Efficacy Between Different Two Self-Expanding Nitinol Stents For The Atherosclerotic Femoro-Popliteal Arterial Disease (SENS-FP)
Efficacy of Self-Expanding Nitinol S.M.A.R.T-CONTROL Stent Versus Complete SE Stent For The Atherosclerotic Femoro-Popliteal Arterial Disease : Prospective, Multicenter, Randomized, Controlled Trial (SENS-FP Trial)
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Five randomized, controlled trials failed to demonstrate any benefit of a stainless-steel stent over angioplasty alone. The nitinol stent has proven superior primary patency than balloon angioplasty in superficial femoral artery (SFA) lesions. Several studies reported stent fractures were associated with a higher risk of in-stent restenosis and reocclusion. In vitro, Stefan et al. reported the 7 different SFA stents showed differences in the incidence of high strain zones, which indicates a potential for stent fracture, as demonstrated by the mechanical fatigue tests. They claimed differences in stent design might play a major role in the appearance of stent strut fracture related to restenosis and reocclusion. Also, in retrospective study, Iida et al. reported there was significant difference in stent fracture between S.M.A.R.T. stent group and Luminexx stent group and primary patency was worse in those associated with stent fracture than in those without stent fractures. Recent stent design improvements focus on decreasing stent fracture rates which can negatively impact patency rates by rearranging strut alignment.
The design of self-expandible nitinol stents might be different depending on the developed time; The first-generation nitinol stents (e.g., LuminexxTM and SmartTM) showed a remarkably high rate of stent strut fracture. A second generation of slotted tube nitinol stents has been developed. These stents had a better flexibility, by reducing the number of connections between cells or crowns, and by configurating spiral orientation of these interconnections. Several studies reported that these nitinol stents are more fracture-resistant and more flexible, some of them providing superior patency rate (e.g., LifeTM and EverflexTM). However, the one of the important limitations to their studies is that those was the non-randomized study of relative small sample size or was confined to in-vitro. Upto date, the multicenter, randomized controlled trial for direct comparison of stent fracture and primary patency between two different nitinol stents has not been done except one study; SMART versus Luminexx stent. SMART and Luminexx stent have been classified into 1st generation self-expandable nitinol stent. Complete-SE stent of Medtronic company was different to Smart stent of Cordis company in that the configuration of interconnection of Complete-SE stent had peak-to-peak connection and more spiral orientation of interconnection, compared to mart stent. On the other hand, Smart stent has the peak-to-valley bridge and in-line interconnection. Medtronics company have claimed Complete's stent crowns have been configured to minimize crown to crown interaction, increasing the stent's flexibility without compromising radial strength. We made the hypothesis that the design of Complete-SE stent might be more fracture-resistant or effective for in-stent restenosis, compared with Smart stent.
On the other hand, to date, in previous many reports, the dual antiplatelet therapy consisted of aspirin and clopidogrel have proven to decrease the incidence of cardiovascular death, myocardial infarction, or revascularization without an increase in major bleeding in patients who underwent percutaneous coronary intervention regardless of stent type (bare metal stent or drug-eluting stent). In 2011 ESC (European society of cardiology) guideline, dual antiplatelet therapy with aspirin and a thienopyridine for at least one month is recommended after infra-inguinal bare metal stent implantation (Class I, Level C). However, there have been no the definite evidence or guideline for the optimal antiplatelet agents after stent implantation one month later. There have been many studies for the efficacy of thienopyridine in peripheral arterial disease. Also, With the potential benefit of cilostazol on vascular function in vitro, there have several previous efforts to prove the efficacy of cilostazol in patients undergoing endovascular therapy or stent implantation in peripheral arterial disease. However, still specific data regarding a variety of antiplatelet regimen are limited. Also, very few trials have effectively nor properly addressed the direct comparison for the efficacy and safety between clopidogrel and cilostazol. This trial designed to evaluate the efficacy and safety between aspirin plus clopidogrel versus aspirin plus cilostazol in patients undergoing stent implantation in femoropopliteal lesions.
