Validation Study of Mean Arterial Pressure (MAP) Parameter of Masimo INVSENSOR00073
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
California
-
Irvine, California, United States, 92618
- Masimo Corporation
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
Influenced by study design:
- Subject is 18 to 40 years of age.
- Subject weighs a minimum of 110 lbs.
- Baseline heart rate ≥ 45 bpm and ≤ 90 bpm.
- Baseline blood pressure:
- Systolic blood pressure ≤ 135 mmHg and ≥ 100 mmHg.
- Diastolic blood pressure ≤ 95 mmHg and ≥ 55 mmHg.
- Valid systolic blood pressure (SBP) auscultatory measurements for lateral difference is ≤ 15 mmHg.
- Valid diastolic blood pressure (DBP) auscultatory measurements for lateral difference is ≤ 10 mmHg.
- Hemoglobin value ≥ 11 g/dL.
- CO value ≤ 3.0% FCOHb.
- Subject is comfortable to read and communicate in English*. * This is to ensure the subject can provide informed consent (as study materials are currently available in English only) and can comply with study procedures.
Exclusion Criteria:
Influenced by device warning labels:
- Subject is pregnant or breastfeeding.
- Subject is experiencing dysrhythmia or arrhythmia.
Influenced by study design/environment:
- Subject is concurrently participating in another research study.
- Subjects not suitable for the investigation at the discretion of the clinical team including but not limited to the items below.
- Subjects who do not pass the health assessment for safe participation in the study procedures.
- Inability to insert or difficulty with inserting an intravenous line in the subject's hand or arm and/or an arterial line in the radial artery of the subject's wrist.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Test Subject
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo INVSENSOR00073 sensor.
|
Noninvasive blood pressure monitor
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Mean Arterial Pressure (MAP) Accuracy and Precision of Masimo INVSENSOR00073
Time Frame: 1-5 hours.
|
Accuracy and precision of mean arterial pressure (MAP) measured by Masimo INVSENSOR00073 in comparison to radial arterial line.
|
1-5 hours.
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- CIP-1091
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.