Non-Crossover Study of the LMA Protector Versus Ambu AuraGain
A Randomized Non-Crossover Study of the LMA ProtectorTM Versus Ambu Aura GainTM in Adult Patients
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Zürich, Switzerland, 8008
- Christian Keller MD, M.Sc.
-
Zürich, Switzerland, 8008
- Schulthess Klinik
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- ASA 1 or 2
- Age 19-65 yr
- Written informed consent
Exclusion Criteria:
- Difficult airway
- Non fasted
- BMI > 35
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: Single
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Active Comparator: LMA ProtectorTM
function tests: ease of insertion, oropharyngeal seal pressure, fiberoptic position, ease of ventilation, occult blood, gastric tube insertion
|
function tests: ease of insertion, oropharyngeal seal pressure, fiberoptic position, ease of ventilation, occult blood, gastric tube insertion
|
|
Active Comparator: Ambu AuraGainTM
function tests: ease of insertion, oropharyngeal seal pressure, fiberoptic position, ease of ventilation, occult blood, gastric tube insertion
|
function tests: ease of insertion, oropharyngeal seal pressure, fiberoptic position, ease of ventilation, occult blood, gastric tube insertion
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
oropharyngeal leak pressure
Time Frame: 5 min
|
Oropharyngeal leak pressure was determined by closing the expiratory valve of the anesthesia breathing system and a fixed gas flow of 3 l minute-1.
The airway pressure at which an equilibrium was reached was noted (maximum allowed 40 cm H2O).
|
5 min
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
anatomic position
Time Frame: 5 min
|
The anatomic position was assessed by passing a fiberoptic scope through the airway tube stopping 0,5 cm prior to the end of the tube.
The airway tube view was scored using an established scoring system.
|
5 min
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Berthold Moser, MD MBA, Schulthess Klinik
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimate)
First Posted
Study Record Updates
Last Update Posted (Estimate)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- Schulthess_Anä_8
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.