New Indirect Calorimetry Device for Energy Expenditure Measurement (ICALIC)
Development and Validation of a New Indirect Calorimetry Device for Energy Expenditure Measurement in ICU Patients: The ICALIC International Multicentric Study
Energy expenditure (EE) in critically ill patients is highly variable depending on the initial severity of the disease and treatments. Clinicians need to measure EE by indirect calorimetry (IC) to optimize nutritional support. IC devices available on the market have insufficient accuracy for clinical and research use. A new IC is being developed to meet these needs.
Objectives: Validation of a reliable and easy-to-use IC device.
Primary objective:
• Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation
Secondary objectives:
- Accuracy of EE measurements in ICU patients of spontaneous breathing
- Accuracy of EE measurement during ~30 min. vs. the sum of 6 periods of 60 min/12 hrs
- Accuracy of EE measurements in ICU patients on non-invasive mechanical ventilation
Design: Prospective, controlled, observational multicenter study.
Subjects: 182 ICU patients.
Study Overview
Status
Conditions
Detailed Description
Energy expenditure (EE) in critically ill patients is highly variable depending on the initial severity of the disease and treatments. Clinicians need to measure EE by indirect calorimetry (IC) to optimize nutritional support. IC devices available on the market have insufficient accuracy for clinical and research use. A new IC is being developed to meet these needs.
Objectives: Validation of a reliable and easy-to-use IC device.
Primary objective:
• Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation
Secondary objectives:
- Accuracy of EE measurements in ICU patients of spontaneous breathing
- Accuracy of EE measurement during ~30 min. vs. the sum of 6 periods of 60 min/12 hrs
- Accuracy of EE measurements in ICU patients on non-invasive mechanical ventilation Design: Prospective, controlled, observational multicenter study. Subjects: 182 ICU patients. Sample size calculation is based on the estimated accuracy of the new IC device compared to the Deltatrac, when performed simultaneously on one subject. In a previous study, we found a difference of EE between the Deltatrac and the Quark RMR of 24±220 kcal, thus a coefficient of variation (SD/mean) of 8.8 %. We hypothesized a difference of 25±50 kcal when comparing the Deltatrac with the new IC device, thus a coefficient of variation of 2%. This reduction of 6.5% of the coefficient variation, with a power of 90% and a significance level of 0.05 was used to calculate a required sample size of 91 subjects. A drop out rate of 50% is anticipated, leading to 182 patients to be recruited to allow an analysis based on 91 patients with optimal measurements.
Inclusion criteria:
• Adult ICU patients with spontaneous breathing, or on invasive or non-invasive mechanical ventilation.
Exclusion criteria:
- FiO2 > 60%
- Kidney failure requiring Continous veno-venous hemofiltration (CVVH)
- Conditions promoting air leaks: Positive end-expiratory pressure (PEEP) >10, Peak ventilatory pressure > 30
- Chest tube (except if only for liquid drainage = no bubbling)
- Unstable hemodynamic conditions requiring changes in vasoactive drugs either 1 hour before and/or during the IC measurement (i.e. all conditions resulting in unstable pH during IC measurements).
- Change in body temperature >0.5°C either 1 hr before and/or during the IC measurement
- Physical agitation or curarization either 1 hour before and/or during the IC measurement
Outcomes:
Primary outcome:
• EE in intubated ICU patients on mechanical ventilation
Secondary outcomes:
- EE in ICU patients of spontaneous breathing
- EE during ~30 minutes vs. the sum of 6 periods of 60 minutes/12 hours
- EE measurements when using different types of ventilators
- Time to start EE measurement (calibration, warming up)
- Duration for IC disinfection New Calorimeter - Protocol - November 11. 2013 Page 3 sur 5
- Limitations of EE measurements in patients on non-invasive mechanical ventilation
Indirect calorimetry devices and methodology:
- New IC device (Cosmed, Italy) and Deltatrac II (Datex, Finland).
- Briefly, patients are laying down in a supine position on a bed during ongoing measure of ~30 min (see Figure 1) by connecting the calorimeter to the endotracheal tube (invasive mechanical ventilation) or to the facemask (non-invasive mechanical ventilation), or by using a transparent canopy in plexiglas to cover the head while room-air flows through and is mixed with the expirate (spontaneous breathing). This mixture of patient breath and room air is finally collected and analyzed by the IC device for O2 and CO2 concentrations, and used to calculate EE. The first 2 min of data is eliminated as acclimatization artefact. From the remaining ~28 min, a segment of 5 consecutive 1-min measures with 5% coefficient of variation in VO2 and VCO2 is considered as steady state.
