Accuracy Validation of a Pulse Oximetry Monitor
14 de octubre de 2016 actualizado por: Medtronic - MITG
SpO2 Accuracy Validation of a USB Pulse Oximetry Monitor Interface Cable Via Reference Co-Oximetry
The purpose of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry.
Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.
Descripción general del estudio
Estado
Retirado
Condiciones
Descripción detallada
The description of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry.
Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.
Tipo de estudio
De observación
Contactos y Ubicaciones
Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.
Ubicaciones de estudio
-
-
Colorado
-
Louisville, Colorado, Estados Unidos, 80027
- Clinimark
-
-
Criterios de participación
Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.
Criterio de elegibilidad
Edades elegibles para estudiar
18 años a 45 años (Adulto)
Acepta Voluntarios Saludables
Sí
Géneros elegibles para el estudio
Todos
Método de muestreo
Muestra no probabilística
Población de estudio
The study population will be ten to fifteen male and female subjects of any race, ranging in pigmentation from light to dark to meet the study design requirements
Descripción
Inclusion Criteria:
- healthy
- non-smoking, or who have refrained from smoking for 2 days
- male or female of any race
- 18-45 years.
- Subjects must have the ability to understand and provide written informed consent as well as be willing to comply with study procedures.
- To be eligible to participate in the study, the subject must clear a recent physical (12-lead ECG, medical history and blood test for sickle cell disease.) prior to enrollment
Exclusion Criteria:
- - morbid obesity (defined as BMI > 39.5),
- injuries, deformities or abnormalities or piercings that may prevent proper application of the device under test, compromised circulation,
- subjects with known respiratory conditions (smokers with COHb levels >3%, flu, pneumonia, bronchitis, shortness of breath/ respiratory distress, severe asthma, emphysema, COPD),
- subjects with known heart or cardiovascular conditions (uncontrolled hypertension, heart arrhythmias other than sinus arrhythmia, previous heart attack, blocked artery, history of myocardial ischemia, angina, myocardial infarction, congestive heart failure or cardiomyopathy, history of stroke, transient ischemic attack or carotid artery disease),
- other known health condition (diabetes, thyroid disease, kidney disease / chronic renal impairment, history of seizures (except childhood febrile seizures) or epilepsy, or unexplained syncope, recent history of frequent migraine headaches, recent head injury, cancer / chemotherapy
- women actively trying to get pregnant or are pregnant,
- clotting disorders (history of bleeding disorders or current use of blood thinners, Hemophilia, blood clots)
- Sickle cell disease
- Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
- Unwillingness or inability to remove colored nail polish or artificial nails from test digit(s)
- Prior or known allergies to: lidocaine (or similar pharmacological agents, e.g., Novocain) or heparin
- Recent arterial cannulation (i.e., less than 30 days prior to study date), Six or more arterial cannulations of each (right & left) radial artery, history of clinically significant complications from previous arterial cannulation
Plan de estudios
Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.
¿Cómo está diseñado el estudio?
Detalles de diseño
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Periodo de tiempo |
---|---|
The OxiCable pulse rate measurement performance is an ARMS of 3 or better over the full range 60-100% SaO2.
Periodo de tiempo: 2.5 hours
|
2.5 hours
|
Colaboradores e Investigadores
Aquí es donde encontrará personas y organizaciones involucradas en este estudio.
Patrocinador
Colaboradores
Investigadores
- Investigador principal: David Ransom, MD, Clinimark, LLC
Fechas de registro del estudio
Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.
Fechas importantes del estudio
Inicio del estudio
1 de noviembre de 2016
Finalización primaria (Anticipado)
1 de enero de 2017
Finalización del estudio (Anticipado)
1 de enero de 2017
Fechas de registro del estudio
Enviado por primera vez
14 de julio de 2015
Primero enviado que cumplió con los criterios de control de calidad
21 de julio de 2015
Publicado por primera vez (Estimar)
22 de julio de 2015
Actualizaciones de registros de estudio
Última actualización publicada (Estimar)
18 de octubre de 2016
Última actualización enviada que cumplió con los criterios de control de calidad
14 de octubre de 2016
Última verificación
1 de octubre de 2016
Más información
Términos relacionados con este estudio
Palabras clave
Otros números de identificación del estudio
- COVMOPR0499
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
INDECISO
Descripción del plan IPD
This study was cancelled.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .