Wearable Sweat Sensor for High-Concentration Lactate Monitoring in Critically Ill Patients
Development and Preliminary Clinical Validation of a Wearable Sweat Sensor for High-Concentration Lactate Monitoring in Critically Ill Patients: A Prospective Cohort Study
Studienübersicht
Status
Status
Bedingungen
Bedingungen
Intervention / Behandlung
Intervention / Behandlung
Detaillierte Beschreibung
Over the past decade, the wearable electronics industry has seen exponential growth in demand for continuous health monitoring technologies. From smartwatches to electronic skin, wearable devices have expanded their focus beyond biophysical signals to include biochemical signals such as small-molecule substances and macromolecular proteins in sweat. However, wearable sweat sensing technology still faces significant challenges, including discomfort and skin irritation caused by sweat-inducing materials, effective sweat extraction and collection capability, continuity and reliability of sweat sensing data, sensor calibration, and understanding of sweat analysis methods to discover clinically valuable new biomarkers. Therefore, how to translate wearable sweat monitoring devices into stable, continuous clinical use-matching the needs of health monitoring, disease diagnosis, and precision medicine-remains a research hotspot and challenge.
Lactate is abundant in sweat and is a hallmark product of anaerobic metabolism. Abnormal lactate levels are associated with multiple diseases. As an important marker of tissue perfusion, blood lactate level is closely associated with the prognosis of patients with severe sepsis and septic shock. Clinical guidelines explicitly recommend blood lactate monitoring in septic patients and lactate-guided fluid resuscitation when lactate levels are elevated. Dynamic monitoring of serum lactate changes and calculation of lactate clearance can more accurately reflect dynamic changes in tissue oxygen supply and consumption, and evaluate tissue perfusion and oxygen metabolism. Studies have shown that a lactate clearance rate ≥10% within 6 hours of resuscitation may predict lower mortality in septic patients. Clinical research has also shown that serum lactate concentration can be used to guide clinical treatment decisions, serving as both a therapeutic target and a prognostic indicator. However, most existing wearable sweat sensors have a detection limit of only 0-20 mmol/L, suitable only for exercise monitoring under normal physiological conditions, and cannot meet the clinical need for monitoring hyperlactatemia patients, whose lactate levels can reach up to 100 mmol/L. In addition, current wearable sweat lactate sensors suffer from poor reproducibility and are easily affected by environmental factors and complex sweat composition, posing major obstacles to clinical translation.
Based on this, the investigators have designed an intelligent wearable device for high-concentration sweat lactate monitoring, which includes a lactate detection electrode, a pH detection electrode, a temperature detection electrode, a sweat-stimulating electrode, a microfluidic sweat collection device, a hydrogel patch for inducing sweating, a flexible integrated circuit, and an intelligent display system. This wearable sweat lactate sensor achieves a detection range of up to 100 mmol/L, with a sensitivity of 89 nA/(cm²·mM) and a response time of 150 seconds, and achieves non-exercise-induced sweating through electrical stimulation. In addition, given the sensitivity of the lactate oxidase sensor to temperature and pH, corresponding temperature and pH sensors have been designed for calibration purposes.
This study is a prospective cohort study to be conducted in the surgical intensive care unit (ICU). Critically ill patients with blood lactate concentration >1.5 mmol/L will be enrolled and divided into two groups: (1) a sweat collection group, in which sweat will be collected using a commercial sweat collection device (HELA) for laboratory (ELISA) analysis; and (2) a wearable device monitoring group, in which patients will wear the investigator-designed wireless wearable sweat sensor for real-time monitoring. Clinical data, including medical history, imaging, and laboratory results, will be collected concurrently. The correlation between sweat and blood lactate concentrations, as well as the accuracy and safety of the wearable device, will be evaluated using Pearson correlation analysis.
