Tämä sivu käännettiin automaattisesti, eikä käännösten tarkkuutta voida taata. Katso englanninkielinen versio lähdetekstiä varten.

Accuracy and Safety of Anytime 5Pro and Anytime 4Pro CGM Systems in Adults With Diabetes

maanantai 8. kesäkuuta 2026 päivittänyt: Yuwell Group

Accuracy and Safety Evaluation of the Anytime 5Pro and Anytime 4Pro Continuous Glucose Monitoring Systems in Adults With Diabetes: A Post-Market Clinical Follow-up Investigation

This post-market clinical follow-up (PMCF) investigation evaluates the accuracy and safety of two CE-marked continuous glucose monitoring (CGM) systems, the Anytime 5Pro and the Anytime 4Pro, in adults with type 1 or type 2 diabetes. Each participant simultaneously wears one Anytime 5Pro sensor and one Anytime 4Pro sensor, one on each upper arm, for the full labelled wear period of each device (up to 16 and 15 days, respectively). CGM readings are compared against venous blood glucose measured by a laboratory reference analyzer (YSI 2500) during four in-clinic sampling sessions that include controlled glucose manipulation. Participants are masked to real-time CGM readings throughout the study. The investigation is conducted to confirm device performance and safety under real-world conditions of use in a European population, in line with the manufacturer's post-market clinical follow-up obligations under the EU Medical Device Regulation.

Tutkimuksen yleiskatsaus

Yksityiskohtainen kuvaus

This is a prospective, multi-centre, dual-device (paired intra-subject), objective performance criteria PMCF investigation. The investigation will be conducted at two clinical site(s) in Poland. Each participant will simultaneously wear two CGM sensors: one Anytime 5Pro sensor and one Anytime 4Pro sensor, each inserted on the posterior-lateral side of one upper arm (one sensor per arm), each for the duration of its respective labelled wear life. The Anytime 4Pro sensor will be worn from insertion on Day 1 until its removal at Visit 5 on Day 16, following completion of the Session 4 reference sampling and full data synchronisation (15-day labelled wear life). The Anytime 5Pro sensor will be worn from insertion on Day 1 until its removal at Visit 6 on Day 17 (16-day labelled wear life). During the study course participants will not be able to read the real-time CGM glucose readings; the mobile application display will be masked throughout the sensor wear period. Each participant will be provided with sponsor-supplied smartphone(s) with the Yuwell Anytime companion application installed. The smartphone will be carried by the participant throughout the sensor wear period to maintain continuous Bluetooth data transfer from the transmitter. The application display will be masked so that participants cannot view their glucose values, trends, or alerts.

Sensor placement side will be allocated deterministically at Visit 1 based on the participant identifier parity assigned at enrolment. Even-numbered participant identifier (Allocation A): Anytime 5Pro on the right upper arm, Anytime 4Pro on the left upper arm. Odd-numbered participant identifier (Allocation B): Anytime 5Pro on the left upper arm, Anytime 4Pro on the right upper arm. Because participant identifiers are issued sequentially, this rule yields a balanced 1:1 distribution of placement-side allocations across participants and across participating centres, without the need for envelope-based randomisation. The rationale for this deterministic allocation is to control for potential systematic side-related bias (dominant arm use, sleeping posture, adhesion wear) when comparing the two devices within-subject.

A total of 140 sensors will be used in the investigation: 70 Anytime 5Pro sensors and 70 Anytime 4Pro sensors (one of each per participant).

Each participant will attend a total of six clinic visits. Visit 1 (Day 1) is the screening, enrolment and dual sensor insertion visit, during which eligibility assessments will be performed, informed consent obtained, and one Anytime 5Pro sensor and one Anytime 4Pro sensor inserted on the contralateral upper arms in accordance with the deterministic side allocation rule. The Anytime 5Pro sensor will undergo a 45-minute warm-up and the Anytime 4Pro sensor a 60-minute warm-up. The participant will also receive the sponsor-supplied smartphone(s) with the Yuwell Anytime application installed in masked mode. The four subsequent in-clinic venous blood sampling sessions will be scheduled at different phases of the sensor wear period: Session 1 (Visit 2, Day 2) - Beginning; Session 2 (Visit 3, Day 6 or Day 7) - Early middle; Session 3 (Visit 4, Day 10 or Day 11) - Late middle; Session 4 (Visit 5, Day 16, to be started before 09:00) - End. Each session will last 9 hours, with venous blood sampled every 15+/-3 minutes to obtain paired CGM-reference measurements for accuracy assessment. As a pre-defined exception, for 2 participants (1 per site, scheduled in the mid-to-late stage of the investigation), Visit 1 and Visit 2 will be combined and their Session 1 will last 12 hours from sensor insertion, to capture the first 12 hours of wear (during which accuracy may vary per the IFU); participation in the combined visit is an optional element the participant marks in the main informed consent form. At Visit 5 (Day 16), following completion of Session 4 reference sampling and full Anytime 4Pro data synchronisation, the Anytime 4Pro sensor will be removed at the end of its 15-day labelled wear life and its insertion site assessed. Visit 6 (Day 17, after 19:00) will be the Anytime 5Pro sensor removal visit, during which the Anytime 5Pro sensor will be removed, post-removal skin assessment and the usability questionnaires will be performed, and adverse events will be reviewed. Between clinic visits, participants will wear both sensors in the home-use setting under their usual daily conditions. Because the application display is masked, participants will not receive glucose values, trends or alerts from the investigational devices; glucose management decisions during the study will therefore be based on standard-of-care methods (fingerstick blood glucose monitoring or the participant's usual CGM system, if applicable) and not on the investigational device readings.

