User Performance Evaluation of an Investigational Blood Glucose Monitoring System
Ninja 2 User Performance Evaluation
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Indiana
-
Mishawaka, Indiana, United States, 46544
- Bayer HealthCare LLC, Diabetes Care
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Males and females, 18 years of age and above
- Diagnosed as having type 1 or type 2 diabetes
- Currently self-testing blood glucose at home
- Able to speak, read, and understand English
- Willing to complete all study procedures
Exclusion Criteria:
- Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
- Hemophilia or any other bleeding disorder
- Employee of competitive medical device company,immediate family member of such, or living within the household of such
- Pregnancy
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Intended Users of the System
Untrained subjects with diabetes (at least 70% of subjects will be insulin users) use an investigational blood glucose monitoring system (Ninja 2) to self-test capillary blood obtained from fingerstick and palm.
|
The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM).
BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer.
BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results <75mg/dL) or within +/- 20%(for reference BG results >=75mg/dL) of the reference method results (YSI capillary plasma).
|
1 hour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS).
BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer.
BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results <75mg/dL) or +/- 20% (for reference BG results >=75mg/dL) of the YSI capillary plasma reference method results.
|
1 hour
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- R&D-2010-2011.32
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.