An Invasive Controlled Desaturation Study to Determine the Accuracy of the CART-I Pulse Oximeter
An Invasive Controlled Desaturation Study to Determine the Accuracy of the CART-I Pulse Oximeter (Sky Labs Inc., Korea) in Healthy Adult Volunteers
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Pangyo
-
Gyeonggi-do, Pangyo, Korea, Republic of
- Sky Labs
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Healthy adults aged 20-50 years2
- COHb < 3 %
- MetHb < 2 %
- ctHb > 10 g/dl
- Non-smoker
- An individual deemed fit by the investigators to be included in the clinical trial
Exclusion Criteria:
- History of respiratory disease
- History of cardiovascular disease
- Smoker
- Pregnant
- History of fainting
- History of diabetes
- Obesity (defined as BMI > 30 kg/m2)
- Allen's test positive (poor collateral blood supply of hand)
- An individual who is judged to be difficult for the collection of normal PPG signals because CART-I is loose when being worn on fingers other than the thumb and little finger
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: CART-I plus
CART-I pulse oximeter with a ring-type wearable PPG sensor was placed on each volunteer to evaluate the SpO2 accuracy during steady-state, non-motion conditions.
|
CART-I noninvasively measures the oxygen saturation of blood in a local area by screening the bloodstream through the user's finger using PPG signals
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of SpO2
Time Frame: 1 day
|
Mean bias, standard deviation, standard error, maximum/minimum bias, precision are calcuated
|
1 day
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Byung-Moon Choi, MD, PhD, Asan Medical Center
Publications and helpful links
General Publications
- Agarwala P, Salzman SH. Six-Minute Walk Test: Clinical Role, Technique, Coding, and Reimbursement. Chest. 2020 Mar;157(3):603-611. doi: 10.1016/j.chest.2019.10.014. Epub 2019 Nov 2.
- Luks AM, Swenson ER. Pulse Oximetry for Monitoring Patients with COVID-19 at Home. Potential Pitfalls and Practical Guidance. Ann Am Thorac Soc. 2020 Sep;17(9):1040-1046. doi: 10.1513/AnnalsATS.202005-418FR.
- Andres-Blanco AM, Alvarez D, Crespo A, Arroyo CA, Cerezo-Hernandez A, Gutierrez-Tobal GC, Hornero R, Del Campo F. Assessment of automated analysis of portable oximetry as a screening test for moderate-to-severe sleep apnea in patients with chronic obstructive pulmonary disease. PLoS One. 2017 Nov 27;12(11):e0188094. doi: 10.1371/journal.pone.0188094. eCollection 2017.
- Shiao SY, Ou CN. Validation of oxygen saturation monitoring in neonates. Am J Crit Care. 2007 Mar;16(2):168-78.
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
Other Study ID Numbers
- 2021-0153
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.