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

70

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

  • Aikuinen
  • Vanhempi Aikuinen

Hyväksyy terveitä vapaaehtoisia

Ei

Kuvaus

Inclusion Criteria:

  1. Diagnosis of type 1 or type 2 diabetes mellitus, documented in medical records, with diabetes established at least 6 months prior to Day 1.
  2. Age >= 18 years at the time of informed consent.
  3. Stable vital signs at screening (systolic blood pressure 90-180 mmHg, diastolic blood pressure 50-110 mmHg, heart rate 50-100 bpm, body temperature 35.5-37.5 C).
  4. Feasibility of establishing forearm venous access for repeated blood sampling.
  5. Haematocrit <= 70%.
  6. For T1D participants: prior experience with a CGM system (any brand).
  7. Willingness and ability to comply with the investigation procedures, including wearing both sensors (one Anytime 5Pro and one Anytime 4Pro) simultaneously throughout the 17-day participation period, attending six scheduled clinic visits, and participating in glucose manipulation procedures.
  8. Written informed consent obtained prior to any investigation-related procedures.

Exclusion Criteria:

  1. Skin disorders at the prospective sensor insertion site (posterior-lateral upper arm), including scarring, redness, swelling, infection, or tattoos.
  2. Known allergies to alcohol-based disinfectants or medical adhesives.
  3. Coagulation abnormalities or bleeding disorders, or current use of anticoagulant therapy (except low-dose aspirin).
  4. Current pregnancy, suspected pregnancy, or lactation.
  5. History of epilepsy or psychiatric disorders that may compromise investigation compliance or subject safety during glucose manipulation procedures.
  6. Severe hepatic or renal insufficiency.
  7. Current use of medications known to interfere with CGM readings, including ascorbic acid (vitamin C) supplementation >1000 mg/day.
  8. Concurrent participation in another interventional device or drug trial, or participation within the past 30 days.
  9. Known drug or alcohol abuse.
  10. Any condition that, in the judgement of the investigator, would make the subject unsuitable for participation in this investigation or would compromise the subject's safety.
  11. Planned magnetic resonance imaging (MRI), computed tomography (CT), X-ray examination, or diathermic therapy during the sensor wear period.

    -

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Diagnostiikka
  • Jako: Ei käytössä
  • Inventiomalli: Yksittäinen ryhmätehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: Dual CGM wear (Anytime 5Pro + Anytime 4Pro)
During the four in-clinic reference sampling sessions, participants undergo controlled manipulation of blood glucose to obtain paired CGM-reference measurements spanning a wide glucose range, including hypoglycaemic and hyperglycaemic levels. Glucose is raised and/or lowered under continuous medical supervision using standard clinical means (e.g., carbohydrate intake and/or insulin administration) according to the protocol, with predefined safety limits and stopping rules. Each session lasts up to 9 hours, with venous blood sampled every 15 (+/-3) minutes via an indwelling catheter for reference glucose measurement on the YSI 2500 analyzer. The purpose is to ensure adequate representation of low, mid-range, and high glucose values for the accuracy endpoints (MARD, agreement rates, and Consensus Error Grid analysis), rather than to treat the participant's diabetes.

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Mean Absolute Relative Difference (MARD) between CGM and reference glucose
Aikaikkuna: Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)
MARD across the reportable glucose range (30-500 mg/dL), reported separately for each device. Calculated as the mean of the absolute values of (CGM value - reference value) / reference value, expressed as a percentage, across all evaluable paired CGM-reference points. Reported for the Anytime 5Pro and the Anytime 4Pro independently.
Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)
Agreement rate within +/-20%/+/-20 mg/dL of reference glucose
Aikaikkuna: Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)
Proportion of paired CGM-reference points within +/-20 mg/dL of the reference for reference values <=80 mg/dL, or within +/-20% of the reference for reference values >80 mg/dL, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)
Proportion of paired points in Consensus Error Grid zones A and B
Aikaikkuna: Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)
Percentage of paired CGM-reference points falling within zones A and B of the Consensus Error Grid analysis, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Paired measurements collected during four in-clinic sessions across the sensor wear period (Days 2 to 16)

