Effects of Inspiration Rise Time on Work of Breathing and Comfort of Conscious Patients on Mechanical Ventilation
Introduction Pressure Support Ventilation is widely used in patients in the ICU. Matching the patient's respiratory needs with adequate ventilator settings is necessary to ensure a low work of breathing (WOB) and maximal patient comfort. The inspiratory rise time (IRT) determines the time to reach the selected airway pressure. A short IRT results in a high peak inspiratory flow and a short time to reach that peak, but is also associated with the development of turbulent flow, resulting in increased WOB. Aim of this study is to investigate the effects of different IRT settings on WOB and patient comfort during pressure support ventilation.
Methods We will performed a prospective, single blind cohort study in patients on Pressure Support Ventilation. 10 healthy adult patients admitted to the ICU after elective facial or neck surgery will be included. All patients will be ventilated in pressure support mode using a Servo 300 ventilator (Siemens. Elema, Solna, Sweden), with a positive end expiratory pressure of 5 cm H2O, pressure support level of 12 cm H2O above PEEP and an inspiratory oxygen fraction of 0.40. Patients have to be awake and cooperative (Ramsay 2). WOB will be measured with an esophageal balloon, and miniature flowmeter (Bicore system). Breathing comfort will be evaluated using a visual analogue scale (VAS) ranging from 1 to 10. WOB and patient comfort will be measured (in random order) at 0, 5, and 10% IRT. For statistical analysis the two-way analysis of variance will used. A p value of < 0.05 will be considered statistically significant.
Přehled studie
Postavení
Podmínky
Intervence / Léčba
Detailní popis
Introduction Pressure Support Ventilation is widely used in patients in the ICU. Matching the patient's respiratory needs with adequate ventilator settings is necessary to ensure a low work of breathing (WOB) and maximal patient comfort. The inspiratory rise time (IRT) determines the time to reach the selected airway pressure. A short IRT results in a high peak inspiratory flow and a short time to reach that peak, but is also associated with the development of turbulent flow, resulting in increased WOB. Aim of this study is to investigate the effects of different IRT settings on WOB and patient comfort during pressure support ventilation.
Methods We will performed a prospective, single blind cohort study in patients on Pressure Support Ventilation. 10 healthy adult patients admitted to the ICU after elective facial or neck surgery will be included. All patients will be ventilated in pressure support mode using a Servo 300 ventilator (Siemens. Elema, Solna, Sweden), with a positive end expiratory pressure of 5 cm H2O, pressure support level of 12 cm H2O above PEEP and an inspiratory oxygen fraction of 0.40. Patients have to be awake and cooperative (Ramsay 2). WOB will be measured with an esophageal balloon, and miniature flowmeter (Bicore system). Breathing comfort will be evaluated using a visual analogue scale (VAS) ranging from 1 to 10. WOB and patient comfort will be measured (in random order) at 0, 5, and 10% IRT. For statistical analysis the two-way analysis of variance will used. A p value of < 0.05 will be considered statistically significant.
Typ studie
Zápis (Očekávaný)
Fáze
- Nelze použít
Kontakty a umístění
Studijní místa
-
-
-
Nijmegen, Holandsko, 6500 HB
- Radboud University Nijmegen Medical Centre
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Popis
Inclusion Criteria:
- awake and cooperative patient
- ICU admission following surgery of head or neck
Exclusion Criteria:
- coagulopathy
- inability to sit
- hypoxemia and/or hypercapnia and /or hypocapnia
- Pain score > 5 (on scale 1-10)
- esophageal abnormalities
- Pregnancy
- COPD
Studijní plán
Jak je studie koncipována?
Detaily designu
- Přidělení: Randomizované
- Intervenční model: Crossover Assignment
- Maskování: Singl
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
---|
Work of breathing
|
Sekundární výstupní opatření
Měření výsledku |
---|
patient comfort
|
Spolupracovníci a vyšetřovatelé
Vyšetřovatelé
- Vrchní vyšetřovatel: Cornelia WE Hoedemaekers, MD PhD, Radboud University Medical Center
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia
Primární dokončení
Dokončení studie (Aktuální)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Odhad)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Odhad)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další identifikační čísla studie
- LBCH 0601
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
produkt vyrobený a vyvážený z USA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .