ExoAtlet II For SCI Patients (EXOATLET2SCI)
ExoAtlet II Lower Extremity Exoskeleton Safety and Efficacy Use With Spinal Cord Injured Individuals
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Moscow, Russian Federation, 129110
- Moscow Regional Research Clinical Institute. M.F. Vladimirsky
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- 18 Years of age or older in general good health
- Weight, no more than 220lbs (100kg)
- Skin must be healthy where it touches the ExoAtlet 2.0
- Able to stand using a device such as a standing frame
- Have enough strength in your hands and shoulders to support yourself standing and walking using crutches or a walker
- Have good control of upper body
- Determined to have enough bone health to walk full weight bearing without risk of fracture. Meeting of this condition is at the discretion of your personal MD
- Passive range of motion (PROM) at shoulders, trunk, upper extremities and lower extremities within functional limits for safe gait and use of appropriate assistive device/stability aid
- Hip width no greater than 18" (46 cm) measured when sitting.
- Femur length between 37 cm and 49 cm measured between mid patellar tendon and the floor.
- In general, good health and able to tolerate moderate levels of activity.
- Blood pressure and heart rate within established guidelines for locomotive training: At rest; Systolic 150 or less Diastolic 90 or less and Heart rate 100 or less Exercise; Systolic 180 or less, Diastolic 105 or less and Heart Rate 145 or less
Exclusion Criteria:
- Spinal cord injury level higher than T4 or C7 ASIA D
- Severe muscle stiffness/tightness
- Significant spasticity (Modified Ashworth Scale score of 3 or above)
- Trunk or lower extremity pressure ulcer
- Open Wounds/tissue hypersensitivity
- Unstable spine, un-healed limbs, or fractures
- Severe sensitivity to touch
- Presence of bone in soft tissue where bone normally does not exist (heterotopic ossification), limiting range of motion in the hip or knee joints
- Joint instability, dislocation, moderate to severe hip dysplasia
- Significant scoliosis (>40 degrees)
- Hardware, implant, or any external device impeding with safe fitting or use of ExoAtlet 2.0
- Femoral or tibial rotation deformity (>15 degrees)
- Significant flexion contractures limited to 35º at the hip and 20º at the knee
- Uncontrolled seizures, musculoskeletal injury, fracture or lower-limb surgery in past year.
- Known history of pulmonary disease limiting exercise tolerance or history of cardiac disease
- Psychopathology, or other condition that the physician or investigator, in his or her clinical judgment, considers to be exclusionary to safely use an exoskeleton.
- Severe concurrent medical diseases, illness, systemic or peripheral infection
- Dizziness or headache with standing
- History of autonomic dysreflexia
- Orthostatic Hypotension: Decrease in Systolic BP > 20, Diastolic BP > 10 upon standing from a seated position.
- Pregnant or lactating females
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: ExoAtlet II Safety and Efficacy
20 participants with paraplegia from SCI will participate in gait training using ExoAtlet II powered exoskeleton.
|
Subjects will learn to use the ExoAtlet II exoskeleton, with the assistance of a trained medical professional, to perform functional activities (including stand-up, sit-down, stepping forward, and walking).
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Number of adverse events
Time Frame: 0-8 weeks
|
Document and track any adverse events, both minor and major, during exoskeleton sessions.
|
0-8 weeks
|
|
Skin integrity
Time Frame: 0-8 weeks
|
A visual inspection of the participant's skin, at all areas of contact with the exoskeleton, will be conducted at the beginning and end of each session.
|
0-8 weeks
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
10 Meter Walk Test
Time Frame: 1 Day
|
Assesses walking speed, in the exoskeleton, in meters per second over a distance of 10 meters
|
1 Day
|
|
6 Minute Walk Test
Time Frame: 1 Day
|
Record the distance covered during a 6 minute walk test utilizing the exoskeleton, to evaluate users endurance
|
1 Day
|
|
Transfer in and out
Time Frame: 2 Days
|
Percentage of patients able to successfully transfer in and out of the exoskeleton with or without assistance
|
2 Days
|
|
Perform Sit to Stand to Sit
Time Frame: 2 Days
|
Percentage of patients that successfully completed both sit to stand and stand to sit in the exoskeleton
|
2 Days
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Sergey Kotov, MD, Moscow Regional Research and Clinical Institute Moniki n.a. M.F. Vladimirskiy
Publications and helpful links
General Publications
- Miller LE, Zimmermann AK, Herbert WG. Clinical effectiveness and safety of powered exoskeleton-assisted walking in patients with spinal cord injury: systematic review with meta-analysis. Med Devices (Auckl). 2016 Mar 22;9:455-66. doi: 10.2147/MDER.S103102. eCollection 2016.
- Kozlowski AJ, Bryce TN, Dijkers MP. Time and Effort Required by Persons with Spinal Cord Injury to Learn to Use a Powered Exoskeleton for Assisted Walking. Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):110-21. doi: 10.1310/sci2102-110. Epub 2015 Apr 12.
- Gagnon DH, Cunha JD, Boyer-Delestre M, Bosquet L, Duclos C. How does wearable robotic exoskeleton affect overground walking performance measured with the 10-m and six-minute walk tests after a basic locomotor training in healthy individuals? Gait Posture. 2017 Oct;58:340-345. doi: 10.1016/j.gaitpost.2017.08.027. Epub 2017 Aug 26.
- Tefertiller C, Hays K, Jones J, Jayaraman A, Hartigan C, Bushnik T, Forrest GF. Initial Outcomes from a Multicenter Study Utilizing the Indego Powered Exoskeleton in Spinal Cord Injury. Top Spinal Cord Inj Rehabil. 2018 Winter;24(1):78-85. doi: 10.1310/sci17-00014. Epub 2017 Nov 20.
- Hartigan C, Kandilakis C, Dalley S, Clausen M, Wilson E, Morrison S, Etheridge S, Farris R. Mobility Outcomes Following Five Training Sessions with a Powered Exoskeleton. Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):93-9. doi: 10.1310/sci2102-93. Epub 2015 Apr 12.
- McIntosh K, Charbonneau R, Bensaada Y, Bhatiya U, Ho C. The Safety and Feasibility of Exoskeletal-Assisted Walking in Acute Rehabilitation After Spinal Cord Injury. Arch Phys Med Rehabil. 2020 Jan;101(1):113-120. doi: 10.1016/j.apmr.2019.09.005. Epub 2019 Sep 27.
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- EA22019SCI
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.