Barometric pressure is recorded. EE is calculated using the modified Weir equation.
- Calibration of the gas analyzers is carried out before /after each measurement according to the manufacture instructions, including ethanol burning for the Deltatrac II.
- EE measurements is performed simultaneously by the 2 IC devices.
Data collected:
- Sex, Age (yr), height (±2 cm), dry weight (±1kg)
- Heart rate, Body temperature
- Admission diagnosis, Major treatments
- SAPS II and SOFA on ICU admission day; SOFA on the study day
- Ventilator type (name) and ventilation mode (Minute ventilation, FiO2)
- Inspiratory and expiratory O2, CO2, air flow (ml/min) measured by the IC
Statistics and data analysis:
The Coordinating center analyses the data transferred by the Investigation centers
- All data are reported as means±SD. The parameters (see above) of the two IC devices is compared by Bland and Altman plots
- Reliability and adequacy (sensitivity, specificity, positive and negative predictive values) between the IC devices, is tested using ROC curves with kappa coefficient (reliability coefficient). For coefficient of variation, ANOVA is used when applicable. Results are adjusted using Greenhouse-Geisser correction in case of non-sphericity. Post-hoc tests using Bonferroni correction compared the results between two IC devices. The prediction of EE differences between devices are analyzed by multivariate regression. As 91 are expected to complete the entire protocol, 9 factors are introduced in the model (Peduzzi P, J Clin Epidemiol 1996, 49, 1373-9), i.e. ventilator type, ventilator mode, minute ventilation, FiO2, BMI, SOFA, heart rate, body temperature, medical/surgical category.
Statistical analysis significance is set at p<0.05.
For the subgroup of patients with prolonged EE measurements or on non-invasive ventilation:
Univariate and multivariate analysis are used to analyze the factors associated with observed variations.
Value of the study for clinical practice practice and research:
An accurate and reliable IC for nutrition specialists will allow to optimize energy prescription, a factor shown to improve the clinical outcome. An easy-to-use and affordable (10'000 $) IC will reduce both the burden on human resources and costs.
Study originality:
- Development of an IC device meeting technical characteristics defined by physicians (bottom-up strategy)
- Demonstration of the clinical relevance of short EE measurements, practically convenient and less human-resources consuming, versus longer EE measurements New Calorimeter - Protocol - November 11. 2013 Page 4 sur 5
- Exploratory comparative tests, whenever possible, to evaluate IC in a subgroup of patients with non-invasive ventilation New Calorimeter - Protocol - November 11. 2013 Page 5 sur 5
Agenda of actions:
- The IC prototype will be made available the 18th of May 2014.
- Recruitment of subjects will begin in May-June 2014 and will continue for 1 year, or until sufficient subjects are included in the study.
Study Type
Enrollment (Anticipated)
Contacts and Locations
Study Contact
- Name: claude Pichard, Prof
- Phone Number: +41 22 372 9349
- Email: claude.pichard@unige.ch
Study Contact Backup
- Name: Claudia P Heidegger, MD
- Phone Number: +41 22 372 7440
- Email: claudia-paula.heidegger@hcuge.ch
Study Locations
-
-
-
Geneva, Switzerland, 1211
- Recruiting
- Geneva University Hospital- Nutrition Unit
-
Contact:
- claude Pichard, Prof
- Phone Number: +41 22 372 9349
- Email: claude.pichard@unige.ch
-
Contact:
- Claudia P Heudegger, MD
- Phone Number: +41 22 372 7440
- Email: claudia-paula.heidegger@hcuge.ch
-
Sub-Investigator:
- Claudia P Heidegger, MD
-
Sub-Investigator:
- Anne Berit Guttormsen, MD
-
Sub-Investigator:
- Michael Hiesmayr, MD
-
Sub-Investigator:
- Pierre Singer, MD
-
Sub-Investigator:
- Jan Wernerman, MD
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Adult ICU patients with spontaneous breathing, or on invasive or non-invasive mechanical ventilation.
