Comparison Between the Octaray and Pentaray Catheters in Patients Undergoing Ablation for Atrial Tachycardia. (OCTOPUS-AT)
Comparison Between the Multi-electrode Octaray and Pentaray Catheters in Patients Undergoing Ablation for Atrial Tachycardia Using Coherent Mapping.
The goal of this randomized clinical trial is to compare the 3D electroanatomical activation map-quality and -efficiency in patients undergoing atrial tachycardia (AT) catheter ablation. Primary objective of the study is to contrast CARTO-Coherent mapping quality (electrograms per map) using Octaray versus Pentaray catheter for left and right atrium ATs, separately.
Participants will be randomized with 1:1 ratio for undergoing 3D anatomy map with either Octaray (intervention group) or Pentaray (control) group. Next, all patients will undergo activation map of the AT with both catheters. For ablation in the intervention group, the Ocatray made activation map of the AT will be used, and the Pentary made AT activation map will be used in the control group.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This is a randomized open-label cross-over trial including consecutive patients undergoing ablation for atrial tachycardia of the left and right atrium. The sample-sizes of the left (N=20) and right (N=20) atrium will be equal. Half of the eligible patients with either left or right atrium AT will be randomized to undergo Coherent Mapping with Pentaray catheter first followed by Coherent Mapping with the Octaray catheter. The other half will undergo initial Coherent Mapping with the Octaray catheter followed by Pentaray catheter. This way all study participants (total N=40) will undergo the Coherent Mapping with both catheters (Figure 1). Only the second map created with either Octaray (n active arm left atrium =10 and n active arm right atrium =10) or Pentaray (n control arm left atrium =10 and n control arm right atrium =10) will be used for the subsequent ablation procedure.
The main outcome of the trial is the performance of the Coherent Mapping quality (electrograms per map) with Octaray catheter versus Pentaray catheter for left and right atrium, separately. A map will be considered complete when the fill threshold reach 5 mm for the entire surface area of the left or right atrium, that is the interpolation between points will be <5 mm as previously reported. The high- and low-pass filters settings will be harmonized between the catheters. The following data will be collected: 1) number of electrograms per map; 2) Voltage based signal-to-noise ratio (vSNR),17 calculated as the ratio between the mean of the peak-to-peak amplitude of the true local bipolar atrial signals and the mean of the SD of the peak-to-peak of all other signals considered as far-field or ambient lab noise; 3) electrograms acquisition rate (i.e., the acquired number of points per minute of mapping); 4) electrograms acquisition density (i.e., the acquired number of electrograms per cm2 of the left or right atrium); 5) number of electrograms acquired per beat; 6) the total ablation and fluoroscopy times; 7) acute mapping success rate, defined as termination of AT at the target region; 8) number of entrainments attempted and the best post-pacing interval value at the ablation site when available. Successful entrainment is sometimes not possible (i.e., focal AT) and some investigator may prefer not to entrain. Therefore, the decision to use entrainment and the number of entrainment attempts will be left to the discretion of the operator.
Study Type
Study Type
Enrollment (Anticipated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Bakhtawar Mahmoodi, MD PhD MPH
- Phone Number: +31611828388
- Email: b.mahmoodi@erasmusmc.nl
Study Contact Backup
- Name: Tamas Szili-Torok, MD PhD
- Email: t.szilitorok@erasmusmc.nl
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Age ≥18 years
- Planned for catheter ablation of either left- or right-atrium AT/flutter will be eligible for inclusion. In the group of right atrium AT patients with typical flutters will also be eligible for inclusion.
Exclusion Criteria:
- Individuals not able to provide informed consent for any reason including language and mental disabilities
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Randomized
- Interventional Model: Crossover Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Octaray (intervention)
Octaray created 3D map quality
|
Newly designed multi-electrode catheter for creating 3D electroanatomical CARTO maps.
|
|
Active Comparator: Pentaray (control)
Pentaray created 3D map quality
|
Current standard multi-electrode catheter for creating 3D electroanatomical CARTO maps.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
Number of electrograms per map
Time Frame: during procedure
|
during procedure
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Voltage based signal-to-noise ratio.
Time Frame: during procedure
|
during procedure
|
|
|
electrograms acquisition rate
Time Frame: during procedure
|
the acquired number of points per minute of mapping)
|
during procedure
|
|
electrogram acquisition density
Time Frame: during procedure
|
the acquired number of electrograms per cm2 of the left or right atrium
|
during procedure
|
|
number of electrograms acquired per beat
Time Frame: during procedure
|
during procedure
|
|
|
the total ablation and fluoroscopy times
Time Frame: during procedure
|
during procedure
|
|
|
acute mapping success rate
Time Frame: during procedure
|
defined as termination of AT at the target region
|
during procedure
|
|
number of entrainments attempted and the best post-pacing interval value at the ablation site
Time Frame: during procedure
|
May not be available especially in focal AT's
|
during procedure
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Anticipated)
Study Start
Primary Completion (Anticipated)
Primary Completion
Study Completion (Anticipated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- MEC-2022-0269
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.