The Effect of Different Measurement-manufacturing Techniques on Load Distribution in Amputees
27. marts 2020 opdateret af: Marmara University
The Effect of Different Measurement-manufacturing Techniques on Stump and Soles Load Distribution in Transtibial Knee Prostheses
Alignment of the prosthesis with the stump is performed through the socket.
Good socket stump fit provides near-normal posture and weight-transfer pattern.
It also ensures that the walk is painless and low-energy.
Recently, studies have been carried out to use 3D printers in socket production.
3D method is accepted as an effective and reliable method in prosthetic socket production.
The purpose of this study is to compare the pressures distrubution in the sockets produced by conventional and 3D scanners and printers in the transtibial amputee.
Studieoversigt
Status
Ukendt
Betingelser
Intervention / Behandling
Undersøgelsestype
Observationel
Tilmelding (Forventet)
30
Kontakter og lokationer
Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.
Studiesteder
-
-
-
Istanbul, Kalkun, 34000
- Marmara University
-
-
Deltagelseskriterier
Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.
Berettigelseskriterier
Aldre berettiget til at studere
18 år til 70 år (Voksen, Ældre voksen)
Tager imod sunde frivillige
Ingen
Køn, der er berettiget til at studere
Alle
Prøveudtagningsmetode
Sandsynlighedsprøve
Studiebefolkning
Unilateral transtibial amputees
Beskrivelse
Inclusion Criteria:
- Being unilateral transtibial amputation
- To have enough perception level to understand and apply commands
Exclusion Criteria:
- To have orthopedic, neurological, etc. disorder other than amputation that will affect posture or gait
- The presence of pain that will affect the posture or walking pattern
- Wound / tenderness on the stump
- Visual impairment that will affect the posture or gait pattern
- Use of any assistive device (walking stick, canary, etc.) during walking
Studieplan
Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
Kohorter og interventioner
Gruppe / kohorte |
Intervention / Behandling |
---|---|
Group 1
Unilateral transtibial amputees using conventinional manufactured socket
|
Presssure disturubition will be measured in conventional manufactured socket during walking and stairing up and down activities in unilateral transtibial amputees
|
Group 2
Unilateral transtibial amputees using 3D manufactured socket
|
Presssure disturubition will be measured in 3D manufactured socket during walking and stairing up and down activities in unilateral transtibial amputees
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
---|---|---|
Measurement of load distrubution on stump in conventional manufactured socket
Tidsramme: Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during walking
|
Load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during walking.
|
Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during walking
|
Measurement of load distrubution on stump in 3D manufactured socket
Tidsramme: Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during walking
|
Load distrubution (mmHg/N/Kg) on stump will be measured with F-socket sensories during walking.
|
Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during walking
|
Measurement of load distrubution on stump in conventional manufactured socket
Tidsramme: Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs.
|
Load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs.
|
Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs.
|
Measurement of load distrubution on stump in 3D-manufactured socket
Tidsramme: Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs
|
Load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs
|
Change from load distrubution (mmHg/N/Kg) on stump will be measured with F-socket film sensories during climbing up and down stairs
|
Measurement of sole load distrubution (mmHg/N/Kg) in conventional manufactured socket
Tidsramme: Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during walking
|
Sole load distrubution (mmHg/N/Kg) will be measured with F-scan sensories during walking
|
Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during walking
|
Measurement of sole load distrubution (mmHg/N/Kg) in 3D-manufactured socket
Tidsramme: Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during walking
|
Sole load distrubution (mmHg/N/Kg) will be measured with F-scan sensories during walking
|
Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during walking
|
Measurement of sole load distrubution (mmHg/N/Kg) in conventional manufactured socket
Tidsramme: Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during climbing up and down stairs
|
Sole load distrubution (mmHg/N/Kg) will be measured with F-scan sensories during climbing up and down stairs
|
Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during climbing up and down stairs
|
Measurement of sole load distrubution (mmHg/N/Kg) in 3D-manufactured socket
Tidsramme: Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during climbing up and down stairs
|
Sole load distrubution (mmHg/N/Kg) will be measured with F-scan sensories during climbing up and down stairs
|
Change from sole load distrubution (mmHg/N/Kg) will be measured with F-socket film sensories during climbing up and down stairs
|
Samarbejdspartnere og efterforskere
Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.
Sponsor
Efterforskere
- Studieleder: yasar tatar, Marmara University
Publikationer og nyttige links
Den person, der er ansvarlig for at indtaste oplysninger om undersøgelsen, leverer frivilligt disse publikationer. Disse kan handle om alt relateret til undersøgelsen.
Generelle publikationer
- Dou P, Jia X, Suo S, Wang R, Zhang M. Pressure distribution at the stump/socket interface in transtibial amputees during walking on stairs, slope and non-flat road. Clin Biomech (Bristol, Avon). 2006 Dec;21(10):1067-73. doi: 10.1016/j.clinbiomech.2006.06.004. Epub 2006 Aug 17.
- Jia X, Suo S, Meng F, Wang R. Effects of alignment on interface pressure for transtibial amputee during walking. Disabil Rehabil Assist Technol. 2008 Nov;3(6):339-43. doi: 10.1080/17483100802044634.
- Xiaohong J, Xiaobing L, Peng D, Ming Z. The Influence of Dynamic Trans-tibial Prosthetic Alignment on Standing Plantar Foot Pressure. Conf Proc IEEE Eng Med Biol Soc. 2005;2005:6916-8. doi: 10.1109/IEMBS.2005.1616096.
- Lee VS, Solomonidis SE, Spence WD. Stump-socket interface pressure as an aid to socket design in prostheses for trans-femoral amputees--a preliminary study. Proc Inst Mech Eng H. 1997;211(2):167-80. doi: 10.1243/0954411971534287.
Datoer for undersøgelser
Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.
Studer store datoer
Studiestart (Faktiske)
6. september 2019
Primær færdiggørelse (Forventet)
10. april 2020
Studieafslutning (Forventet)
30. juni 2020
Datoer for studieregistrering
Først indsendt
25. marts 2020
Først indsendt, der opfyldte QC-kriterier
27. marts 2020
Først opslået (Faktiske)
30. marts 2020
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
30. marts 2020
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
27. marts 2020
Sidst verificeret
1. april 2019
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Andre undersøgelses-id-numre
- 09.2019.359
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
UBESLUTET
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Ingen
Studerer et amerikansk FDA-reguleret enhedsprodukt
Ingen
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Transtibial amputeret
-
US Department of Veterans AffairsUniversity of MichiganAfsluttetTranstibial amputationForenede Stater
-
Massachusetts Institute of TechnologyTilmelding efter invitationTranstibial amputation - UnilateralForenede Stater
-
Sheba Medical CenterAfsluttetUnilateral transtibial amputeret
-
University of WashingtonAfsluttet
-
University of OklahomaAfsluttet
-
Orthocare Innovations, LLCEunice Kennedy Shriver National Institute of Child Health and Human Development...Trukket tilbageTranstibial amputation | Kunstige lemmerForenede Stater
-
University of WashingtonAfsluttetTranstibial amputeretForenede Stater
-
VA Office of Research and DevelopmentAfsluttetTranstibial amputationForenede Stater
-
Southern California Institute for Research and...California State University, Long Beach; California State University, Dominguez...UkendtTranstibial amputationForenede Stater
-
Mayo ClinicAfsluttetTranstibial amputation - UnilateralForenede Stater
Kliniske forsøg med conventional manufactured socket
-
Owlet Baby Care, Inc.Afsluttet