Tämä sivu käännettiin automaattisesti, eikä käännösten tarkkuutta voida taata. Katso englanninkielinen versio lähdetekstiä varten.

Evaluation of the Temperature Difference of Medical Gases Between the Heated Inspiratory Limb and the Tracheal Carina in Patients Undergoing Invasive Mechanical Ventilation With Active Humidification: A Crossover Clinical Study.

perjantai 13. marraskuuta 2026 päivittänyt: Riccardo Re, Azienda Sanitaria Universitaria Integrata del Trentino

The aim of this study is to assess the actual temperature drop of medical gases from the distal end of the heated ventilator circuit to the tracheal carina during invasive mechanical ventilation with active humidification. This analysis reflects the investigators' intention to observe the temperature drop under real-world clinical practice conditions, while assessing potential interactions with patient body temperature, ambient temperature, and the size of unheated ventilatory connectors. These findings may provide insight into the appropriate temperature and humidity targets, thereby further optimizing active humidification during invasive mechanical ventilation.

Tutkimuksen yleiskatsaus

Tila

Rekrytointi

Ehdot

Interventio / Hoito

Yksityiskohtainen kuvaus

Study setting: Patients undergoing invasive mechanical ventilation via orotracheal tube, nasotracheal tube, or tracheostomy cannula.

Research Hypothesis: The temperature drop between the distal end of the heated inspiratory circuit and the distal end of the tracheal tube does not exceed 2°C; therefore, heating medical gases above 37°C is not necessary.

Study Objective: 1) To assess the temperature difference between the distal end of the heated inspiratory limb and the patient's tracheal carina. 2) To identify variables that may influence the temperature drop, including connector size (dead space), the presence of a mainstream end-tidal CO₂ sensor, type, dimension and depth of tracheal tube (orotracheal, nasotracheal, tracheostomy), ventilation mode and minute volume, patient body temperature, and ambient temperature.

Study Design: Crossover clincal study Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation.

Protocol: The study protocol involves measuring, in all patients meeting the inclusion criteria, the temperature at the end of the heated ventilator circuit ("Tpatient") and the air temperature at the level of the tracheal carina. Tpatient will be recorded as the direct reading from the humidifier display. Tracheal temperature will be measured by inserting a sterile temperature probe through the ventilator circuit. To obtain an accurate measurement of air temperature without the probe contacting the tracheal mucosa, a graduated probe will be used, calibrated according to the correct insertion depth corresponding to the tracheal tube in use. The temperature recorded by the probe will then be displayed via a dedicated electrical cable on the bedside multiparameter monitor. The heat humidifier used in the study will be the device currently available in ICU 1 of Santa Chiara Hospital in Trento, namely the Dräger Aquapor H300® humidifier. The device will be set in FREE mode, allowing the investigators to freely adjust temperature settings according to the study protocol. For each enrolled patient, two sets of measurements will be performed using different humidification settings. The first set will use a Tpatient of 37°C with a -2°C gradient (Tpatient 37°C - chamber temperature 35°C), while the second set will use a Tpatient of 40°C with a -3°C gradient (Tpatient 40°C - chamber temperature 37°C). The order in which the first or second measurement set is performed will be randomly assigned using a web-based randomization tool (random.org). Before temperature measurements are performed, correct attainment of the target humidifier temperature will be verified. For each observation, the following variables will also be recorded: age and male patient core body temperature, type of airway device, tube size, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, ventilation mode, minute ventilation at the time of measurement, and ambient temperature. At the end of the measurements, the probe will be removed and placed on the patient's pillow to measure the ambient temperature in the immediate vicinity of the patient.

Timeframe: The study aims to enroll 50 patients within a maximum period of one year from the first measurement.

Statistical Analysis: During the statistical analysis, the mean, standard deviation, median, and range of the temperature differences between Tpatient and the air temperature at the tracheal carina will be calculated for the two humidification settings (37°C and 40°C). These values will then be compared using the Bland-Altman method. In addition, potential correlations between temperature variations and the study's objective variables will be analysed, including core body temperature, type and size of the airway device, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, minute ventilation at the time of measurement, and ambient temperature.

Expected Results: The temperature drop induced by unheated ventilator connectors (catheter mount and tracheal tube/cannula) does not exceed 2°C and is influenced by both ambient temperature and patient body temperature. Accordingly, a heated inspiratory circuit temperature setting above 37°C is not necessary.

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

50

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

Opiskelupaikat

    • Italy
      • Trento, Italy, Italia, 38100
        • Rekrytointi
        • Santa Chiara Hospital of Trento - ASUIT
        • Ottaa yhteyttä:

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

18 vuotta ja vanhemmat (Aikuinen, Vanhempi Aikuinen)

Hyväksyy terveitä vapaaehtoisia

Ei

Sukupuolet, jotka voivat opiskella

Kaikki

Kuvaus

Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Hoito
  • Jako: Satunnaistettu
  • Inventiomalli: Crossover-tehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: Temperature drop 37°C

Measure of the temperarure drop in two different humidification setting: 37°C - 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

Kokeellinen: Temperature drop 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Temperature drop
Aikaikkuna: 50 patients or 1 year

Measure the real temperature drop from the end of the heated respiratory tube and the tracheal carina in the patient with mechanical invasive ventilation and active humidification.

50 patients or 1 year

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Sponsori

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Maanantai 22. kesäkuuta 2026

Ensisijainen valmistuminen (Arvioitu)

Tiistai 22. kesäkuuta 2027

Opintojen valmistuminen (Arvioitu)

Tiistai 22. kesäkuuta 2027

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Perjantai 13. marraskuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Perjantai 13. marraskuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Keskiviikko 18. marraskuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Keskiviikko 18. marraskuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Perjantai 13. marraskuuta 2026

Viimeksi vahvistettu

Maanantai 1. kesäkuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muut tutkimustunnusnumerot

  • CARINA TEMP

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .