High Flow Ventilation With Volume Guarantee
Pilot Study: High Flow Ventilation With Volume Guarantee
Study Overview
Status
Intervention / Treatment
Detailed Description
HFV is a way to deliver mechanical breathing support to patients with respiratory failure applied in Neonatal Intensive Care Units (NICU's). In some NICU's, this mode of ventilation is applied when conventional ventilation fails, whereas in other units HFV is used as the primary ventilation mode.
HFV is a ventilator mode where high frequency pressure changes, typically 6-12 per second (6 - 12 Hertz (Hz)), is used to exchange air between the lungs (alveoli) and the ventilator tubing.
Concerns have been raised regarding fluctuating PCO2 values in babies receiving HFV. It is believed that stable normocapnia (normal PCO2 values) as opposed to hyper- and hypocapnia is associated to better short and long-term outcomes.
In conventional ventilator modes, CO2 diffusion is a function of the product from volume per breath (Vt) and the frequency (f) of breathing (Vt*f = minute volume (MV)). In HFV the CO2 diffusion is a function of a product from the small tidal volumes generated (Vthf) squared, and the oscillation frequency (Hz) delivered by the ventilator (vthf2 * F=DCO2). The small tidal volumes are being adjusted by altering the pressure range (amplitude) and the frequency by which pressure changes is generated. One challenge in using HFV has been unwanted fluctuations in PCO2 values, which has been shown to be associated with clinical important complications, in particular cerebral hemorrhages. In order to avoid substantial fluctuation in PCO2 a meticulous adjustment of the amplitude is required.
Modern neonatal ventilators now offer the ability to automatically adjust the amplitude of the high frequency breaths to obtain a stable high frequency tidal volume. The ventilator measures the Vthf and increase or decrease the power of each breath to obtain the set Vthf. This mode is called high frequency ventilation with volume guarantee (HFV+VG).
Automatically adjustment of the amplitude is a novel option in high frequency ventlators and has not been studied in randomized trials.
This study will investigate compare HFV with and without VG. Patients treated with HFV will be randomized to conventional HFV or HFV+VG. After 24 hours the mode will be changed to HFV+VG for those starting on HFV and to HFV for those starting on HFV+VG for 24 hours. After the trial period of 24 x 2 hours - the HFV mode will be at the clinician's preference.
To compare the efficiency of both modes of HFV, measured PCO2 values, oxygen need and variables from the ventilator will be sampled. These variables describe the effort from the ventilator (frequency, amplitude) as well as variables associated with gas exchange in the lungs (Vthf, DCO2).
Study Type
Enrollment (Anticipated)
Phase
- Not Applicable
Contacts and Locations
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- all neonates in need of high frequency ventilator
Exclusion Criteria:
- neonates expected to need high frequency ventilator less than 48 hours because of planned surgery etc
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Crossover Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: HFV+vg
volume guarantee given
|
High frequency ventilation with and without volume guarantuee
|
|
Experimental: HFV-VG
Volume guarantee not given
|
High frequency ventilation with and without volume guarantuee
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
stability of High Frequency Tidal volumes
Time Frame: during study, ie. 48 hours.
|
HFVVt are measured by ventilator.
|
during study, ie. 48 hours.
|
|
Stability of CO2 measurements
Time Frame: during study, ie. 48 hours.
|
Measured CO2 values will be measured as usual
|
during study, ie. 48 hours.
|
|
frequency of needed manual adjustments on the ventilator
Time Frame: during study, ie 48 hours
|
Adjustments made on ventilator will be automatically sampled by the ventilator
|
during study, ie 48 hours
|
Collaborators and Investigators
Sponsor
Collaborators
Investigators
- Principal Investigator: Bjorn e Ogland, md dr philos, Oslo University Hospital
Study record dates
Study Major Dates
Study Start
Primary Completion (Anticipated)
Study Completion (Anticipated)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- OsloUH_bogland
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.