Evaluate Safety And Effectiveness Of The Tivoli® DES and The Firebird2® DES For Treatment Coronary Revascularization
A Prospective, Open Label, Randomized Study to Evaluate Safety And Effectiveness Of The Tivoli® Biodegradable Polymer Rapamycin-Eluting Stent and The FIREBIRD2® Rapamycin-Eluting Coronary CoCr Stent For Treatment Coronary Revascularization
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Anticipated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Liaoning
-
Shenyang, Liaoning, China, 86-110016
- Recruiting
- The General Hospital Of Shenyang Military Region
-
Contact:
- Han Yaling, Prof.
- Phone Number: (86-024) 28851120
- Email: hanyaling@263.net
-
Principal Investigator:
- Han Yaling, Prof.
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
• The patient must be ≥18 of age;
- Symptomatic ischemic heart disease and/or objective evidence of myocardial ischemia including chronic stable coronary artery disease, or acute coronary syndrome including non-ST-elevation and ST-elevation myocardial infarction;
- Acceptable candidate for CABG;
- The patient is willing to comply with specified follow-up evaluations;4.Patients who agree to accept the follow-up visits.
- Patients can understand the study objectives psychologically and linguistically and show the sufficient compliance to the study protocol. Patients present acceptance of the risks and benefits described in the informed consent form.
Angiographic Inclusion criteria:
- At least one lesion with a diameter stenosis >70% or more suitable for coronary stent implantation in a vessel with a reference diameter ranging from 2.5 mm to 4.0 mm;
- Patients with multi-lesion or multi-vessel coronary disease, must have successful treatment (<30% diameter stenosis visual estimate) of the first treated lesion prior to treatment of other lesions at the same brand of stent. Staged (planned) procedures must be performed within 30 days of the index procedure.
Exclusion Criteria:
General Exclusion Criteria
- Pregnant or nursing patients and those who plan pregnancy in the period up to 1 year following index procedure.
- Patient has a history of bleeding diathesis or coagulopathy or patients in whom anti-platelet and/or anticoagulant therapy is contraindicated or in which patient will not be able to comply with dual antiplatelet therapy for at least 1 year;
- Patient has other medical illness (e.g., cancer, known malignancy, congestive heart failure, organ transplant recipient or candidate) or known history of substance abuse (alcohol, cocaine, heroin etc.) that may cause non-compliance with the protocol, confound the data interpretation or is associated with a limited life expectancy (i.e., less than 1 year);
- Left ventricular function <40%;
- Cardiogenic shock or hemodynamic compromise requiring pressors and/or inotropes or mechanical support;
- Patient has a known hypersensitivity or contraindication to aspirin, heparin, clopidogrel/ticlopidine, stainless steel alloy, cobalt chromium, Rapamycinry , styrene-butylenes-styrene or poly-lactic acid (PLA) polymer, and/or contrast sensitivity that cannot be adequately pre-medicated;
- Any significant medical condition which in the Investigator's opinion may interfere with the patient's optimal participation in the study;
- Currently participating in another investigational drug or device study or patient in inclusion in another investigational drug or device study during follow-up.
- Existing impairment in liver and kidney.
- Extremely tortuous and/or calcification lesions.
Angiographic exclusion criteria:
• Two or more chronic total occlusions in the proximal half of the epicardial coronary artery which cannot be recannalized.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Active Comparator: Firebird2® DES
Device:Firebird2® DES The Firebird2® rapamycin-eluting stent (Firebird2® DES) is an open cell balloon expandable cobalt chromium stent coated with a biocompatible durable styrene-butylenes-styrene (SBS) polymer containing rapamycin at a dose of 9 micrograms per millimeter of stent length.
|
Device:Tivoli® DES The Tivoli® DES drug eluting stent(Tivoli® DES) is an open cell balloon expandable cobalt chromium stent coated with a bio-gradable polymer (PLGA) containing rapamycin at a dose of 8 micrograms per millimeter of stent length.
In vitro assays have shown that about 50% of the drug releasing in 1 week and approximately 80% releasing in 28 days.
