Diese Seite wurde automatisch übersetzt und die Genauigkeit der Übersetzung wird nicht garantiert. Bitte wende dich an die englische Version für einen Quelltext.

Accuracy Validation of a Pulse Oximetry Monitor

14. Oktober 2016 aktualisiert von: Medtronic - MITG

SpO2 Accuracy Validation of a USB Pulse Oximetry Monitor Interface Cable Via Reference Co-Oximetry

The purpose of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.

Studienübersicht

Status

Zurückgezogen

Bedingungen

Detaillierte Beschreibung

The description of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.

Studientyp

Beobachtungs

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienorte

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

18 Jahre bis 45 Jahre (Erwachsene)

Akzeptiert gesunde Freiwillige

Ja

Studienberechtigte Geschlechter

Alle

Probenahmeverfahren

Nicht-Wahrscheinlichkeitsprobe

Studienpopulation

The study population will be ten to fifteen male and female subjects of any race, ranging in pigmentation from light to dark to meet the study design requirements

Beschreibung

Inclusion Criteria:

  • healthy
  • non-smoking, or who have refrained from smoking for 2 days
  • male or female of any race
  • 18-45 years.
  • Subjects must have the ability to understand and provide written informed consent as well as be willing to comply with study procedures.
  • To be eligible to participate in the study, the subject must clear a recent physical (12-lead ECG, medical history and blood test for sickle cell disease.) prior to enrollment

Exclusion Criteria:

  • - morbid obesity (defined as BMI > 39.5),
  • injuries, deformities or abnormalities or piercings that may prevent proper application of the device under test, compromised circulation,
  • subjects with known respiratory conditions (smokers with COHb levels >3%, flu, pneumonia, bronchitis, shortness of breath/ respiratory distress, severe asthma, emphysema, COPD),
  • subjects with known heart or cardiovascular conditions (uncontrolled hypertension, heart arrhythmias other than sinus arrhythmia, previous heart attack, blocked artery, history of myocardial ischemia, angina, myocardial infarction, congestive heart failure or cardiomyopathy, history of stroke, transient ischemic attack or carotid artery disease),
  • other known health condition (diabetes, thyroid disease, kidney disease / chronic renal impairment, history of seizures (except childhood febrile seizures) or epilepsy, or unexplained syncope, recent history of frequent migraine headaches, recent head injury, cancer / chemotherapy
  • women actively trying to get pregnant or are pregnant,
  • clotting disorders (history of bleeding disorders or current use of blood thinners, Hemophilia, blood clots)
  • Sickle cell disease
  • Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
  • Unwillingness or inability to remove colored nail polish or artificial nails from test digit(s)
  • Prior or known allergies to: lidocaine (or similar pharmacological agents, e.g., Novocain) or heparin
  • Recent arterial cannulation (i.e., less than 30 days prior to study date), Six or more arterial cannulations of each (right & left) radial artery, history of clinically significant complications from previous arterial cannulation

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Zeitfenster
The OxiCable pulse rate measurement performance is an ARMS of 3 or better over the full range 60-100% SaO2.
Zeitfenster: 2.5 hours
2.5 hours

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Mitarbeiter

Ermittler

  • Hauptermittler: David Ransom, MD, Clinimark, LLC

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn

1. November 2016

Primärer Abschluss (Voraussichtlich)

1. Januar 2017

Studienabschluss (Voraussichtlich)

1. Januar 2017

Studienanmeldedaten

Zuerst eingereicht

14. Juli 2015

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

21. Juli 2015

Zuerst gepostet (Schätzen)

22. Juli 2015

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Schätzen)

18. Oktober 2016

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

14. Oktober 2016

Zuletzt verifiziert

1. Oktober 2016

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Schlüsselwörter

Andere Studien-ID-Nummern

  • COVMOPR0499

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

UNENTSCHIEDEN

Beschreibung des IPD-Plans

This study was cancelled.

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .

3
Abonnieren