Study Type
Study Type
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Seoul, Korea, Republic of, 152-703
- Korea University Guro Hospital
-
Seoul, Korea, Republic of, 152-703
- Cardiovascular Center, Korea University Guro Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
Clinical criteria
- Age 20 years of older
- Symptomatic peripheral-artery disease with (Rutherford 2 - 6); moderate to severe claudication (Rutherford 2-3), chronic critical limb ischemia with pain while was at rest (Rutherford 4), or with ischemic ulcers (Rutherford 5-6)
- Patients with signed informed consent
Anatomical criteria
- Stenosis of >50% or occlusive atherosclerotic lesion of the ipsilateral femoropopliteal artery
- Patent (≤50% stenosis) ipsilateral iliac artery or concomitantly treatable ipsilateral iliac lesions (≤30% residual stenosis),
- At least one patent (less than 50% stenosed) tibioperoneal run-off vessel.
Exclusion Criteria:
- Disagree with written informed consent
- Major bleeding history within prior 2 months
- Known hypersensitivity or contraindication to any of the following medication: heparin, aspirin, clopidogrel, cilostazol, or contrast agent
- Acute limb ischemia
- Previous bypass surgery or stenting of the ipsilateral femoropopliteal artery
- Untreated inflow disease of the ipsilateral pelvic arteries (more than 50% stenosis or occlusion)
- Patients that major amputation ("above the ankle" amputation) has been done, is planned or required
- Patients with life expectancy <1 year due to comorbidity
Study Plan
How is the study designed?
Design Details
- Primary Purpose: TREATMENT
- Allocation: RANDOMIZED
- Interventional Model: PARALLEL
- Masking: NONE
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
EXPERIMENTAL: Complete SE Stent
study design is 2x2 randomization design.
First, before randomization, stratification will be performed according to lesion length 15cm criteria at web-based computerized program.
Patients will be randomized in a 1:1 manner according to different two (SMART versus Complete-SE) stents.
And then, patients received aspirin and clopidogrel during one month.
After one month from index procedure, patients were randomized to receive either clopidogrel group (clopidogrel will not be changed but continue) or cilostazol group (clopidogrel will be changed into cilostazol) in separate groups of SMART group and Complete SE group.
Randomization procedure will be performed using a web-based program
|
same to SMART CONTROL Stent
|
|
ACTIVE_COMPARATOR: SMART CONTROL Stent
same to Complete SE
|
Provisional stenting should be performed; the case that optimal ballooning response is not obtained (sub-optimal balloon response) should be enrolled.
The procedure is usually done, as follows; After the guidewire is passed through the target lesion, predilation of the target lesion with an optimally sized balloon will be performed prior to stent implantation.
Recommended, minimal balloon dilation time is 120 seconds.
The sub-optimal balloon response is defined as a residual pressure gradient of >15 mmHg, residual stenosis of >30%, and flow-limiting dissection.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The rate of binary restenosis
Time Frame: one year
|
the rate of binary restenosis (stenosis of at least 50 percent of the luminal diameter) or PSVR ≥ 2.5 or zero (PSVR=peak systolic velocity within the area of stenosis divided by peak systolic velocity in a normal adjacent proximal artery segment) in the treated segment at 12 months after intervention as determined by catheter angiography or Duplex ultrasound
|
one year
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
stent fracture rate, clinical outcomes, angiographic outcomes, ankle-brachial index
Time Frame: 1 year
|
|
1 year
|
Collaborators and Investigators
Sponsor
Sponsor
Publications and helpful links
Study record dates
Study Major Dates
Study Start (ANTICIPATED)
Study Start
Primary Completion (ANTICIPATED)
Primary Completion
Study Completion (ANTICIPATED)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ESTIMATE)
First Posted
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- SENS-FP
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.