Combined and hybrid marker models for radiostereometry assessment of polyethylene liner motion in dual mobility hip prosthesis: a proof-of-concept study

Peter Bo Jørgensen, Bart L Kaptein, Kjeld Søballe, Stig S Jakobsen, Maiken Stilling, Peter Bo Jørgensen, Bart L Kaptein, Kjeld Søballe, Stig S Jakobsen, Maiken Stilling

Abstract

Background: Investigation of polyethylene liner movement in total hip arthroplasty requires bead-marking for radiographic visibility of the liner. However, occlusion of markers poses a challenge for marker registration in radiographs.

Methods: The polyethylene of a dual mobility acetabular system was marked with twelve 1-mm tantalum markers (four groups of three markers) using a custom-made drill guide. Liner motion in a phantom and a patient was investigated with dynamic radiostereometry analysis (dRSA) at 1-year follow-up and static radiostereometry analysis (sRSA) postoperatively and at 1- and 2-year follow-up. A combined marker configuration (CMC) model was calculated from the registered positions of the liner markers and the femoral head in several images. Furthermore, the CMC model and the theoretic marker positions from computer-assisted models of the drill guide were combined in a hybrid model.

Results: The CMC model included eleven markers in the phantom and nine markers in the patient, which was sufficient for dRSA. Liner movement in the phantom followed liner contact with the femoral neck, while liner movement in the patient was independent. The hybrid model was necessary to determine liner orientation in sRSA recordings, which clearly changed from postoperative to 1- and 2-year follow-up even though the patient was positioned similarly.

Conclusion: Polyethylene liner motion in dual mobility hip prosthesis can be assessed with CMC models in dRSA recordings. In sRSA, the liner position between follow-ups is unpredictable and analysis requires inclusion of all markers in the model, accomplished with a hybrid marker model.

Trial registration: ClinicalTrials.gov [ NCT02301182 ], 25 October 2015.

Keywords: Hip prosthesis; Phantoms (imaging); Polyethylene; Radiography; Radiostereometric analysis.

Conflict of interest statement

None of the authors has conflicts of interest regarding this article.

© 2021. The Author(s) under exclusive licence to European Society of Radiology.

Figures

Fig. 1
Fig. 1
The occlusion of liner markers by the shell and the head/neck/stem (only the left radiostereometric analysis image frame is shown). Despite a high number of markers in the liner (n = 12) (a), they tend to be occluded by the head/neck (b) and cup (c) in RSA recordings. Marker information can be used as a simple marker model (green markers) (d), a combined marker configuration model (e) that merges marker information from several recordings (blue markers) or a hybrid model (f) that adds the theoretic marker positions (red markers) from the computer-assisted drawings of the drill guide used to insert the tantalum markers
Fig. 2
Fig. 2
A customised tool with a drill guide for inserting markers in the individual liner sizes of the system was developed (left) and machined in stainless steel (right). Three markers with increased depth ensured distinctive marker groups
Fig. 3
Fig. 3
Radiostereometric analysis set-up with a direct anteroposterior angle for static recordings (a) and a 45-degree recording angle (b) for dynamic recordings in order to obtain an optimal view of the polyethylene liner
Fig. 4
Fig. 4
Liner rotation and stem rotation of the phantom (a) and patient (b). Dashed lines show stem movements as indicated by the pictogram. In the phantom, liner movement (solid lines) occurs in the end range of modified FABER motion (at 4 and 12 s) and of modified FADIF motion (at 8 s) when the liner/neck angle (black) approaches 36.6° (solid red). In the patient, liner movement occurs in the end range of modified FABER motion (at 4 s) without the liner/neck angle approaching 36.6°
Fig. 5
Fig. 5
Example of anatomic dual mobility cup with a combined marker configuration model. The y-axis is shown for the femoral neck (blue), the cup (red) and the liner (green). ∠A indicates the liner/neck angle. The red circles indicate the detected marker projections in the image, as well as the centre of the femoral head
Fig. 6
Fig. 6
Radiographic anteversion and inclination of the cup. A Acetabular axis. A’ Projected acetabular axis. ∠RI indicates the radiographic inclination, defined as the angle between the longitudinal axis (Y) and the acetabular axis (norm of cup) projected perpendicular on the coronal plane (A’). ∠RA indicates the radiographic anteversion, defined as the angle between the acetabular axis (A) and the coronal plane. X, Y, and Z represent the coordinate system of the radiostereometric analysis recording

