Comparison Between Enlite and Flash Glucose Monitoring
A Comparative Study on the Accuracy, Ease of Use and Skin Reactions Between the Enlite Sensor With Guardian 2 Link Transmitter and the FreeStyle Libre Flash Glucose Sensor
- The accuracy of the sensors (Flash vs Enlite glucose monitoring) will be evaluated with a standardized breakfast test. Patients receive a standardized breakfast. The usual insulin bolus will be administered at breakfast. After breakfast, a venous blood sample is taken for 2 hours every 15 minutes to determine the blood sugar. At the same time, the glucose value shown by the sensors will be noted. A comparison of the blood glucose value with the data recorded by the sensor can evaluate the accuracy of the sensor and also estimate the lag-time between the sensor glucose and the venous glucose.
- Patient satisfaction will be evaluated using a questionnaire that will be completed after the Enlite sensor has been worn for 1 month.
- The development of skin reactions will be checked by a short questionnaire, supplemented with a picture of possible skin phenomena.
- The data recorded by the FreeStyle sensor (average glucose,% above target,% within target,% under target, amount of hypoglycemia) in the month prior to sensor switching will be compared to the same data recorded by the Enlite sensor during the first month of use.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
-
Hasselt, Belgium, 3500
- Jessa Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Children (5 to 18 years)
- Diagnosis of type 1 diabetes mellitus
- Treatment is an insulin pump and wearing a FreeStyle Flash Libre glucosesensor
Exclusion Criteria:
-
Study Plan
How is the study designed?
Design Details
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Target group
Children (5 to 18 years) with type 1 diabetes mellitus treated with an insulin pump and wearing a FreeStyle Flash Libre glucosesensor will switch to the Enlite sensor communicating with Minimed 640G pump
|
Children (5 to 18 years) with type 1 diabetes mellitus treated with an insulin pump and wearing a FreeStyle Flash Libre glucosesensor will switch to the Enlite sensor communicating with Minimed 640G pump
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of Enlite sensor
Time Frame: 2 hours
|
The accuracy of the sensors (flash vs enlite glucose monitoring) will be evaluated with a standardized breakfast test.
Patients receive a standardized breakfast.
The usual insulin bolus will be administered at breakfast.
After breakfast, a venous blood sample is taken for 2 hours every 15 minutes to determine the blood sugar.
At the same time, the glucose value shown by the sensors will be noted.
A comparison of the blood glucose value with the data recorded by the sensor can evaluate the accuracy of the sensor and also estimate the lag-time between the sensor glucose and the venous glucose.
|
2 hours
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Usability of Enlite sensor
Time Frame: 1 month
|
Patient satisfaction will be evaluated using a questionnaire that will be completed after the Enlite sensor has been worn for 1 month.
|
1 month
|
|
skin reactions wih Enlite sensor
Time Frame: 1 month
|
The development of skin reactions will be checked by a short questionnaire, supplemented with a picture of possible skin phenomena.
|
1 month
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Guy Massa, MD, PhD, Jessa Hospital, Hasselt, Belgium
Study record dates
Study Major Dates
Study Start (ACTUAL)
Study Start
Primary Completion (ACTUAL)
Primary Completion
Study Completion (ACTUAL)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ACTUAL)
First Posted
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- JessaH_B243201732756
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Type1 Diabetes Mellitus
-
NCT03262415CompletedType 2 Diabetes Mellitus | Type1 Diabetes Mellitus
-
NCT03011021RecruitingType1 Diabetes Mellitus | Autoimmune Diabetes
-
NCT06606509Not yet recruitingType1 Diabetes Mellitus
-
NCT05418569RecruitingType1 Diabetes Mellitus
-
NCT03117998Completed
-
NCT04102202WithdrawnType1 Diabetes Mellitus
-
NCT03738982CompletedType1 Diabetes Mellitus
Clinical Trials on Enlite sensor communicating with Minimed 640G pump
-
NCT02657213Completed
-
NCT03103867CompletedDiabetes Mellitus, Type 1
-
NCT03653533Terminated
-
NCT05574062CompletedDiabetes type1 | Children, Only
-
NCT02246582CompletedType 1 Diabetes | Type 2 Diabetes
-
NCT01726621CompletedType 1 Diabetes | Type 2 Diabetes
-
NCT01991548CompletedType 1 Diabetes | Type 2 Diabetes
-
NCT02490098Withdrawn
-
NCT03815487CompletedDrug Therapy | Patient Care