Alice PDx User/Validation Extended Trial

October 16, 2018 updated by: Philips Respironics

The objective of this study is to test the usability of the Alice PDx device by both the end user (patient/participant) and the sleep technician. Data will be collected via questionnaires regarding the participant's ability to understand the participant user manual under home conditions and the technician's ability to set up the Alice PDx according to the health care provider manual. This study will also compare the physiological data recorded by the Alice PDx to the physiological data recorded by its predicate device the Alice 5 System and validate its equivalence. This will be evaluated by comparing the manual study scores for detected apnea and hypopnea events and on an event by event basis during a standard polysomnography (PSG)/ sleep study. The study will test the ability of the Alice PDx to record good quality data as determined by the Good Study Indicator (GSI) function of the Alice PDx firmware.

The primary hypothesis is that the participant will be able to successfully apply the Alice PDx basic leads following the participant user manual instructions and the sleep technician will be able to successfully apply the advanced leads in preparation for the sleep study. In addition, it will be demonstrated that the Alice PDx is equivalent to the Alice 5 System in its intent to record, display and print physiological data to the clinician/physician in order to aid in the screening/follow-up diagnosis of respiratory sleep disorders.

Study Overview

Detailed Description

Alice PDx is a portable, diagnostic recording device. It may be used for obstructive sleep screening as well as follow-up and diagnostic assessment. The device may be used in a sleep lab or clinical setting by trained professionals and it may be used at home by patients as directed by their health care provider. The Alice PDx device is not currently approved by the Food and Drug Administration (FDA).

The Alice PDx device is capable of recording various physiological inputs and storing the data locally on a removable storage card. The device may also be connected directly to a PC (personal computer) running the Alice Sleepware software application. Sleepware can display live or pre-recorded data in a resolution consistent with the computer hardware specifications.

Study Type

Interventional

Enrollment (Actual)

32

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Pennsylvania
      • Indiana, Pennsylvania, United States, 15701
        • Indiana Regional Medical Center Sleep Disorders Center
      • Monroeville, Pennsylvania, United States, 15146
        • Sleep Center of Greater Pittsburgh

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Patients must call into a sleep lab or be referred to a sleep lab for a diagnostic polysomnogram
  • Able to follow directions
  • Able to provide informed consent

Exclusion Criteria:

  • Participants requiring supplemental oxygen therapy
  • History of having a previous sleep study/polysomnogram performed
  • Current positive airway pressure (PAP) users
  • Unable or unwilling to perform a polysomnogram

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: Alice PDx with only written instructions
Participants will be asked to follow the Alice PDx user instructions to apply basic leads and sensors and undergo a sleep study in their home. Participants will receive little or no instruction concerning the set up of the Alice PDx device.
Participants will be asked to follow the Alice PDx user instructions to apply basic leads and sensors and undergo a sleep study in their home. Participants will receive little or no instruction concerning the set up of the Alice PDx device. After the Alice PDx home study has proven successful, participants will return to the sleep center to undergo a second sleep study, the Lab Night, in which simultaneous monitoring with the Alice PDx and Alice 5 will be performed.
Other Names:
  • PDx
Other: Alice PDx with written and verbal instructions
Participants will be asked to follow the Alice PDx user instructions to apply basic leads and sensors and undergo a sleep study in their home. Participants will receive little or no instruction concerning the set up of the Alice PDx device.
Participants will be provided instruction by a trained sleep professional on the set up of the Alice PDx device. They will then be asked to follow the instructions for the Alice PDx device, if necessary, and undergo a sleep study in their home. After the Alice PDx home study has proven successful, participants will return to the sleep center to undergo a second sleep study, the Lab Night, in which simultaneous monitoring with the Alice PDx and Alice 5 will be performed.
Other Names:
  • PDx

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
This Study Will Compare the Apnea Hypopnea Index Recorded by the Alice PDx to the Physiological Data Recorded by Its Predicate Device the Alice 5 System and Validate Its Equivalence.
Time Frame: Lab Night
The Apnea-Hypopnea Index (AHI) is an index used to indicate the severity of sleep apnea. It is represented by the number of apnea and hypopnea events per hour of sleep. The AHI detected events of Alice 5 system was compared to the AHI events of the Alice PDx.
Lab Night
This Study Will Compare the Central Apnea Index, Hypopnea Index, Mixed Apnea Index and Obstructive Apnea Index Recorded by the Alice PDx to the Physiological Data Recorded by Its Predicate Device the Alice 5 System and Validate Its Equivalence.
Time Frame: Lab Night

The central apnea index, hypopnea index, mixed apnea index and obstructive apnea indexes are all values that are calculated by determining the type of apneic event (apnea, hypopnea, mixed or obstructive). Each type of event is added up over the night and divided by the number of hours.

For this analysis each index was compared between both devices.

Lab Night
This Study Will Compare Different Measurements Recorded by the Alice PDx to the Measurement Data Recorded by Its Predicate Device the Alice 5 System and Validate Its Equivalence.
Time Frame: Lab Night
This study will compare different measurements recorded which include Time in Bed, Stage N1, Stage N2, Stage N3, REM, Sleep Onset Latency, Total Sleep Time, Wake Time in Bed and Wake After Sleep Onset by the Alice PDx to the measurement data recorded by its predicate device the Alice 5 System and validate its equivalence.
Lab Night

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
The Secondary Objective of This Study is to Compare the Measurements of the Alice PDx When Patients Complete the Set-up of the Device at Home and When Patients Are Set up by a Sleep Technician in the Sleep Laboratory.
Time Frame: Lab Night
The Alice PDx incorporates a unique Good Study Indicator (GSI). The GSI is a predicated on airflow and oximeter signal quality and displays the amount of "good quality data" needed for a study to be complete and valid. The GSI visually displays the amount of good quality data in 25-percent increments on the Alice PDx screen. For purposes of this secondary objective this number was compared from participants who set-up and wore the device at home and those that wore the device in a sleep lab set up by a sleep technician.
Lab Night

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Imran Bajwa, MD, Indiana Regional Medical Center Sleep Disorders Center

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

September 1, 2008

Primary Completion (Actual)

June 1, 2009

Study Completion (Actual)

September 1, 2009

Study Registration Dates

First Submitted

February 11, 2009

First Submitted That Met QC Criteria

February 11, 2009

First Posted (Estimate)

February 12, 2009

Study Record Updates

Last Update Posted (Actual)

November 19, 2018

Last Update Submitted That Met QC Criteria

October 16, 2018

Last Verified

October 1, 2018

More Information

Terms related to this study

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Subscribe