The polymer may be degraded in 3 to 6 months in human body.
This stent is supplied pre-mounted on a delivery catheter and is available in the following sizes: lengths of 10、15、18、21、25、30、35 mm and diameters of 2.5、2.75、3.0、3.5、4.0 mm.
The Firebird2® rapamycin-eluting stent (Firebird2® DES) is an open cell balloon expandable cobalt chromium stent coated with a biocompatible durable styrene-butylenes-styrene (SBS) polymer containing rapamycin at a dose of 9 micrograms per millimeter of stent length.
In vitro assays have shown that about 50% of the drug elutes in 1 week and approximately 90% elutes in 1 month.
This stent is supplied pre-mounted on a delivery catheter and is available in the following sizes: lengths of 13, 18, 23, 29, 33 mm and diameters of 2.25, 2.5, 2.75, 3.0, 3.5, 4.0 mm.
|
|
Experimental: Tivoli® DES
Device:Tivoli® DES The Tivoli® DES is an open cell balloon expandable cobalt chromium stent coated with a bio-gradable polymer (PLGA) containing rapamycin at a dose of 8 micrograms per millimeter of stent length.
|
Device:Tivoli® DES The Tivoli® DES drug eluting stent(Tivoli® DES) is an open cell balloon expandable cobalt chromium stent coated with a bio-gradable polymer (PLGA) containing rapamycin at a dose of 8 micrograms per millimeter of stent length.
In vitro assays have shown that about 50% of the drug releasing in 1 week and approximately 80% releasing in 28 days.
The polymer may be degraded in 3 to 6 months in human body.
This stent is supplied pre-mounted on a delivery catheter and is available in the following sizes: lengths of 10、15、18、21、25、30、35 mm and diameters of 2.5、2.75、3.0、3.5、4.0 mm.
The Firebird2® rapamycin-eluting stent (Firebird2® DES) is an open cell balloon expandable cobalt chromium stent coated with a biocompatible durable styrene-butylenes-styrene (SBS) polymer containing rapamycin at a dose of 9 micrograms per millimeter of stent length.
In vitro assays have shown that about 50% of the drug elutes in 1 week and approximately 90% elutes in 1 month.
This stent is supplied pre-mounted on a delivery catheter and is available in the following sizes: lengths of 13, 18, 23, 29, 33 mm and diameters of 2.25, 2.5, 2.75, 3.0, 3.5, 4.0 mm.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
Ischemia-driven Target Lesion Failure (TLF) which is a composite of cardiac death, myocardial infarction (Q and non-Q wave) and target lesion revascularization (TLR) at 12 months post-procedure.
Time Frame: 12 months
|
12 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Han Yaling, Prof., The General Hospital Of Shenyang Military Region
Publications and helpful links
General Publications
- Han Y, Xu B, Xu K, Guan C, Jing Q, Zheng Q, Li X, Zhao X, Wang H, Zhao X, Li X, Yu P, Zang H, Wang Z, Cao X, Zhang J, Pang W, Li J, Yang Y, Dangas GD. Six Versus 12 Months of Dual Antiplatelet Therapy After Implantation of Biodegradable Polymer Sirolimus-Eluting Stent: Randomized Substudy of the I-LOVE-IT 2 Trial. Circ Cardiovasc Interv. 2016 Feb;9(2):e003145. doi: 10.1161/CIRCINTERVENTIONS.115.003145.
- Han Y, Xu B, Jing Q, Lu S, Yang L, Xu K, Li Y, Li J, Guan C, Kirtane AJ, Yang Y; I-LOVE-IT 2 Investigators. A randomized comparison of novel biodegradable polymer- and durable polymer-coated cobalt-chromium sirolimus-eluting stents. JACC Cardiovasc Interv. 2014 Dec;7(12):1352-60. doi: 10.1016/j.jcin.2014.09.001. Epub 2014 Nov 12.
Study record dates
Study Major Dates
Study Start
Study Start
Primary Completion (Anticipated)
Primary Completion
Study Completion (Anticipated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- I-LOVE-IT2
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.