A Performance Evaluation of the LumiraDx Point of Care CRP Assay (REACT)
A Performance Evaluation of the LumiraDx Point of Care CRP Assay.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This is a performance evaluation study designed to assess the precision and accuracy of the LumiraDx point of Care (POC) C-reactive Protein (CRP) test when used in patients presenting with symptoms of infection, tissue injury or inflammatory disorder.
The LumiraDx POC CRP test is a quantitative immunoassay providing results in under 10 minutes. The accuracy of the LumiraDx POC CRP test will be assessed using capillary whole blood, venous blood, and plasma samples, by comparison to the CRP results obtained from the same individuals as analysed by trained laboratory professionals using the reference device, Siemens Dimension Xpand Plus CRP assay. The results of this study are intended to be used for regulatory filings for use as an in vitro diagnostic test in the UK, European Union (EU) and other relevant geographies.
Adults (aged 18 and over at time of consent) presenting to the study sites with symptoms of infection, tissue injury or inflammatory disorder will be included in the study. Approximately 250 patients will be recruited to ensure that sufficient suitable patients are enrolled and to allow for a small level of exclusions.
After obtaining informed consent, blood tube samples will be drawn and fingerstick samples of capillary blood will be taken from the subject, these will be applied directly onto unique test strips for immediate measurement of CRP on the LumiraDx POC instrument.
The CRP results obtained via the LumiraDx CRP test will be used in a set of comparative analyses designed to determine the performance characteristics of the test. This study is an observational, cross-sectional design. As an observational study of an investigative device, results obtained during the course of the study will not inform or alter patient management or treatment decisions in any way.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Edinburgh, United Kingdom
- Royal Infirmary Edinburgh
-
London, United Kingdom
- University College London Hospital
-
London, United Kingdom
- St Georges Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
- Patients 18 years and over at the time of consent
- Willing and able to provide written informed consent and comply with study procedures
- Presenting with symptoms of infection, tissue injury, inflammatory disorders. This includes but not limited to : respiratory tract infection (upper and lower), rheumatoid arthritis, Lupus, burns, trauma, inflammatory bowel disease.
Exclusion Criteria:
- The patient is currently receiving or has received within the past thirty (30) days of the study visit an experimental biologic or drug including either treatment or therapy.
- Skin lesions or conditions that would preclude a fingerstick and/or a venous blood draw.
- Patient has previously participated in this study
- Any patient with critical illness or requiring a time critical intervention
- Patient with end of life or palliative care
- Patient suffering from the following conditions: Myeloma, monoclonal gammopathy, extreme lipaemia
- Patient is deemed medically unfit to participate.
Study Plan
How is the study designed?
Design Details
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Symptoms of inflammation, Infection, tissue injury
Venous blood draw of up to 24mL and up to 6 capillary fingersticks
|
Venous blood draw
Capillary blood draw
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
To determine the performance of the LumiraDx POC CRP assay when compared to the Siemens Dimension Xpand Plus CRP in patients with symptoms of infection, inflammation, or injury.
Time Frame: 2 months
|
Measurement of blood samples from patients with symptoms of inflammation, infection or injury in a reference method and in the LumiraDx method to assess accuracy of the LumiraDx method
|
2 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Alasdair Gray, MD, NHS Lothian
- Principal Investigator: James Harnett, University College London Hospital NHS Foundation Trust
- Principal Investigator: Phil Moss, St Georges University Hospitals NHS Foundation Trust
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- S-CLIN-PROT-00008
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.