Physician-Sponsored IDE for the Talent Endoluminal Stent Graft System for the Treatment of Thoracic Lesions

September 10, 2023 updated by: Rodney A. White, M.D.
The primary objective of this trial is to determine whether the Talent Thoracic Stent Graft, an investigational device, is a safe and effective method of treating thoracic aortic aneurysms (abnormal ballooning of the vessel wall) and other thoracic lesions (dissections, transections, pseudoaneurysms, penetrating ulcers, etc.). The endovascular method is a substitute for the major operation that is performed to treat the lesions.

Study Overview

Detailed Description

The endovascular method consists of the insertion of catheters (vinyl tubes) into both groin arteries. These catheters then allow positioning of artificial tubes or stent graft(s) into the diseased blood vessel without the need for major surgery which involves opening the chest. The blood vessel problem (thoracic aortic aneurysm/lesion) will be treated using an artificial bypass tube (stent graft) that will be placed inside the diseased artery. The stent graft(s) that will be used are made of polyester graft fabric sewn to a self-expanding nickel-titanium (Nitinol) wire frame and is manufactured by Medtronic AVE, Santa Rosa California.

Aneurysms/lesions can be located in any position along the length of the aorta. This study evaluates only aneurysms/lesions that occur in the descending part of the thoracic aorta as it traverses the chest and enters the abdominal cavity. The risk of thoracic aneurysms/lesions is that they rupture without warning. The risk of rupture increases as the aneurysm size increases. Following rupture, almost all patients expire within the first 24 hours. For this reason treatment of thoracic aneurysms/lesions is recommended by conventional surgical means if the patient is a candidate for an operation. The conventional operation involves occluding the aorta and replacing the aneurysm with a cloth tube that is sewn to replace the diseased part of the aorta. The major surgery and occlusion of the aorta that is part of the conventional surgery is directly related to several complications that have been reported in the literature. In addition, many patients are too ill for conventional surgery due to concomitant illnesses.

After the procedure, the patient will be followed at regular intervals (pre-discharge, 1 month, 6 months, 12 months and every year thereafter for life) as part of the evaluation of the experimental treatment. During this time, the patient will have several tests performed to evaluate the function of the repair. The tests will include clinical examination, x-rays of the chest and spiral CT scan. The experimental part is the placement of the TALENT Stent-Graft System using the catheter (endovascular) methods. The risk of the experimental catheter bypasses is that the procedure may not be successful. In most cases, this would require a standard operation to repair the problem. Sometimes there are other risks such as injury to the vessels. Small pieces of diseased arteries may be dislodged which may require removal. Bleeding may occur from introduction sites, and rarely, infection may develop. If a catheter bypass is unsuccessful, there can be added discomfort because the procedure may last longer.

The benefits of the procedure are that the patient might avoid some of of the pain and discomfort associated with standard operations. Major surgery is avoided and hospital stay may be much shorter. Patients may be able to return to usual daily activities sooner. In addition, patients who are too high risk for conventional surgical treatment can have their aneurysm treated using the endovascular graft if their anatomy is appropriate

Study Type

Interventional

Enrollment (Actual)

304

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • California
      • Torrance, California, United States, 90502
        • LAC Harbor-UCLA Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

Subjects who participate in this study as study patients must fulfill the following criteria:

  • Subject is > 18 years of age.
  • Subject is not pregnant or lactating. Females of child-bearing potential must practice a reliable method of contraception.
  • Subject is diagnosed with one of the following conditions of the descending thoracic aorta. All conditions must be verified by diagnostic imaging [ultrasonography, computed tomography (CT), magnetic resonance imaging (MRI) or angiography].

    • A true (i.e., atherosclerotic) supraceliac aneurysm (fusiform or saccular type) with or without a co-existing aortic dissection or penetrating aortic ulcer
    • Aortic dissection of DeBakey Type I or II (Stanford A, proximal) in the absence of an aneurysm; OR
    • Penetrating aortic ulcer in the absence of an aneurysm; OR
    • Traumatic transection; OR
    • Pseudoaneurysm - traumatic or degenerative (i.e., one that does not involve all layers of the vessel and is not atherosclerotic in origin).
  • Subject's anatomy is suitable for placement of the Medtronic/Talent Stent-Graft, with a distinct proximal aneurysm neck of 10 mm or more in length and a distal aneurysm neck of at least 10 mm.
  • Subject has a TAA that is dilated to > 5 cm in diameter, > 1.5 times the diameter of the adjacent native/non-aneurysmal aorta, or is symptomatic.
  • Subject has a proximal and distal aortic neck diameter > 18 mm and < 42 mm.
  • Subject has an arterial access site, either peripherally or via infrarenal abdominal aorta that is adequate for introduction of the stent-graft delivery system.
  • Subject is competent to give informed consent.
  • Subject will be available for the periodic follow-up (surveillance) after the procedure.

Exclusion Criteria:

  • Subjects who would participate as study subjects and who fulfill any of the following criteria may not participate in this study:
  • Subject has TAA with less than 10 mm proximal fixation length.
  • Subject has an aneurysm that would require exclusion by the stent-graft of the segment of the aorta that gives rise to dominant spinal cord/intercostal arteries.
  • Subject has a lesion that prevents delivery or expansion of the device.
  • Subject has systemic infection, or is suspected of having systemic infection.
  • Subject has a known mycotic aneurysm.
  • Subject is not available or is not willing to come back for periodic follow-up (surveillance) after the procedure.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Treatment of Thoracic Lesions
Endoluminal treatment of thoracic lesions
Endoluminal treament of thoracic lesions with a thoracic stent-graft

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Participants With Successful Endoluminal Treatment of Thoracic Lesion
Time Frame: Throughout study completion, an average of 1 year
Evaluate the number of patients who had successful treatment of their thoracic lesion using the Medtronic Talent Endoluminal Stent Graft..
Throughout study completion, an average of 1 year

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Participants With Successful Device Delivery and Deployment
Time Frame: Throughout study completion, an average of 1 year
Evaluate the number of patients where the thoracic device was successfully delivered and deployed within the patients' thoracic lesion.
Throughout study completion, an average of 1 year

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Rodney A. White, M.D., Lundquist Institute for Biomedical Innovation at Harbor-UCLA Medical Center

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

May 1, 2002

Primary Completion (Actual)

January 1, 2021

Study Completion (Actual)

January 1, 2021

Study Registration Dates

First Submitted

December 28, 2007

First Submitted That Met QC Criteria

January 17, 2008

First Posted (Estimated)

January 18, 2008

Study Record Updates

Last Update Posted (Actual)

September 14, 2023

Last Update Submitted That Met QC Criteria

September 10, 2023

Last Verified

June 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Subscribe