Comparing Medical Body Composition Analyzer Seca mBCA 555 With Seca mBCA 515
Clinical Accuracy Validation of the Medical Body Composition Analyzer Seca mBCA 555 in Comparison to Seca mBCA 515
The mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations (BCA-01, BCA-02, BCA-20) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated based on the previous study BCA-03 conducted at the UKE.
The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.
In order to apply prediction equations and normal ranges from previous clinical investigations BCA-01, BCA-02, BCA-03 and BCA-20 to the new mBCA 555 it shall be compared to the mBCA 515.
Study Overview
Status
Status
Conditions
Conditions
Detailed Description
The use of bioelectrical impedance analysis (BIA) is widespread both in healthy subjects and patients. BIA allows the determination of the fat-free mass (FFM), skeletal muscle mass (SMM), fat mass (FM), extracellular water (ECW) and total body water (TBW) when using appropriate population, age or pathology-specific BIA equations and established procedures. BIA equations are based on validation studies where BIA is compared with reference methods such as Air Displacement Plethysmography (ADP), Dual-Energy X-Ray Absorptiometry (DXA), Magnetic Resonance Imaging (MRI) and isotope dilution.
The predicate device seca mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations in Kiel (Germany) and New York City (USA) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated in a previous study conducted at the University Hospital Hamburg-Eppendorf.
The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.
Primary objective is to test whether the mBCA 515 prediction equations can be applied to the new mBCA 555. If necessary, to proof substantial equivalence, prediction equations should be adjusted for the mBCA 555.
Secondary objective is to compare Impedance, phase angle, reactance and resistance for different frequencies in order to clarify if normal ranges for BIVA and phase angle generated with the mBCA 515 need to be adjusted for the new mBCA 555.
Bioimpedance was measured using the seca mBCA 555 and mBCA 515. Body height was measured using the seca 274.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
-
Hamburg, Germany, 20246
- University Medical Center Hamburg-Eppendorf - Institute of Transfusion Medicine
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- male and female subjects,
- 18 years and older,
- subjects need to be able to sign the informed consent and privacy policy
Exclusion Criteria:
- acute disease
- pregnancy
- intake of diuretics
- edema diagnosed by inspection and palpation of lower limbs
- paralysis e.g. after a stroke
- neurodegenerative diseases e.g. ALS
- tumors in treatment
- amputation
- electronic implants e.g. pacemaker
- probands who cannot provide an ICF by themselves
- probands who might be dependent from the sponsor or the investigation site
- current alcohol abuse
- active prostheses
- electronic life-support systems, e.g. artificial heart, artificial lung
- portable electronic medical devices, e.g. ECG devices or infusion pumps
- metallic implants
Study Plan
How is the study designed?
Design Details
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
SMM
Time Frame: 10min
|
Skeletal muscle mass, bias of mBCA 555 in comparison to mBCA 515
|
10min
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Phase Angle
Time Frame: 10 minutes
|
Phase angle at 50 kHz, bias of mBCA 555 in comparison to mBCA 515
|
10 minutes
|
Other Outcome Measures
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
R50RS
Time Frame: 5 minutes
|
resistance right body side, 50 kHz, bias of mBCA 555 in comparison to mBCA 515
|
5 minutes
|
|
R5RS
Time Frame: 5 minutes
|
resistance right body side, 5 kHz, bias of mBCA 555 in comparison to mBCA 515
|
5 minutes
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (ACTUAL)
Study Start
Primary Completion (ACTUAL)
Primary Completion
Study Completion (ACTUAL)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ACTUAL)
First Posted
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
Other Study ID Numbers
- BCA-21
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Bioelectrical Impedance Measurement of Healthy Adults
-
NCT01471938CompletedBioelectrical Impedance Measurement of Healthy Adults
-
NCT01368640CompletedBioelectrical Impedance Measurement of Healthy Adults
-
NCT03779932CompletedBioelectrical Impedance Measurement of Healthy Adults
-
NCT06845995CompletedOlder Adults | Body Composition Measurement | Cognitive Health | Metabolic Health | Bioelectrical Impedance | Appetite Hormones
-
NCT01681836CompletedFocus of Study of Healthy Adults
-
NCT06612164Completed
-
NCT02380599UnknownPhysiology of Manipulation / Mobilization in Healthy Adults
-
NCT03552666CompletedBioequivalence of Vitamin D in Healthy Adults
-
NCT02862236CompletedHealthy Adults | Subclinical Anxious Adults | Nootropic Potential of Hypericum and Alprazolam
-
NCT02003950CompletedInvestigate the Maladaptive Eating Behaviors of Normal Healthy Adults