SCT02 Clinical Accuracy Validation (WIBOFE)
Clinical Accuracy Validation for the Withings BeamO (SCT02) Smart Thermometer
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Florida
-
Miami, Florida, United States, 33175
- Diverse Clinical Research
-
-
Indiana
-
Indianapolis, Indiana, United States, 46237
- Dragonfly Primary Care
-
-
New York
-
New York, New York, United States, 10029
- New York City Health and Hospitals Corporation
-
-
Ohio
-
Cincinnati, Ohio, United States, 45206
- Integrated Family Care
-
Mason, Ohio, United States, 45040
- SES
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria
- Consent: Demonstrates understanding of the study and willingness to participate as evidenced by voluntary written informed consent and has received a signed and dated copy of the informed consent form.
- Age: Participant age is among the categories
- Compliance: Understands and is willing, able and likely to comply with all study procedures and restrictions.
Exclusion Criteria
- Personnel: An employee of the sponsor or the study site or members of their immediate family.
Medical Condition:
- Preterm infants (current age plus length of gestation is less than 40 weeks)
- Has taken antipyretic in previous 2 hours
- Inflammation at the measurement site
- Medical conditions that affect temperature readings (barbiturates, thyroid, antipsychotics, recent immunizations)
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Basic Science
- Allocation: Non-Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Febrile - Investigational Device
Temperature recording of patient with fever with investigational device
|
temperature measurement with investigational device (Withings SCT02)
|
|
Active Comparator: Febrile - Comparator
Temperature recording of patient with fever with predicate device
|
temperature measurement with reference device (Withings SCT01)
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Clinical Bias
Time Frame: 1 visit
|
Mean of the per-subject differences (Investigational - Reference), computed using the first measurement obtained from each subject with each device.
A bias closer to 0 °F indicates closer agreement between devices; the sign indicates direction (a negative value means the investigational device read lower, on average, than the reference).
|
1 visit
|
|
Limits of Agreement
Time Frame: 1 visit
|
Defined per ISO (International Standards Organization) 80601-2-56 as 2 × the standard deviation of the per-subject (Investigational - Reference) differences.
A smaller value indicates tighter agreement between the two devices around the mean bias.
|
1 visit
|
|
Clinical Repeatability
Time Frame: 1 visit
|
Pooled standard deviation of three repeated measurements taken on the same subject with the investigational device, calculated as σr = √[(σ1² + σ2² + … + σN²) / N], where σi is the within-subject standard deviation of the three repeats for subject i and N is the number of subjects with repeated measurements.
A smaller value (closer to 0 °F) indicates greater consistency of repeated measurements on the same subject.
|
1 visit
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- WIBOFE
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.