Accuracy and Reliability of Fetal Heart Rate and Uterine Contraction Monitoring Method
Study to Determine Substantial Equivalence of Electrohysterography (EHG) Monitoring to Tocodynanometer Monitoring
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
New York
-
New York, New York, United States, 11432
- Queens Hospital Center
-
-
Pennsylvania
-
Philadelphia, Pennsylvania, United States, 19140
- Temple University
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Pregnant, age 15-40
- Term gestation (>36 completed weeks)
- Singleton fetus
- Active labor
- Vertex presentation
- Requiring internal monitoring
Exclusion Criteria:
- Clinical contraindication for Intra Uterine Pressure Catheter
- Major fetal malformation
Study Plan
How is the study designed?
Design Details
- Primary Purpose: DIAGNOSTIC
- Allocation: RANDOMIZED
- Interventional Model: SINGLE_GROUP
- Masking: SINGLE
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
ACTIVE_COMPARATOR: Monitor Philips 50XM (K954351)
CTG Fetal Monitor If not confident of Monica AN24 displayed data then remove Monica AN24 monitor and continue monitoring with Philips 50XM
|
If not confident of Monica AN24 displayed data then remove Monica AN24 monitor and continue monitoring with Philips 50XM
|
|
EXPERIMENTAL: Monica AN24 (K101801)
EHG Fetal Monitor
|
If not confident of Monica AN24 displayed data then remove Monica AN24 monitor and continue monitoring with Philips 50XM
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The Mean Positive Percentage Agreement (PPA) for Fetal Heart Between Device 1 Monica AN24 & Device 2 Philips 50XM
Time Frame: during labor and delivery the waveforms were measured for between 35 minutes and 15 hours during the first and second stage of labour
|
During Labor & delivery, fetal heart rate was measured between the Monica AN24 & the Philips 50XM and the waveforms of the 2 devices were measured to see the percentage of time they were in agreement.
|
during labor and delivery the waveforms were measured for between 35 minutes and 15 hours during the first and second stage of labour
|
|
The Mean Positive Percentage Agreement (PPA) for Maternal Heart Between Device 1 Monica AN24 & Device 2 Philips 50XM
Time Frame: during labor and delivery the waveforms were measured for between 35 minutes and 15 hours during the first and second stage of labour
|
During Labor & delivery, maternal heart rate was measured between the Monica AN24 & the Philips 50XM and the waveforms of the 2 devices were measured to see the percentage of time they were in agreement.
|
during labor and delivery the waveforms were measured for between 35 minutes and 15 hours during the first and second stage of labour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The Mean Positive Percentage Agreement for Uterine Contractions Between the Monica AN24 & The Philips 50XM
Time Frame: between 35 mins & 15hrs during first & second stage labor
|
During labor and delivery uterine contractions were measured between the Monica AN24 & the philips 50XM, the waveforms of the two devices were measured to see the percentage of time they were in agreement
|
between 35 mins & 15hrs during first & second stage labor
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Wayne Cohen, MD, Queens Hospital Centre
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (ACTUAL)
Primary Completion
Study Completion (ACTUAL)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ESTIMATE)
First Posted
Study Record Updates
Last Update Posted (ESTIMATE)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- MHL101801
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.