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Sensor accuracy stability across wear phases (MARD and +/-20%/+/-20 mg/dL agreement)
Aikaikkuna: Days 1 to 16
MARD and the +/-20%/+/-20 mg/dL agreement rate calculated at each phase of the wear period (beginning, early middle, late middle, end), reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Point accuracy stratified by rate of change (RoC) of glucose
Aikaikkuna: Days 1 to 16
Point accuracy stratified by rate of change of glucose concentration, categorised as < -2, -2 to < -1, -1 to < 0, 0 to 1, > 1 to 2, and > 2 mg/dL per minute, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Sensor accuracy across glucose concentration ranges (iCGM special control approach)
Aikaikkuna: Days 1 to 16
Accuracy across glucose ranges (<54, 54-70, 70-180, >180 mg/dL), following the integrated CGM (iCGM) special control approach: proportion of CGM readings within the absolute (mg/dL) criterion for reference values <=70 mg/dL and within the relative (%) criterion for reference values >70 mg/dL, with corresponding 95% lower confidence limits. Reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Mean Absolute Difference (MAD) for glucose values <=70 mg/dL
Aikaikkuna: Days 1 to 16
Mean absolute difference (MAD) in mg/dL between CGM and reference for reference glucose values <=70 mg/dL, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Hypoglycaemic alert performance
Aikaikkuna: Days 1 to 16
True alert rate and detection rate at the hypoglycaemic threshold of 70 mg/dL (3.9 mmol/L), evaluated within a +/-15 minute window, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Hyperglycaemic alert performance
Aikaikkuna: Days 1 to 16
True alert rate and detection rate at the hyperglycaemic threshold of 200 mg/dL (11.1 mmol/L), evaluated within a +/-15 minute window, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16
Agreement rates at +/-15%/+/-15 mg/dL and +/-30%/+/-30 mg/dL
Aikaikkuna: Days 1 to 16
Proportion of paired CGM-reference points within +/-15%/+/-15 mg/dL and within +/-30%/+/-30 mg/dL of reference values, applying the same absolute/relative threshold methodology as the +/-20%/+/-20 mg/dL primary endpoint. Reported separately for the Anytime 5Pro and the Anytime 4Pro.
Days 1 to 16

Muut tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Incidence, severity, and relatedness of adverse events and serious adverse events
Aikaikkuna: Day 1 to Day 17
Incidence, severity, and relatedness of all adverse events (AEs) and serious adverse events (SAEs), including adverse device effects. Safety findings attributed to the relevant device (Anytime 5Pro, Anytime 4Pro, or ambiguous/both) and reported per device in addition to the overall pooled summary.
Day 1 to Day 17
Incidence of device deficiencies and device-related adverse events
Aikaikkuna: Day 1 to Day 17
Incidence of device deficiencies and device-related adverse events, reported separately for the Anytime 5Pro and the Anytime 4Pro.
Day 1 to Day 17
Skin assessment findings at sensor insertion sites
Aikaikkuna: Day 1 to Day 17
Skin assessment findings at each of the two sensor insertion sites (Anytime 5Pro site and Anytime 4Pro site, evaluated independently) at each clinic visit and upon sensor removal: erythema, oedema, induration, pruritus, bruising, and infection.
Day 1 to Day 17
Participant satisfaction and usability (QUSS-CGM-based questionnaires)
Aikaikkuna: Day 17 (Visit 6)
Participant satisfaction and usability assessed by two device-specific questionnaires (one per device) based on the QUSS-CGM and adapted for the masked-application context, covering ease of sensor insertion and removal, wearing comfort, daily-life impact, sensor size and adhesive performance, and overall user acceptance. Analysed and reported separately for the Anytime 5Pro and the Anytime 4Pro.
Day 17 (Visit 6)
Sensor functional survival (Kaplan-Meier)
Aikaikkuna: Through sensor wear period (up to 16 days for Anytime 5Pro; up to 15 days for Anytime 4Pro)
Sensor functional survival estimated using Kaplan-Meier methodology, calculated separately for the Anytime 5Pro (over the 16-day labelled wear period) and the Anytime 4Pro (over the 15-day labelled wear period).
Through sensor wear period (up to 16 days for Anytime 5Pro; up to 15 days for Anytime 4Pro)
Lag time between interstitial fluid glucose and venous blood glucose
Aikaikkuna: Days 1 to 16
Lag time between interstitial fluid glucose measured by each CGM system and venous blood glucose measured by the reference analyzer, evaluated separately for each device in accordance with the IFCC Working Group recommendations on the clinical performance evaluation of CGM systems.
Days 1 to 16

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Arvioitu)

Maanantai 20. heinäkuuta 2026

Ensisijainen valmistuminen (Arvioitu)

Sunnuntai 31. tammikuuta 2027

Opintojen valmistuminen (Arvioitu)

Sunnuntai 31. tammikuuta 2027

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Maanantai 8. kesäkuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Maanantai 8. kesäkuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Perjantai 12. kesäkuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Perjantai 12. kesäkuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Maanantai 8. kesäkuuta 2026

Viimeksi vahvistettu

Maanantai 1. kesäkuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

PÄÄTTÄMÄTÖN

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .

Kliiniset tutkimukset Diabetes Mellitus, Type 1 | Diabetes Mellitus, Type 2

Tilaa