruen zones up to 2 years and at 5 years bone stock was still preserved compared with postoperatively (net BMD + 1.2% (95% CI - 0.4 to 2.8)). Increase in BMD occurred mainly in the greater trochanter and distally around the stem with a decrease in the calcar area. Both versions showed excellent clinical outcome up to 5 years.

Conclusion: This short stem seems to preserve proximal bone stock up to 5 years, exhibiting similar behaviour both with and without a collar.

Trial registration number and date of registration: ClinicalTrials.gov, (identifier: NCT01894854). July 10, 2013.

Keywords: Bone density; DXA; Short stem THA; THA.

Conflict of interest statement

GF had paid presentation at symposium with JRI, otherwise the authors have no conflicts of interest related to this work.

© 2021. The Author(s).

Figures

Fig. 1
Fig. 1
Stem design in lateral view, the collared and the collarless version with 133-degree neck shaft angle
Fig. 2
Fig. 2
CONSORT flow diagram
Fig. 3
Fig. 3
Net BMD vs. time
Fig. 4
Fig. 4
Radiograph demonstrating the seven Gruen zones. Graphs represent the change in BMD in percentage (y-axis) over time in years (x-axis). Blue = collarless; red = collared

References

    1. Pitto RP, Mueller LA, Reilly K, Schmidt R, Munro J. Quantitative computer-assisted osteodensitometry in total hip arthroplasty. Int Orthop. 2007;31(4):431–438. doi: 10.1007/s00264-006-0257-x.
    1. Knutsen AR, Lau N, Longjohn DB, Ebramzadeh E, Sangiorgio SN. Periprosthetic femoral bone loss in total hip arthroplasty: systematic analysis of the effect of stem design. Hip Int. 2017;27(1):26–34. doi: 10.5301/hipint.5000413.
    1. Salemyr M, Muren O, Ahl T, Boden H, Eisler T, Stark A, Skoldenberg O. Lower periprosthetic bone loss and good fixation of an ultra-short stem compared to a conventional stem in uncemented total hip arthroplasty. Acta Orthop. 2015;86(6):659–666. doi: 10.3109/17453674.2015.1067087.
    1. Kim YH, Park JW, Kim JS. Ultrashort versus conventional anatomic cementless femoral stems in the same patients younger than 55 years. Clin Orthop Relat Res. 2016;474(9):2008–2017. doi: 10.1007/s11999-016-4902-4.
    1. Liang HD, Yang WY, Pan JK, Huang HT, Luo MH, Zeng LF, Liu J. Are short-stem prostheses superior to conventional stem prostheses in primary total hip arthroplasty? A systematic review and meta-analysis of randomised controlled trials. BMJ Open. 2018;8(9):e021649. doi: 10.1136/bmjopen-2018-021649.
    1. Venesmaa PK, Kroger HP, Miettinen HJ, Jurvelin JS, Suomalainen OT, Alhava EM. Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry–a 3-year follow-up study. J Bone Miner Res. 2001;16(6):1056–1061. doi: 10.1359/jbmr.2001.16.6.1056.
    1. Rahmy AIA, Gosens T, Blake GM, Tonino A, Fogelman IJOI. Periprosthetic bone remodelling of two types of uncemented femoral implant with proximal hydroxyapatite coating: a 3-year follow-up study addressing the influence of prosthesis design and preoperative bone density on periprosthetic bone loss. Osteoporos Int. 2004;15(4):281–289. doi: 10.1007/s00198-003-1546-5.
    1. Shetty AA, Slack R, Tindall A, James KD, Rand C. Results of a hydroxyapatite-coated (Furlong) total hip replacement: a 13- to 15-year follow-up. J Bone Joint Surg Br. 2005;87(8):1050–1054. doi: 10.1302/0301-620X.87B8.16011.
    1. Weber E, Sundberg M, Flivik G. Design modifications of the uncemented Furlong hip stem result in minor early subsidence but do not affect further stability: a randomized controlled RSA study with 5-year follow-up. Acta Orthop. 2014;85(6):556–561. doi: 10.3109/17453674.2014.958810.
    1. Cohen B, Rushton N. Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg Br. 1995;77(3):479–483. doi: 10.1302/0301-620X.77B3.7744941.
