Pulse Oximetry Performance Comparison in Newborns
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Contact
Study Contact
- Name: Anna Klochak
- Phone Number: 3032633702
- Email: anna.l.klochak@medtronic.com
Study Contact Backup
- Name: Ami Stuart
- Email: ami.stuart@medtronic.com
Study Locations
-
-
California
-
Los Angeles, California, United States, 90048
- Cedars-Sinai
-
-
Missouri
-
St Louis, Missouri, United States, 63130
- Washington University in St. Louis
-
-
North Carolina
-
Winston-Salem, North Carolina, United States, 27157
- Atrium Health Wake Forest Baptist
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
- Adult
- Older Adult
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
- Parent or legal guardian is willing and able to provide informed consent prior to any study procedures being performed.
- Subject is expected to be able to participate for the duration of the evaluation in the delivery room. In most instances, the length of the study will not exceed the anticipated duration of ECG monitoring.
- Subject is ≥ 29 weeks gestational age.
- Subject can accommodate multiple sensors.
Exclusion Criteria:
- Subjects that have an existing health condition, in which the investigator determines that safe or accurate oximetry measures may not be obtained at the areas where the PO sensor could be placed. For example, major congenital anomalies, limb anomalies or anyone with abrasions or surface lesions on the contact areas would be excluded.
- Any subject whose treatment or condition would be compromised if they were to participate in this study and any patient whose use of a pulse oximeter and sensor would present a potential for serious risk to health, safety, or welfare.
- Currently participating in any other study expected to confound study results in the opinion of the investigator.
Study Plan
How is the study designed?
Design Details
Number of groups / cohorts
Cohorts and Interventions
Group / CohortGroup / Cohort |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Newborn
A variety of subjects will be targeted for inclusion representing differing stages of development and environments
|
The Nellcor™ N-600x Pulse Oximetry System with N-600X Pulse Oximeter and Nellcor™ Sensors and Cables with OxiMax technology is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin and PR.
The N-600x Pulse Oximeter is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments.
The Masimo Rad-97™ and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin, PR, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or acoustic respiratory rate (RRa).
The Masimo Rad-97™ and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
In addition, the Masimo Rad-97™ and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97™ and accessories for functional oxygen saturation of arterial hemoglobin and PR to multi-parameter devices for the display on those devices.
Market-released ECG monitoring system
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
To compare time to stable signal of heart rate (HR) estimation between two pulse oximeters (PO).
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
|
To compare overall accuracy of heart rate (HR) estimation between two pulse oximeters (PO) as compared to gold standard electrocardiography (ECG).
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
To evaluate instances of false bradycardia.
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
|
To evaluate instances of false bradycardia requiring emergency medical treatment.
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
|
To evaluate variability in PR between the two PO sensors
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
|
To compare the number of repositions and replacements required for the two PO sensors during monitoring.
Time Frame: Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Subject participation is expected to last a minimum of 5 minutes and a maximum of 12 hours.
|
Collaborators and Investigators
Sponsor
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- MDT24005PROLD
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.