User Study of an Investigational Blood Glucose Monitoring System
29. januar 2016 opdateret af: Ascensia Diabetes Care
Tatsu/Tradewind System User Performance Evaluation
This clinical trial is designed to assess an investigational blood glucose monitoring system with fingerstick blood in the hands of untrained intended users.
Studieoversigt
Status
Afsluttet
Betingelser
Intervention / Behandling
Undersøgelsestype
Interventionel
Tilmelding (Faktiske)
118
Fase
- Ikke anvendelig
Kontakter og lokationer
Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.
Studiesteder
-
-
California
-
Escondido, California, Forenede Stater, 92026
- AMCR Institute
-
-
Deltagelseskriterier
Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.
Berettigelseskriterier
Aldre berettiget til at studere
18 år og ældre (Voksen, Ældre voksen)
Tager imod sunde frivillige
Ingen
Køn, der er berettiget til at studere
Alle
Beskrivelse
Inclusion Criteria:
- Type 1 or type 2 diabetes
- 18 years of age and older
- Able to speak, read, and understand English
- Willing to complete all study procedures
Exclusion Criteria:
- Previously participated in a study using this investigational system
- Working for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors
- Pregnancy
- Hemophilia or any other bleeding disorder
- Working for a competitive medical device company or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company
- A condition which, in the opinion of the investigator, would put the person or study conduct at risk (reason for exclusion will be clearly documented by investigator or designee).
Studieplan
Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Diagnostisk
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: Intended Users of the System
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick.
|
Tradewind is a Bayer investigational meter that used an investigational sensor.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
Tidsramme: 1 hour
|
Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor.
BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer.
BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results <75mg/dL) or +/- 20% (for reference BG results >=75mg/dL) of the reference method results (YSI capillary plasma).
|
1 hour
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Number of Study Staff Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method When Testing Subject Blood Glucose (BG)
Tidsramme: 1 hour
|
Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor.
BGM results were compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer.
BG meter results were used to calculate the number of BG results within +/- 15mg/dL (for reference BG results <75mg/dL) or +/- 20% (for reference BG results >=75mg/dL) of the reference method results.
|
1 hour
|
Samarbejdspartnere og efterforskere
Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.
Sponsor
Datoer for undersøgelser
Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.
Studer store datoer
Studiestart
1. september 2011
Primær færdiggørelse (Faktiske)
1. september 2011
Studieafslutning (Faktiske)
1. september 2011
Datoer for studieregistrering
Først indsendt
4. oktober 2011
Først indsendt, der opfyldte QC-kriterier
4. oktober 2011
Først opslået (Skøn)
6. oktober 2011
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Skøn)
29. februar 2016
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
29. januar 2016
Sidst verificeret
1. januar 2016
Mere information
Begreber relateret til denne undersøgelse
Andre undersøgelses-id-numre
- CTD-PRO-2011-005-01
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Tatsu/Tradewind Investigational BG Monitoring System
-
Ascensia Diabetes CareAfsluttet
-
Ascensia Diabetes CareAfsluttetDiabetesForenede Stater
-
Ascensia Diabetes CareAfsluttet
-
Ascensia Diabetes CareAfsluttet
-
Ascensia Diabetes CareAfsluttet
-
Ascensia Diabetes CareAfsluttetDiabetesForenede Stater
-
Ascensia Diabetes CareAfsluttetDiabetesForenede Stater
-
Ascensia Diabetes CareAfsluttetDiabetesForenede Stater
-
Ascensia Diabetes CareAfsluttetDiabetesForenede Stater
-
Emory UniversityAmerican Heart Association; Abbott Diabetes CareIkke rekrutterer endnu