Validation of Vital Signs Recording with VT-Patch Connected Devices in Children (VT)
Validation of Vital Signs Recording with VT-Patch® Connected Devices in Children
: Patient readmission to PICU is a major point of concern for physicians because, although being a rare event, it has been strongly associated with more morbidity and death. Patient monitoring once they have left the PICU is drastically modified, and small connected devices could be an option to early identify patients at risk of PICU readmission. Such devices have been manufactured for the recording of vital signs in adults, but they do not exist for children.
Besides, studies have recently shown that devices that use optical sensors may be less accurate in individuals with dark skin pigmentation since dark skin tones impose a limitation on optical biometric sensing. The wearable and medical device industry has not sufficiently addressed this issue yet. This is one of the main challenges that need to be addressed in this area to make sure the connected devices using this technology will work with all different skin colours including very dark skin.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Main objective: Assess the accuracy of VT-Patch® to measure oxygen saturation (SpO2) in children with diverse skin colour.
Secondary objectives: Assess the accuracy of VT-Patch® to measure 3 additional physiologic parameters (heart rate (HR), respiratory rate (RR), and temperature. Exploratory assessment of blood pressure (BP) measurement, ECG signal, and movement capture.
Design: Prospective validation study of VT-Patch® for the recording of 4 physiologic parameters and exploratory assessment of three additional parameters.
Methods: The study will be performed in two successive phases:
Run-in phase: Testing VT-Patch® on four children to assess the feasibility of the protocol.
Main study: validation phase of the devices on 12 children per age group (2 children per skin colour group) to assess its accuracy (VT-Patch® for less than 2 years old). The patients of the run-in phase will be included in the main study if no modification of the protocol is required. 5 physiologic parameters (HR, SpO2, RR, temperature, and ECG) will be prospectively and simultaneously recorded for 8 hours with VT-Patch® and the regular monitoring system implemented in the unit High-resolution database that will be the gold standard for evaluation. Vital signs recorded with the connected device will not be used for the clinical care of patients. The Skin colour of each patient will be assessed by the colorimetric Monk scale and as shown below.
Study Type
Study Type
Enrollment (Estimated)
Enrollment
Contacts and Locations
Study Contact
Study Contact
- Name: Azadeh Dastmalchi, PhD
- Phone Number: +1(613)2761413
- Email: azadeh.dastmalchi@vitaltracer.com
Study Locations
-
-
Quebec
-
Montreal, Quebec, Canada, H3T 1C5
- Recruiting
- CHU Sainte Justin Hospital
-
Contact:
- Azadeh Dastmalchi, PhD
- Phone Number: +1(613)2761413
- Email: azadeh.datmalchi@vitaltracer.com
-
Contact:
- Philippe Jouvet, MD, PhD
- Phone Number: (514) 345-4931
- Email: philippe.jouvet.med@ssss.gouv.qc.ca
-
Contact:
- Philippe Jouvet, MD, PhD
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Sampling Method
Study Population
In order to achieve at least 80% match between physiological data's recorded with VT-Patch®, or VT-Watch® and with the standard of care (tolerated difference: 2% for SpO2), this study will include 10 patients per age group (see below) including 4 patients per age group with invasive arterial pressure monitoring (arterial line).
- Neonates: 0 - 28 days
- Infants:29 days - 2 years
- Children: 2 - 12 years
- Adolescents:12 - 18 years
Description
Inclusion Criteria:
- PICU monitoring for the next 24 hours
Exclusion Criteria:
- Recent cardiac or thoracic surgery
- Thoracic skin lesion that contraindicates the VT patch
- No parental consent
- Intermittent presence of one study observer in the patient room considered inappropriate by the physician or the nurse in charge because of the child's medical condition
Study Plan
How is the study designed?
Design Details
- Observational Models: Case-Only
- Time Perspectives: Retrospective
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of VitalTracer devices Oxygen Saturation (SPO2) measurement for different skin colour
Time Frame: 8 hours per subject
|
Percentage of the difference between the absolute value of SpO2 recorded by VT-Patch and the one recorded in high resolution database.
|
8 hours per subject
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of VitalTracer devices blood pressure, heart rate, skin temperature, and Electrocardiogram (ECG) measurement
Time Frame: 8 hours per subject
|
Percentage of the difference between the absolute value of Blood Pressure, Heart Rate and skin temperature and the quality of ECG signals recorded by the VT-Patch and the one recorded in high resolution database.
|
8 hours per subject
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Estimated)
Primary Completion
Study Completion (Estimated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- VT001
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.