    1. Nilsdotter AK, Lohmander LS, Klassbo M, Roos EM. Hip disability and osteoarthritis outcome score (HOOS)–validity and responsiveness in total hip replacement. BMC Musculoskelet Disord. 2003;4:10. doi: 10.1186/1471-2474-4-10.
    1. Weber E, Flivik C, Sundberg M, Flivik G (2019) Migration pattern of a short uncemented stem with or without collar: a randomised RSA-study with 2 years follow-up. Hip Int:1120700019888471. 10.1177/1120700019888471
    1. Levadnyi I, Awrejcewicz J, Gubaua JE, Pereira JT. Numerical evaluation of bone remodelling and adaptation considering different hip prosthesis designs. Clin Biomech (Bristol, Avon) 2017;50:122–129. doi: 10.1016/j.clinbiomech.2017.10.015.
    1. Boden HS, Skoldenberg OG, Salemyr MO, Lundberg HJ, Adolphson PY. Continuous bone loss around a tapered uncemented femoral stem: a long-term evaluation with DEXA. Acta Orthop. 2006;77(6):877–885. doi: 10.1080/17453670610013169.
    1. Schilcher J, Ivarsson I, Perlbach R, Palm L. No difference in periprosthetic bone loss and fixation between a standard-length stem and a shorter version in cementless total hip arthroplasty. A randomized controlled trial. J Arthroplasty. 2017;32(4):1220–1226. doi: 10.1016/j.arth.2016.11.015.
    1. Freitag T, Hein MA, Wernerus D, Reichel H, Bieger R. Bone remodelling after femoral short stem implantation in total hip arthroplasty: 1-year results from a randomized DEXA study. Arch Orthop Trauma Surg. 2016;136(1):125–130. doi: 10.1007/s00402-015-2370-z.
    1. Jahnke A, Engl S, Altmeyer C, Jakubowitz E, Seeger JB, Rickert M, Ishaque BA. Changes of periprosthetic bone density after a cementless short hip stem: a clinical and radiological analysis. Int Orthop. 2014;38(10):2045–2050. doi: 10.1007/s00264-014-2370-6.
    1. Falez F, Papalia M, Granata G, Longo D, Ciompi A, Casella F, Mazzotta G, Favetti F. Bone remodelling and integration of two different types of short stem: a dual-energy X-ray - absorptiometry study. Int Orthop. 2020;44(5):839–846. doi: 10.1007/s00264-020-04545-6.
    1. Yan SG, Weber P, Steinbrück A, Hua X, Jansson V, Schmidutz FJIO. Periprosthetic bone remodelling of short-stem total hip arthroplasty: a systematic review. Int Orthop. 2018;42(9):2077–2086. doi: 10.1007/s00264-017-3691-z.
    1. Nam D, Salih R, Barrack RL, Nunley RM. An evaluation of proximal femur bone density in young, active patients undergoing total hip arthroplasty at one year postoperatively. Hip Int. 2019;29(1):51–57. doi: 10.1177/1120700018761152.
    1. Khosla S, Riggs BL. Pathophysiology of age-related bone loss and osteoporosis. Endocrinol Metab Clin NAm. 2005;34(4):1015–1030. doi: 10.1016/j.ecl.2005.07.009.
    1. Lamb JN, Baetz J, Messer-Hannemann P, Adekanmbi I, van Duren BH, Redmond A, West RM, Morlock MM, Pandit HG (2019) A calcar collar is protective against early periprosthetic femoral fracture around cementless femoral components in primary total hip arthroplasty: a registry study with biomechanical validation. Bone Joint J 101-B (7):779–786. 10.1302/0301-620X.101B7.BJJ-2018-1422.R1
    1. Brew CJ, Stockley I, Grainger AJ, Stone MH (2011) Iliopsoas tendonitis caused by overhang of a collared femoral prosthesis. J Arthroplasty 26 (3):504 e517–509. 10.1016/j.arth.2009.12.020
    1. Meding JB, Ritter MA, Keating EM, Faris PM. Comparison of collared and collarless femoral components in primary uncemented total hip arthroplasty. J Arthroplasty. 1997;12(3):273–280. doi: 10.1016/s0883-5403(97)90023-1.
    1. Albanese CV, Santori FS, Pavan L, Learmonth ID, Passariello R (2009) Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems: 37 patients followed for 3 years. Acta Orthop 80 (3):291–297. 10.3109/17453670903074467

Source: PubMed

3
Abonneren