Contactless Measurement of Heart Rate, Heart Rate Variability and Breathing Rate Using Remote Photoplethysmography

Contactless Measurement of Heart Rate, Heart Rate Variability and Breathing Rate Using Face Video Analysis (Remote Photoplethysmography) - Validation of Shen.AI Vitals Software.

The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability and breathing rate measurement with the use of Shen.AI Vitals software developed by MX Labs.

Study Overview

Detailed Description

Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).

The measurements will be taken with various camera-equipped devices (smartphones, tablets, laptops) using either front camera (self-measurement) or rear camera (measurement made by someone else), in two different lighting conditions (artificial light or natural daylight), with various levels of device stability (fully stable or held in hand/s), and in two different body positions (sitting or supine). In total, each participant will have 8 measurements taken.

Standard deviation of normal heartbeat intervals (SDNN) will be used as the main HRV index.

Apart from the average HR (averaged over the whole 1-minute measurement), instantaneous values of HR (averaged over 10 s) will also be determined and displayed every 1 s by the Shen.AI VItals software.

In all cases, reference measurements will be taken simultaneously (chest ECG for the calculation of average and instantaneous HR as well as HRV; impedance pneumography for the determination of BR).

The reference measurements will be taken with Finapres Nova with the ECG+Resp module (Finapres Medical Systems B.V.).

For each measured parameter, the results from both methods will be compared using the Bland-Altman analysis or non-parametric limits of agreement.

Study Type

Interventional

Enrollment (Actual)

136

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Locations

    • Dolnośląskie
      • Wrocław, Dolnośląskie, Poland, 50-556
        • Uniwersyteckie Centrum Wsparcia Badań Klinicznych we Wrocławiu

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • age ≥ 18 years
  • ability to read and write in Polish or English
  • ability to operate a smartphone, tablet and laptop without the need for glasses
  • ability to understand the study procedure
  • ability to communicate and follow the instructions of the study staff
  • written consent to participate in the study and to the processing of personal data by the research team and the sponsor of the study

Exclusion Criteria:

  • failure to provide the required information in the participant's information survey
  • failure to comply with the study procedure or the instructions of the study stuff
  • extensive facial skin damage (including abrasions, wounds, burns)
  • a disease process involving a large part of the face that makes the video- based measurement impossible
  • extensive face dressing
  • extensive facial skin tattoo or face painting
  • marked deformation of the face (e.g. a tumor)
  • inability to keep the head stable in the required position during the measurement
  • persistent cough making it impossible to remain still during the measurement
  • respiratory dysfunction in the form of shortness of breath or irregular or shallow breathing
  • diagnosed arrhythmia (except sinus bradycardia or tachycardia)
  • moderate or severe anemia
  • diagnosed: heart failure, left ventricular systolic dysfunction, aortic stenosis or other diseases or disorders within the heart vascular system leading to decreased stroke volume, low pressure amplitude or the so-called paradoxical pulse

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: HR, HRV and BR measured with Shen.AI and reference method

Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).

Reference measurements will be taken simultaneously (chest ECG for the calculation of HR and HRV; impedance pneumography for BR).

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Accuracy of heart rate (HR) measurement with the software under test compared with the ECG-based measurement
Time Frame: day 1 (single visit, approximately 30 minutes)

In each participant, average HR will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)
Accuracy of heart rate variability index (SDNN) measurement with the software under test compared with the ECG-based measurement
Time Frame: day 1 (single visit, approximately 30 minutes)

In each participant, the standard deviation of normal heartbeat intervals (SDNN) will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of breathing rate (BR) measurement with the software under test compared with the impedance pneumography-based measurement
Time Frame: day 1 (single visit, approximately 30 minutes)

In each participant, the breathing rate will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded impedance pneumography signal. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of instantaneous heart rate measurement with the software under test compared with the ECG-based measurement
Time Frame: day 1 (single visit, approximately 30 minutes)

In each participant, multiple instantaneous heart rate values (averaged over 10 s) will be estimated from the software under test and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held by someone else (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of HR using the rear camera).

For reference, HR will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held by someone else (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of SDNN using the rear camera).

For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held by someone else (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of BR using the rear camera).

For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held in hand/s (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of HR using the rear camera).

For reference, HR will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held in hand/s (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of SDNN using the rear camera).

For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held in hand/s (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of BR using the rear camera).

For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate (HR) with the software under test in daylight conditions (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of HR using the front smartphone camera).

For reference, HR will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in daylight conditions (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of SDNN using the front smartphone camera).

For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of breathing rate (BR) with the software under test in daylight conditions (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of BR using the front smartphone camera).

For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate (HR) with the software under test in the supine position (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in the supine position (1-minute measurement of HR using the front smartphone camera).

For reference, HR will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using non-parametric limits of agreement.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in the supine position (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in the supine position (1-minute measurement of SDNN using the front smartphone camera).

For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Accuracy of measurement breathing rate (BR) with the software under test in the supine position (compared with reference measurements).
Time Frame: day 1 (single visit, approximately 30 minutes)

Each participant will have one facial video measurement performed in the supine position (1-minute measurement of BR using the front smartphone camera).

For reference BR will be obtained from the simultaneously recorded impedance pneumography signal.

The results from both methods will be compared using the Bland-Altman analysis.

day 1 (single visit, approximately 30 minutes)
Utility of the software under test for self-measurement of physiological parameters (assessed with a questionnaire).
Time Frame: day 1 (after a single visit, approximately 30 minutes)

The utility of the software under test for self-measurement of physiological parameters (HR, SDNN, and BR) will be assessed based on of the average score in the participant's questionnaire completed by each participant after the test procedure (i.e. after all measurements in the given participant).

The scale will be 1-5, with higher scores meaning better outcome.

day 1 (after a single visit, approximately 30 minutes)
Utility of the software under test for the measurement of someone's physiological parameters (assessed with a questionnaire).
Time Frame: day 1 (after a single visit, approximately 30 minutes)

The utility of the software under test for the video-based measurement of someone's physiological parameters will be assessed based on the average score in the participant's questionnaire (the part of the questionnaire related to the measurement with the rear camera of a smartphone) completed by each participant after the test procedure(i.e. after all measurements in the given participant).

The scale will be 1-5, with higher scores meaning better outcome.

day 1 (after a single visit, approximately 30 minutes)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Bartłomiej Paleczny, Wrocław Medical University

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

April 26, 2023

Primary Completion (Actual)

June 6, 2023

Study Completion (Actual)

June 6, 2023

Study Registration Dates

First Submitted

April 17, 2023

First Submitted That Met QC Criteria

June 2, 2023

First Posted (Actual)

June 5, 2023

Study Record Updates

Last Update Posted (Actual)

March 28, 2024

Last Update Submitted That Met QC Criteria

March 27, 2024

Last Verified

June 1, 2023

More Information

Terms related to this study

Other Study ID Numbers

  • MDF-01-008-01

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.