Abstract

Objective: To compare a new device (Innocor) for non-invasive measurement of cardiac output (CO) by foreign gas rebreathing method with conventional techniques used in the measurements of cardiac function.

Methods: Cardiac outputs measured by Innocor (CO(RB)) were compared with CO obtained by echocardiography (CO(EC)), Swan-Ganz thermodilution (CO(TD)), and left ventricle radiography (CO(LVR)) in 34 patients subjected to cardiac catheterization. Values obtained from the four methods were analyzed by linear regression and paired values were compared by the method of Bland and Altman in SPSS.

Results: There was strong positive correlation (r=0.94) between Innocor cardiac output values and the corresponding values obtained by thermodilution and between CO(EC) and CO(LVR) values. Thermodilution appears to overestimate cardiac output when compared to the values obtained with Innocor by (0.66+/-0.22) L/min (P<0.0001). There was no correlation between data obtained by Innocor and the corresponding CO(EC) and CO(LVR) values.

Conclusion: Innocor CO(RB) is an easy, safe and well established method for non-invasive measurement of cardiac output with good prospects for clinical application in heart disease patients.

Figures

Fig. 1
Fig. 1
(a) Scatterplot of COTD against CORB; (b) Bland and Altman plot of COTD against CORB
Fig. 1
Fig. 1
(a) Scatterplot of COTD against CORB; (b) Bland and Altman plot of COTD against CORB
Fig. 2
Fig. 2
(a) Scatterplot of CORB against COEC; (b) Bland and Altman plot of CORB against COEC
Fig. 2
Fig. 2
(a) Scatterplot of CORB against COEC; (b) Bland and Altman plot of CORB against COEC
Fig. 3
Fig. 3
(a) Scatterplot of CORB against LVR; (b) Bland and Altman plot of CORB against LVR
Fig. 3
Fig. 3
(a) Scatterplot of CORB against LVR; (b) Bland and Altman plot of CORB against LVR
Fig. 4
Fig. 4
Scatterplot of COEC against Swan-Ganz
Fig. 5
Fig. 5
Scatterplot of COEC against LVR