Validation of Software for Assessment of Atrial Fibrillation From Data Acquired by a Wearable Smartwatch
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Locations
-
-
-
Kallithea, Greece
- Onassis Cardiac Surgery Center
-
-
-
-
District of Columbia
-
Washington, District of Columbia, United States, 20010
- MedStar Washington Hospital Center
-
-
Florida
-
Orlando, Florida, United States, 32806
- Bioclinica Research Orlando
-
The Villages, Florida, United States, 32162
- Bioclinica Research The Villages
-
-
Illinois
-
Winfield, Illinois, United States, 60190
- PMG Research of DuPage
-
-
Iowa
-
Ames, Iowa, United States, 50010
- PMG Research of McFarland
-
-
North Carolina
-
Rocky Mount, North Carolina, United States, 27804
- PMG Research of Rocky Mount
-
Statesville, North Carolina, United States, 28625
- PMG Research of Piedmont Healthcare
-
Wilmington, North Carolina, United States, 28401
- PMG Research of Wilmington
-
-
South Carolina
-
Mount Pleasant, South Carolina, United States, 29407
- PMG Research of Charleston
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Women and men 22 years of age or older
- Capable of giving informed consent
- Sufficient manual dexterity to capture an ECG from the Fitbit wrist-worn device.
- Subjects in the atrial fibrillation (AF) cohort must have a known history of AF and must be in -AF at the time of enrollment based on the screening ECG
Exclusion Criteria:
- Anti-arrhythmic (rhythm control) medication use (such as amiodarone or flecainide)
- Cardiac pacemaker or implantable cardioverter-defibrillator
- History of a life-threatening rhythm such as ventricular tachycardia or ventricular fibrillation
- Subjects in the sinus rhythm (SR) cohort should not have any history of atrial fibrillation and should not be in atrial fibrillation or other atrial arrhythmia at the time of enrollment based on the screening ECG
Study Plan
How is the study designed?
Design Details
- Observational Models: Case-Control
- Time Perspectives: Prospective
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Group 1: SR
Group 1: Subjects in the sinus rhythm (SR) cohort should not have any history of atrial fibrillation and should not be in atrial fibrillation or other atrial arrhythmia at the time of enrollment based on the screening ECG.
|
Diagnostic Test: Simultaneous assessment of heart rhythm by 12-lead ECG and wearable device characterization of heart rhythm.
|
|
Group 2: AF
Group 2: Subjects in the atrial fibrillation (AF) cohort must have a known history of AF and must be in AF at the time of enrollment based on the screening ECG.
|
Diagnostic Test: Simultaneous assessment of heart rhythm by 12-lead ECG and wearable device characterization of heart rhythm.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Evaluate the software algorithm's ability to classify a heart rhythm as sinus rhythm or atrial fibrillation from a waveform equivalent to a Lead I ECG.
Time Frame: Through study completion, an average of 4 months
|
Sensitivity and specificity of the software algorithm in classifying sinus rhythm and atrial fibrillation from the equivalent of a Lead I ECG waveform, compared with physician's interpretation of a standard 12-lead ECG.
Performance goals are 90% (sensitivity) and 92% (specificity)
|
Through study completion, an average of 4 months
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Confirm the software's ability to produce a waveform clinically equivalent to a Lead I ECG from a clinical reference device (12-lead ECG).
Time Frame: Through study completion, an average of 4 months
|
Qualitative assessment of the software produced waveform to reference Lead I from 12-lead ECG. Goal is >0.80 of strips appear to overlay to the unaided eye. Comparison of R-wave amplitudes between the Fitbit waveform and the clinical reference. Goal is >0.80 of R-wave amplitudes to be within 2mm. |
Through study completion, an average of 4 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Venkatesh Raman, MD, Medstar Health Research Institute
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 129-0360-01
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.