Enhanced Lung Protective Ventilation With ECCO2R During ARDS (PROVE)

December 5, 2023 updated by: Jerome Allardet-Servent, MD, Hôpital Européen Marseille

Enhanced Lung Protective Ventilation With Extracorporeal CO2 Removal During Acute Respiratory Distress Syndrome

Acute Respiratory Distress Syndrome (ARDS) is associated with a mortality rate of 30 - 45 % and required invasive mechanical ventilation (MV) in almost 85 % of patients[1]. During controlled MV, driving pressure (i.e., the difference between end-inspiratory and end-expiratory airway pressure) depends of both tidal volume and respiratory system compliance. Either excessive tidal volume or reduced lung aeration may increase the driving pressure. ARDS patients receiving tidal volume of 6 ml/kg predicted body weight (PBW) and having a day-1 driving pressure ≥ 14 cmH2O have an increased risk of death in the hospital[2]. Seemly, in the LUNG SAFE observational cohort, ARDS patients having a day-1 driving pressure < 11 cmH2O had the lowest risk of death in the hospital[1]. Hence, driving pressure acts as a major contributor of mortality in ARDS, and probably reflects excessive regional lung distension resulting in pro-inflammatory and fibrotic biological processes. Whether decreasing the driving pressure by an intervention change mortality remains an hypothesis; but one of means is to decrease the tidal volume from 6 to 4 ml/ kg predicted body weight (PBW). However, this strategy promotes hypercarbia, at constant respiratory rate, by decreasing the alveolar ventilation. In this setting, implementing an extracorporeal CO2 removal (ECCO2R) therapy prevents from hypercarbia. A number of low-flow ECCO2R devices are now available and some of those use renal replacement therapy (RRT) platform. The investigators previously reported that combining a membrane oxygenator (0.65 m²) within a hemofiltration circuit provides efficacious low flow ECCO2R and blood purification in patients presenting with both ARDS and Acute Kidney injury[3].

This study aims to investigate the efficacy of an original ECCO2R system combining a 0.67 m² membrane oxygenator (Lilliput 2, SORIN) inserted within a specific circuit (HP-X, BAXTER) and mounted on a RRT monitor (PrismafleX, BAXTER). Such a therapy only aims to provide decarboxylation but not blood purification and has the huge advantage to be potentially implemented in most ICUs without requiring a specific ECCO2R device. The study will consist in three periods:

  • The first period will address the efficacy of this original ECCO2R system at tidal volume of 6 and 4 ml/kg PBW using an off-on-off design.
  • The second part will investigate the effect of varying the sweep gas flow (0-2-4-6-8-10 l/min) and the mixture of the sweep gas (Air/O2) on the CO2 removal rate.
  • The third part will compare three ventilatory strategies applied in a crossover design:

    1. Minimal distension: Tidal volume 4 ml/kg PBW and positive end-expiratory pressure (PEEP) based on the ARDSNet PEEP/FiO2 table (ARMA).
    2. Maximal recruitment: 4 ml/kg PBW and PEEP adjusted to maintain a plateau pressure between 23 - 25 cmH2O.
    3. Standard: Tidal volume 6 ml/kg and PEEP based on the ARDSNet PEEP/FiO2 table (ARMA).

Study Overview

Study Type

Interventional

Enrollment (Actual)

3

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Marseille, France
        • Service de REANIMATION, HOPITAL EUROPEEN MARSEILLE

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • ARDS moderate or severe (Berlin criteria)
  • Onset < 48 h
  • Driving pressure ≥ 11 cmH2O

Exclusion Criteria:

  • Lack of consent or social protection
  • Chronic respiratory failure (requiring Oxygen or NIPPV)
  • Severe hypoxemia: PaO2/FIO2 < 80 with PEEP ≥ 18 cmH2O AND FIO2= 1
  • Acute Renal Failure requiring RRT
  • DNR order or death expected within the next 72 hours
  • Planned surgery or transport out-of-ICU expected within the next 72 hours
  • Heparin allergy
  • Contraindication to jugular vein catheterization
  • Intracranial Hypertension

