Study of an Investigational Glucose Meter System
Performance of the G3 Blood Glucose Monitoring System With TATSU Strip
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
New Jersey
-
Fairfield, New Jersey, United States, 07004
- Consumer Product Testing Co.
-
-
New Mexico
-
Santa Fe, New Mexico, United States, 87505
- Southwest Clinical Research Center
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Males and females, 18 years of age and older
- Type 1 or type 2 diabetes
- Able to speak, read, and understand English
- Willing to complete all study procedures
Exclusion Criteria:
- Pregnancy
- Hemophilia or any other bleeding disorder
- Previously participated in a study using the G3 system
- Working for a medical laboratory, hospital or other clinical setting that involves training on and clinical use of blood glucose monitors.
- A condition, which in the opinion of the investigator, would put the person or study conduct at risk (reason for exclusion will be clearly documented by investigator or designee).
- Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Intended Users of the Monitoring System
Untrained subjects with diabetes use the G3 investigational blood glucose monitoring system.
|
Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and AST of the palm using the G3 meter and an investigational sensor.
Study staff test subject venous blood and all BG results are compared to a reference laboratory glucose method.
Untrained subjects utilize some additional features of the meter using the User Guide and provide feedback.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS).
BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer.
YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 5to15mg/dL (<100mg/dL YSI capillary plasma) or +/- 5to15% (>=100mg/dL YSI capillary plasma).
Site staff tested in parallel after subjects.
|
1 hour
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method
Time Frame: 1 hour
|
Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS).
BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer.
YSI capillary plasma BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) or +/- 20% (>=75mg/dL YSI capillary plasma).
|
1 hour
|
|
Percent of Venous Blood Glucose Results Within +/- 5to15mg/dL (<75mg/dL) or Within +/- 5to20% (>=75mg/dL) of Laboratory Glucose Method
Time Frame: 1 hour
|
Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS).
Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer.
YSI venous plasma results are used to calculate the number of BGMS results within +/- 5to15mg/dL (<75mg/dL YSI venous plasma) or +/- 5to20% (>=75mg/dL YSI venous plasma).
|
1 hour
|
|
Number of Subjects Able to Perform Given Tasks Using Product Labeling for Instruction
Time Frame: 1 hour
|
After reading the instructions for use, and without assistance from the study staff, subjects use the BGMS to utilize some of the additional features of the system.
Study staff documents Yes or No 'Did the subject complete the task successfully?'
|
1 hour
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Robert Bernstein, MD, FACE, Southwest Clinical Research Center
- Principal Investigator: Michael Caswell, PhD, Consumer Product Testing Co.
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- CTD-PRO-2010-009-01
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.