Exclusion Criteria:
FiO2 > 60%
- Kidney failure requiring Continous veno-venous hemofiltration (CVVH)
- Conditions promoting air leaks: Positive end-expiratory pressure (PEEP) >10, Peak ventilatory pressure > 30
- Chest tube (except if only for liquid drainage = no bubbling)
- Unstable hemodynamic conditions requiring changes in vasoactive drugs either 1 hour before and/or during the IC measurement (i.e. all conditions resulting in unstable pH during IC measurements).
- Change in body temperature >0.5°C either 1 hr before and/or during the IC measurement
- Physical agitation or curarization either 1 hour before and/or during the IC measurement
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
---|
indirect calorimetry measurement
Measurement of energy expenditure in made by indirect calorimetry in patients laying down in a supine position on a bed during ongoing measure by connecting the calorimeter to the endotracheal tube (invasive mechanical ventilation) or to the facemask (non-invasive mechanical ventilation), or by using a transparent canopy in plexiglas to cover the head while room-air flows through and is mixed with the expirate (spontaneous breathing).
This mixture of patient breath and room air is finally collected and analyzed by the IC device for O2 and CO2 concentrations, and used to calculate EE.
EE is calculated using the modified Weir equation.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
calories reference IC devices in intubated ICU patients on mechanical ventilation
Time Frame: 30 minutes
|
Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation
|
30 minutes
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
calories
Time Frame: 30 minutes
|
Accuracy of EE measurements in ICU patients of spontaneous breathing
|
30 minutes
|
calories
Time Frame: 60 minutes
|
Accuracy of EE measurement during ~30 min.
vs. the sum of 6 periods of 60 min/12 hrs
|
60 minutes
|
calories
Time Frame: 30 minutes
|
Accuracy of EE measurements in ICU patients on non-invasive mechanical ventilation
|
30 minutes
|
Collaborators and Investigators
Sponsor
Collaborators
Investigators
- Principal Investigator: Claude Pichard, Prof, University Hospital, Geneva
Study record dates
Study Major Dates
Study Start (ACTUAL)
Primary Completion (ANTICIPATED)
Study Completion (ANTICIPATED)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (ESTIMATE)
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- ICALIC
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Acute Respiratory Failure
-
Efficacy Care R&D LtdMemorial Hermann Hospital; CRG Medical, Inc.UnknownShock | Shock, Septic | Respiratory Failure | Respiratory Distress Syndrome | Shock, Cardiogenic | Acute Cardiac Failure | Acute Respiratory Failure | Acute Kidney Failure | Multi Organ Failure | Respiratory Arrest | Acute Respiratory Failure With Hypoxia | Acute Respiratory Failure Requiring Reintubation | Acute... and other conditionsUnited States
-
Dr. Behcet Uz Children's HospitalCompletedAcute Respiratory Failure | Acute Hypoxemic Respiratory Failure | Acute Hypoxemic and Hypercapnic Respiratory FailureTurkey
-
Siriraj HospitalRecruitingAcute Hypoxemic Respiratory Failure | Acute Hypercapnic Respiratory FailureThailand
-
Fisher and Paykel HealthcareCentre hospitalier de l'Université de Montréal (CHUM); Institut universitaire...Not yet recruitingAcute Hypoxemic Respiratory Failure | Acute Hypercapnic Respiratory Failure
-
Yonsei UniversityNot yet recruitingRespiratory Failure | Respiratory Insufficiency | Hypercapnic Respiratory Failure | Acute Hypoxemic Respiratory Failure | Acute Hypercapnic Respiratory Failure | Ventilatory Depression | Hypoxemic Respiratory Failure | Respiratory Depression | Hypoxemic Acute Respiratory Failure | Hypercapnic AcuteKorea, Republic of
-
Beijing Chao Yang HospitalRecruitingPatients With Acute Hypoxemic Respiratory FailureChina
-
Raffaele ScalaRecruitingAcute Respiratory Failure | Airway Clearance Impairment | Acute-on-chronic Respiratory FailureItaly
-
Catholic University of the Sacred HeartFisher and Paykel HealthcareCompletedWeaning Failure | Acute Respiratory FailureFrance, Greece, Italy, Spain
-
Poitiers University HospitalRecruitingAcute Respiratory Failure | Hypoxemic Respiratory FailureFrance
-
Assiut UniversityNot yet recruitingAcute Hypoxemic Respiratory Failure | Acute Hypercapnic Respiratory FailureEgypt