Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)

2026年11月15日 更新者:University Hospital, Montpellier

SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia

SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.

Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.

The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.

This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.

調査の概要

状態

まだ募集していません

条件

介入・治療

研究の種類

介入

入学 (推定)

40

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

      • Montpellier、フランス
        • University hospital Montpellier
        • コンタクト:
        • 主任研究者:
          • Eva Tamburro, MD

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • Adult subject
  • Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
  • Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.

Exclusion Criteria:

  • Subject lacking the cognitive capacity to perform and understand the experimental protocol
  • Subject unable to read and/or write
  • Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
  • Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
  • Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
  • Pregnant or breastfeeding woman.
  • Absence of free and informed consent
  • Subject not affiliated with a social security scheme
  • Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
  • Subject deprived of liberty by judicial or administrative decision
  • Persons participating in another study with an exclusion period still ongoing.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:他の
  • 割り当て:なし
  • 介入モデル:単一グループの割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:Instrumented spasticity assessment

Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility.

Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
時間枠:Day 1 and Day 2

Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2.

Day 1 and Day 2

二次結果の測定

結果測定
メジャーの説明
時間枠
Difference in total elbow range of motion between the affected and unaffected upper limbs
時間枠:Day 1

Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean stretch velocity between the affected and unaffected upper limbs
時間枠:Day 1

Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the maximum resistance force between the affected and unaffected upper limbs
時間枠:Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean resistance force between the affected and unaffected upper limbs
時間枠:Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
時間枠:Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Comparison of Instrumental and Clinical Measures of total elbow range of motion
時間枠:Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental and Clinical Measures of stretch velocity
時間枠:Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
時間枠:Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
時間枠:Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
時間枠:Day 1

Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds.

Day 1
Feasibility of post hoc biomechanical parameter extraction from video recordings
時間枠:Day 1

Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed.

Day 1
Comparison of Video-Derived and Instrumental total range of motion
時間枠:Day 1

Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental mean stretch velocity
時間枠:Day 1

Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental maximum resistance force
時間枠:Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of mean resistance force
時間枠:Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
時間枠:Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Clinical Measures of total range of motion
時間枠:Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of stretch velocity
時間枠:Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
時間枠:Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Video-Derived and Clinical Resistance Force Measurements
時間枠:Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Eva Tamburro, MD、University Hospital, Montpellier

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2027年2月1日

一次修了 (推定)

2029年2月1日

研究の完了 (推定)

2029年2月1日

試験登録日

最初に提出

2026年8月25日

QC基準を満たした最初の提出物

2026年11月15日

最初の投稿 (実際)

2026年11月20日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月20日

QC基準を満たした最後の更新が送信されました

2026年11月15日

最終確認日

2026年11月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • RECHMPL25_0311
  • 2026-A01370-51 (その他の識別子:IDRCB)

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

未定

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。