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Minimal distension
Tidal volume 4 ml/kg PBW and positive end-expiratory pressure (PEEP) based on the ARDSNet PEEP/FiO2 table (ARMA) + ECCO2R (sweep gas = 8 L/min, blood flow = 400 mL/min)
Low flow Extracorporeal CO2 removal using a 0.67 m² membrane oxygenator (Lilliput 2) and a specific circuit (HP-X) mounted on a RRT monitor (PrismafleX)
Experimental: Maximal recruitment
Tidal volume 4 ml/kg PBW and PEEP adjusted to maintain a plateau pressure between 23 - 25 cmH2O + ECCO2R (sweep gas = 8 L/min, blood flow = 400 mL/min)
Low flow Extracorporeal CO2 removal using a 0.67 m² membrane oxygenator (Lilliput 2) and a specific circuit (HP-X) mounted on a RRT monitor (PrismafleX)
Active Comparator: Standard
Tidal volume 6 ml/kg PBW and positive end-expiratory pressure (PEEP) based on the ARDSNet PEEP/FiO2 table (ARMA) without ECCO2R (no sweep gas flow, blood flow = 400 mL/min)
Low flow Extracorporeal CO2 removal using a 0.67 m² membrane oxygenator (Lilliput 2) and a specific circuit (HP-X) mounted on a RRT monitor (PrismafleX)

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in PaCO2
Time Frame: 15 minutes after initiation of ECCO2R at tidal volume of 4 ml/kg PBW.
20 % decrease in PaCO2 after initiation of ECCO2R at tidal volume of 4 ml/kg PBW (as compared to 4 ml/kg without ECCO2R)
15 minutes after initiation of ECCO2R at tidal volume of 4 ml/kg PBW.

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
PaCO2
Time Frame: each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Arterial blood gas analyser (RAPIDPoint 500)
each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
CO2 removal rate
Time Frame: each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using measurements from both the blood side and the gas side (two methods)
each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Transpulmonary pressure and work of breathing
Time Frame: each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using an oesophageal balloon catheter (NutriVent catheter) and a dedicated monitor (FluxMed, MBMed)
each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Regional tidal ventilation
Time Frame: each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using an Electrical Impedance Tomography device (BB², Swisstom)
each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
End-expiratory Lung Volume
Time Frame: each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using the nitrogen wash-in wash-out method (Engstrom GE)
each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Plasma Cytokines
Time Frame: Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using Elisa custom kit (Qiagen) from plasma samples
Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Pulmonary Cytokines
Time Frame: Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using Elisa custom kit (Qiagen) from BAL samples
Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Type III Procollagen
Time Frame: Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using both RIA and Elisa methods from plasma and BAL samples
Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Pulmonary Inflammatory and Fibrotic pathway
Time Frame: Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.
Using mRNA custom kit RT-PCR analysis (Qiagen) from BAL samples
Only in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Plasma Free Hemoglobin
Time Frame: every 24 hours, up to 72 hours.
serum samples
every 24 hours, up to 72 hours.
Haptoglobin
Time Frame: every 24 hours, up to 72 hours.
serum samples
every 24 hours, up to 72 hours.
Lacticodéshydrogenase (LDH)
Time Frame: every 24 hours, up to 72 hours.
serum samples
every 24 hours, up to 72 hours.
schizocytes
Time Frame: every 24 hours, up to 72 hours.
serum samples
every 24 hours, up to 72 hours.
Bilirubin
Time Frame: every 24 hours, up to 72 hours.
serum samples
every 24 hours, up to 72 hours.

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Jérôme ALLARDET-SERVENT, MD, Hopital Europeen Marseille

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 23, 2018

Primary Completion (Actual)

December 5, 2023

Study Completion (Actual)

December 5, 2023

Study Registration Dates

First Submitted

May 3, 2018

First Submitted That Met QC Criteria

May 3, 2018

First Posted (Actual)

May 16, 2018

Study Record Updates

Last Update Posted (Estimated)

December 12, 2023

Last Update Submitted That Met QC Criteria

December 5, 2023

Last Verified

December 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.