Comparison of the Performance of ICM and CIED in Detecting AF (ID-AF)
Comparison of the Performance of Implantable Cardiac Monitors and Cardiac Implantable Electronic Devices in Detecting Atrial Fibrillation
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Atrial fibrillation (AF) is the most common abnormal heart rhythm disturbance, affecting 1-2 million people in the UK. The irregular heartbeat caused by AF can make the heart pump blood less efficiently. As a result, AF significantly increases the risk of having a stroke, heart failure and dementia. However, a significantly proportion of people have no symptoms, and they may be only found to be in AF after having a stroke. Therefore, diagnosis largely relies on accurate electrocardiogram (ECG) monitoring.
AF episodes can be unpredictable and easily missed unless continuous ECG monitoring is undertaken. Pacemakers can continuously record the heart's electrical activity and detect AF with a high degree of confidence. However, they are invasive and rely on electrical wires implanted inside the heart, and hence less suited to be used as primary monitoring devices. To overcome these difficulties, implantable cardiac monitors (ICM) have been designed to be placed under the skin.
The new generation of ICMs can now be injected under the skin. Moreover, they connect with the patient's smartphone and transmit recordings. Despite these significant improvements, their ability to reliably capture AF has never been tested against pacemakers.
The aim of this project is to study the performance of the two commonly used ICMs in detecting AF episodes and explore how the new connectivity can empower patients and improve patient care. The investigators plan to inject an ICM in 30 patients with AF and pre-existing pacemaker. After a period of 6 months, the investigators will compare how many episodes were detected in each device. With the information collected the investigators will try to understand the pitfalls in the current technology and develop strategies to improve it.
Accurate, minimally invasive long-term ECG monitor can have far reaching benefits for patients, both in routine clinical practice and research.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Andre Briosa e Gala, MD
- Phone Number: +441865220255
- Email: andre.gala@ouh.nhs.uk
Study Locations
-
-
-
Oxford, United Kingdom
- John Radcliffe Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Participant is willing and able to give informed consent for participation in the trial.
- Male or Female.
- Aged 18 years or above.
- History of paroxysmal and persistent AF.
- Dual-chamber pacemaker, Implantable cardioverter defibrillator (ICD) or cardiac resynchronisation therapy (CRT) device with a functioning atrial lead, able to record electrograms and remote monitoring.
Exclusion Criteria:
- Diagnosis of permanent AF.
- Contra-indications for implantable cardiac monitor.
- Unable to comply with the follow-up schedule.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Other: Patients implanted with the Abbott Confirm Rx ICM
Patient with cardiac implantable electronic devices with an atrial lead will have Confirm Rx Implantable Cardiac Monitor injected in the anterior chest wall.
|
Confirm Rx implanted in patients with atrial fibrillation detected on their cardiac implantable electronic device (CIED)
|
|
Other: Patients implanted with the Medtronic LINQ ICM
Patient with cardiac implantable electronic devices with an atrial lead will have LINQ Implantable Cardiac Monitor injected in the anterior chest wall.
|
Reveal LINQ implanted in patients with atrial fibrillation detected on their cardiac implantable electronic device (CIED)
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Assessment of Performance of Confirm Rx ICM and Reveal LINQ ICM in Detecting AF Episodes in Comparison to CIED
Time Frame: During follow-up (6 months)
|
To assess the performance of the Confirm Rx™ ICM and Reveal LINQ™ ICM in detecting clinically significant AF episodes (episodes longer than 6 minutes) in comparison to CIED (gold standard).
Confirm Rx and Reveal LINQ™ ICM sensitivity, specificity, positive predictive value and negative predictive value will be calculated for AF episodes longer than 6 minutes.
|
During follow-up (6 months)
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Comparison of Changes in R-wave Amplitude Following Implantation in Both ICMs.
Time Frame: During follow-up (6 months)
|
A comparison will be made between the percentage of change in R-wave amplitudes between both ICMs.
|
During follow-up (6 months)
|
|
To Compare the Transmission Success Rate of Both Devices.
Time Frame: During follow-up (6 months)
|
The percentage of successful transmission in both ICMs will be calculated.
|
During follow-up (6 months)
|
|
To Compare the Number of Patient-activated Recordings That Contains Symptoms.
Time Frame: During follow-up (6 months)
|
The percentage of patient-activated recordings with symptoms attributed to those episodes will be calculated.
|
During follow-up (6 months)
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Tim Betts, MD FRCP, Oxford University Hospitals NHS Trust
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- 14558
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.