Studientyp
Studientyp
Einschreibung (Geschätzt)
Einschreibung
Kontakte und Standorte
Studienkontakt
Studienkontakt
- Name: Kai Qu
- Telefonnummer: +8613609117104
- E-Mail: qukai001@xjtu.edu.cn
Teilnahmekriterien
Zulassungskriterien
Zulassungskriterien
Studienberechtigtes Alter
- Kind
- Erwachsene
- Älterer Erwachsener
Akzeptiert gesunde Freiwillige
Probenahmeverfahren
Studienpopulation
Beschreibung
Inclusion Criteria:
- Admitted to the surgical intensive care unit (ICU)
- Blood lactate concentration > 1.5 mmol/L
- Age ≥ 16 years
- Willing to voluntarily donate samples and provide signed informed consent
Exclusion Criteria:
- Pregnant women
- Pre-existing hematologic disease prior to infection
- Long-term use of immunosuppressive agents or presence of immunodeficiency
- Skin too dry to allow adequate sweat collection
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
Anzahl der Gruppen / Kohorten
Kohorten und Interventionen
Gruppe / KohorteGruppe / Kohorte |
Intervention / BehandlungIntervention / Behandlung |
|---|---|
|
Wearable Device Monitoring Group
|
A flexible wireless wearable device integrating a lactate detection electrode, pH detection electrode, temperature detection electrode, sweat-stimulating electrode, microfluidic sweat collection module, sweat-inducing hydrogel patch, and flexible integrated circuit with an intelligent display system.
The device induces localized non-exercise sweating via electrical stimulation and enables real-time detection of sweat lactate concentration up to 100 mmol/L.
|
|
Sweat Collection Group
|
A commercially available sweat collection device (HELA) used to collect sweat samples from critically ill patients at the time of routine blood draws for clinical lactate testing.
Collected sweat samples are subsequently analyzed for lactate concentration using ELISA in the laboratory.
|
Was misst die Studie?
Primäre Ergebnismessungen
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Sweat Lactate Concentration
Zeitfenster: At the time of enrollment, concurrent with clinical blood lactate testing (within the ICU stay, up to approximately 12 months of enrollment period)
|
Sweat lactate concentration will be measured in critically ill patients with blood lactate concentration >1.5 mmol/L, either via laboratory (ELISA) analysis of sweat samples collected using a commercial sweat collection device, or via real-time detection using the investigator-designed wearable sweat sensor.
The correlation between sweat and blood lactate concentrations will be analyzed using Pearson correlation to establish the sweat lactate concentration range corresponding to hyperlactatemia.
|
At the time of enrollment, concurrent with clinical blood lactate testing (within the ICU stay, up to approximately 12 months of enrollment period)
|
Sekundäre Ergebnismessungen
Sekundäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Correlation Between Sweat Lactate Concentration Measured by the Wearable Device and Blood Lactate Concentration Measured by Standard Clinical Testing
Zeitfenster: Throughout the device-monitoring period during ICU stay (up to 12 months)
|
Sweat lactate concentration (mmol/L) obtained from the wearable device will be compared with corresponding blood lactate concentration (mmol/L) obtained through standard clinical laboratory testing, using correlation analysis (e.g., Pearson or Spearman correlation coefficient) and Bland-Altman agreement analysis.
|
Throughout the device-monitoring period during ICU stay (up to 12 months)
|
|
Number of Participants With Device-Related Skin Adverse Events
Zeitfenster: Throughout the device-monitoring period during ICU stay (up to 12 months)
|
Number of participants experiencing skin irritation, discomfort, or other adverse events at the sensor application site, assessed through clinical observation during device use.
|
Throughout the device-monitoring period during ICU stay (up to 12 months)
|
Mitarbeiter und Ermittler
Sponsor
Sponsor
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn (Geschätzt)
Studienbeginn
Primärer Abschluss (Geschätzt)
Primärer Abschluss
Studienabschluss (Geschätzt)
Studienabschluss
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Tatsächlich)
Zuerst gepostet
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Tatsächlich)
Letztes Update gepostet
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Zusätzliche relevante MeSH-Bedingungen
Andere Studien-ID-Nummern
Andere Studien-ID-Nummern
- XJTUSAH-WB-001
Arzneimittel- und Geräteinformationen, Studienunterlagen
Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt
Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .