Comparison Between Different Investigations Used in Intraocular Lens Calculation in High Myopic Cataractous Patients
Optical Biometry Versus Ultrasonic Biometry in Intraocular Lens Calculation in High Myopic Cataractous Patients: A Comparative Study
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
- Prospective interventional (QUASI experimental) comparative study
- This study will be conducted on high myopic cataractous patients scheduled for phacoemulsification and IOL implantation in department of ophthalmology, Sohag university
- patients will be divided into 2 groups one will be subjected to intraocular lens (IOL) calculation by optical biometry the second one will be subjected to intraocular lens (IOL) calculation by ultrasonic biometry
- postoperative auto refraction after I month
Study Type
Study Type
Enrollment (Anticipated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Esraa A Okasha, Resident
- Phone Number: 01028909988
- Email: esraa011037@med.sohag.edu.eg
Study Contact Backup
- Name: Usama A Mohamed, Professor
- Phone Number: 01005470455
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- patients with simple cataract not associated with other pathologies suitable for phacoemulsification and 1ry IOL implantation With axial length (AXL) equal to or greater than 26.5mm
Exclusion Criteria:
- History of trauma
- Associated pathologies such as optic neuropathy, age related macular degeneration, macular edema, retinal detachment, , ocular inflammation, retinitis pigmentosa, proliferative diabetic retinopathy
- corneal opacities or irregularities, scars, dystrophy or ectasia
- Patients who underwent previous corneal surgery (including refractive surgery)
- Patients having intraoperative complications as inability to achieve secure 'in the bag' placement of the IOL (i.e. due to posterior capsule rupture, vitreous loss, weak zonules, or zonular rupture)
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: Non-Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: optical biometry
|
optical biometry is a fast, noncontact method it uses the method of partial coherence interferometry (PCI) to measure the axial length (AXL), based on reflection of the interference signal of the retinal pigment epithelium.
|
|
Experimental: ultrasonic biometry
|
ultrasonic biometry is contact method depends on ultrasonic transducer producing thin sound beam travelling through different media of the eye when it faces interface of substance dissimilar from that it is travelling through part is reflecting and the other travel through the different substance
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
desired post operative refraction
Time Frame: postoperative refraction 1 month
|
detection of postoperative refraction by the use of autorefractor (one of the objective methods of assessment of refraction) after use of IOL power calculated by different investigations (optical vs ultrasonic)
|
postoperative refraction 1 month
|
Collaborators and Investigators
Sponsor
Sponsor
Publications and helpful links
General Publications
- Tappeiner C, Rohrer K, Frueh BE, Waelti R, Goldblum D. Clinical comparison of biometry using the non-contact optical low coherence reflectometer (Lenstar LS 900) and contact ultrasound biometer (Tomey AL-3000) in cataract eyes. Br J Ophthalmol. 2010 May;94(5):666-7. doi: 10.1136/bjo.2009.167700. No abstract available.
- Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011 Nov;37(11):2018-27. doi: 10.1016/j.jcrs.2011.05.042.
- Olsen T. Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol Scand. 2007 Feb;85(1):84-7. doi: 10.1111/j.1600-0420.2006.00774.x.
Study record dates
Study Major Dates
Study Start (Anticipated)
Study Start
Primary Completion (Anticipated)
Primary Completion
Study Completion (Anticipated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- Soh-Med-21-06-01
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.