References

    1. DHR . National report 2019, The Danish Hip Arthroplasty Register, Regionernes Kliniske Kvalitetsudviklingsprogram. 2019.
    1. Blakeney WG, Epinette JA, Vendittoli PA. Dual mobility total hip arthroplasty: should everyone get one? EFORT Open Rev. 2019;4:541–547. doi: 10.1302/2058-5241.4.180045.
    1. D'Apuzzo MR, Koch CN, Esposito CI, et al. Assessment of damage on a dual mobility acetabular system. J Arthroplasty. 2016;31:1828–1835. doi: 10.1016/j.arth.2016.01.039.
    1. Grazioli A, Ek ET, Rudiger HA. Biomechanical concept and clinical outcome of dual mobility cups. Int Orthop. 2012;36:2411–2418. doi: 10.1007/s00264-012-1678-3.
    1. Hansen L, De Raedt S, Jorgensen PB, et al. Marker free model-based radiostereometric analysis for evaluation of hip joint kinematics: a validation study. Bone Joint Res. 2018;7:379–387. doi: 10.1302/2046-3758.76.BJR-2017-0268.R1.
    1. Digas G, Johansson PE, Karrholm J. Inducible displacements of the cup and the femoral head during active range of motion: dynamic RSA studies of cemented total hip replacements. J Orthop Res. 2013;31:1686–1693. doi: 10.1002/jor.22428.
    1. Zugner R, Tranberg R, Lisovskaja V, Shareghi B, Karrholm J. Validation of gait analysis with dynamic radiostereometric analysis (RSA) in patients operated with total hip arthroplasty. J Orthop Res. 2017;35:1515–1522. doi: 10.1002/jor.23415.
    1. Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JH. A new type of model-based Roentgen stereophotogrammetric analysis for solving the occluded marker problem. J Biomech. 2005;38:2330–2334. doi: 10.1016/j.jbiomech.2004.09.018.
    1. Baad-Hansen T, Kold S, Kaptein BL, Soballe K. High-precision measurements of cementless acetabular components using model-based RSA: an experimental study. Acta Orthop. 2007;78:463–469. doi: 10.1080/17453670710014095.
    1. Borlin N, Rohrl SM, Bragdon CR. RSA wear measurements with or without markers in total hip arthroplasty. J Biomech. 2006;39:1641–1650. doi: 10.1016/j.jbiomech.2005.05.004.
    1. Nebergall AK, Rader K, Palm H, Malchau H, Greene ME. Precision of radiostereometric analysis (RSA) of acetabular cup stability and polyethylene wear improved by adding tantalum beads to the liner. Acta Orthop. 2015;86:563–568. doi: 10.3109/17453674.2015.1054255.
    1. World Medical Association World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013;310:2191–2194. doi: 10.1001/jama.2013.281053.
    1. Mann KS, Kurudirek M, Sidhu GS. Verification of dosimetric materials to be used as tissue-substitutes in radiological diagnosis. Appl Radiat Isot. 2012;70:681–691. doi: 10.1016/j.apradiso.2011.12.008.
    1. Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JHC. A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech. 2003;36:873–882. doi: 10.1016/s0021-9290(03)00002-2.
    1. ISO . International ISO standard ISO 5725: accuracy (trueness and precision) of measurement methods and results. 2019.
    1. GaD VR, Fred L. Python 3 Reference Manual. Scotts Valley, CA: CreateSpace; 2009.
    1. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232. doi: 10.1302/0301-620X.75B2.8444942.
    1. Trozzi C, Kaptein BL, Garling EH, et al. Precision assessment of model-based RSA for a total knee prosthesis in a biplanar set-up. Knee. 2008;15:396–402. doi: 10.1016/j.knee.2008.05.001.
    1. Garling EH, Kaptein BL, Geleijns K, Nelissen RG, Valstar ER. Marker configuration model-based roentgen fluoroscopic analysis. J Biomech. 2005;38:893–901. doi: 10.1016/j.jbiomech.2004.04.026.
    1. Lam-Tin-Cheung K, Yuan X, Nikolov HN, et al. Marker-based technique for visualizing radiolucent implant components in radiographic imaging. J Orthop Res. 2017;35:2017–2022. doi: 10.1002/jor.23475.
    1. Ryd L, Yuan X, Lofgren H. Methods for determining the accuracy of radiostereometric analysis (RSA) Acta Orthop Scand. 2000;71:403–408. doi: 10.1080/000164700317393420.
    1. Valstar ER, Gill R, Ryd L, et al. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563–572. doi: 10.1080/17453670510041574.
    1. ISO . International standard ISO 16087:2013. Implants for surgery — Roentgen stereophotogrammetric analysis for the assessment of migration of orthopaedic implants. 2013.
    1. Zaribaf FP, Gill HS, Pegg EC. Characterisation of the physical, chemical and mechanical properties of a radiopaque polyethylene. J Biomater Appl. 2020;35:215–223. doi: 10.1177/0885328220922809.
    1. Horsager K, Kaptein BL, Jorgensen PB, Jepsen CF, Stilling M. Oxford medial unicompartmental knees display contact-loss during step-cycle motion and bicycle motion: a dynamic radiostereometric study. J Orthop Res. 2018;36:357–364. doi: 10.1002/jor.23625.
    1. Noyer D, Caton JH. Once upon a time.... Dual mobility: history. Int Orthop. 2017;41:611–618. doi: 10.1007/s00264-016-3361-6.

Source: PubMed

3
Abonnere