OM2 Abbreviated Sensor Verification
May 12, 2026 updated by: Medtronic - MITG
Verification of Investigational Pulse Oximetry Devices With the Abbreviated Sensor Line in Adult Volunteers
The verification of an investigational pulse oximetry board to verify pulse rate and saturation accuracy over a specified saturation range in diverse populations.
Study Overview
Status
Completed
Conditions
Detailed Description
To verify the SpO2 and pulse rate accuracy in comparison to reference-standard blood measurements.
To achieve paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of the prototype pulse oximeter on a group of healthy adult volunteers.
The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods.
In this study, the investigational pulse oximetry PCBA system saturation measurements will be compared to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from a diverse pool of healthy human subjects.
Study Type
Observational
Enrollment (Actual)
28
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
Colorado
-
Louisville, Colorado, United States, 80027
- Element
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Yes
Sampling Method
Non-Probability Sample
Study Population
Healthy Male or Female volunteers age 18 to 50 years, weighing greater than 40kgs.
Volunteers must be non-smokers or not smoked within 2 days prior to the study.
Cleared a same day health assessment form and health screen to ensure none of the exclusion criteria was met.
With a successful perfusion index ulnar/ulnar+radial ratio test showing adequate collateral blood flow.
Description
Inclusion Criteria:
- Male or female subjects 18 to 50 years of age
- Subject is willing and able to comply with study procedures and duration
- Subject is willing to sign an informed consent
- Subject weighs >40kg
- Subject is a non-smoker or has not smoked within 2 days prior to the study
- Cleared same day health assessment form and health screening
- Successful perfusion index ulnar/ulnar+radial ratio test showing adequate collateral blood flow.
Exclusion Criteria:
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
- Females of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study
- Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
- tHb < 10 g/dl as assessed by CO-Oximetry during the procedure
- MetHb ≥ 2% as assessed by CO-Oximetry during the procedure
Subjects with known respiratory conditions such as:
- uncontrolled / severe asthma
- flu or influenza type infection
- pneumonia / bronchitis
- shortness of breath / respiratory distress
- unresolved respiratory or lung surgery
- emphysema, COPD, lung disease
- recent COVID (last 2 months)
Subjects with known heart or cardiovascular conditions such as:
- hypertension: systolic >140mmHg, or Diastolic >90mmHg on 3 consecutive readings
- have had cardiovascular surgery
- chest pain (angina)
- heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- implantable active medical device such as pacemaker or automatic defibrillator
Self-reported health conditions as identified in the Health Assessment Form
- diabetes
- uncontrolled thyroid disease
- kidney disease / chronic renal impairment
- history of seizures (except childhood febrile seizures)
- epilepsy
- history of unexplained syncope
- recent history of frequent migraine headaches
- recent symptomatic head injury, within the last 2 months
- cancer requiring chemotherapy, radiation, or current treatment
- subjects with known clotting disorders
- history of bleeding disorders or personal history of prolonged bleeding from injury
- history of blood clots
- hemophilia
- sickle cell trait or disease
- current use of blood thinner: prescription or daily use of aspirin
- subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
- participants with severe allergy to iodine (only applicable if iodine is used)
- subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin
- arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site, left or right)
- history of clinically significant complications from previous arterial cannulation
- unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits
- other known health condition, upon disclosure in Health Assessment form
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
|---|
|
Healthy Volunteers
Observational Study
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Proposed claim verification for pulse rate and saturation accuracy
Time Frame: 2.5 hours
|
Acceptance criteria for
|
2.5 hours
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Investigators
- Principal Investigator: Monica Rabanal, Element Labs
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
February 11, 2024
Primary Completion (Actual)
February 22, 2024
Study Completion (Actual)
February 22, 2024
Study Registration Dates
First Submitted
April 19, 2024
First Submitted That Met QC Criteria
May 14, 2024
First Posted (Actual)
May 16, 2024
Study Record Updates
Last Update Posted (Actual)
May 14, 2026
Last Update Submitted That Met QC Criteria
May 12, 2026
Last Verified
April 1, 2024
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- MDT23027OM2VAB
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
No
product manufactured in and exported from the U.S.
Yes
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hypoxia
-
Direction Centrale du Service de Santé des ArméesNot yet recruitingHypoxia | Hypoxia, Brain | Hypoxia in Healthy Individuals | Hypoxia, Altitude | Altitude Hypoxia | Altitude | Hypoxia Altitude Simulation Test | Hypoxia Brain | Normobaric HypoxiaFrance
-
Western University, CanadaCompletedBrain Hypoxia IschemiaCanada
-
Universite du Littoral Cote d'OpaleNot yet recruitingNormoxia | Intermittent Moderate Hypoxia | Continuous Moderate HypoxiaFrance
-
Jorge Torres MejíasEnrolling by invitationChanges in the Intestinal Microbiota Under Hypobaric Hypoxia Conditions | Changes in Blood Glucose Under Hypobaric Hypoxia Conditions | Changes in Body Composition Under Hypobaric Hypoxia Conditions | Determination of Physiological Stress in Hypobaric HypoxiaChile
-
Yale UniversityTianjin Chest HospitalCompletedBrain Ischemia Hypoxia | Muscle; Ischemic | Muscle HypoxiaChina
-
Fliegerärztliches InstitutCompleted
-
University of Texas at AustinCompletedIntermittent HypoxiaUnited States
-
Owlet Baby Care, Inc.Completed
-
Robert L. OwensCompletedIntermittent HypoxiaUnited States
-
Montreal Heart InstituteUniversité de MontréalCompleted