Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)
SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia
SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.
Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.
The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.
This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.
Tutkimuksen yleiskatsaus
Tila
Tila
Ehdot
Ehdot
Interventio / Hoito
Interventio / Hoito
Opintotyyppi
Opintotyyppi
Ilmoittautuminen (Arvioitu)
Ilmoittautuminen
Vaihe
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskeluyhteys
Opiskeluyhteys
- Nimi: Eva Tamburro, MD
- Puhelinnumero: 33 06 95 16 61 97
- Sähköposti: e-tamburro@chu-montpellier.fr
Opiskelupaikat
-
-
-
Montpellier, Ranska
- University hospital Montpellier
-
Ottaa yhteyttä:
- Eva Tamburro, MD
- Puhelinnumero: 33 06 95 16 61 97
- Sähköposti: e-tamburro@chu-montpellier.fr
-
Päätutkija:
- Eva Tamburro, MD
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Kuvaus
Inclusion Criteria:
- Adult subject
- Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
- Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.
Exclusion Criteria:
- Subject lacking the cognitive capacity to perform and understand the experimental protocol
- Subject unable to read and/or write
- Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
- Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
- Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
- Pregnant or breastfeeding woman.
- Absence of free and informed consent
- Subject not affiliated with a social security scheme
- Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
- Subject deprived of liberty by judicial or administrative decision
- Persons participating in another study with an exclusion period still ongoing.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Muut
- Jako: Ei käytössä
- Inventiomalli: Yksittäinen ryhmätehtävä
- Naamiointi: Ei mitään (avoin tarra)
Aseiden lukumäärä
Aseet ja interventiot
Osallistujaryhmä / ArmOsallistujaryhmä / Arm |
Interventio / HoitoInterventio / Hoito |
|---|---|
|
Kokeellinen: Instrumented spasticity assessment
Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility. |
Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility. |
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
Aikaikkuna: Day 1 and Day 2
|
Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2. |
Day 1 and Day 2
|
Toissijaiset tulostoimenpiteet
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Difference in total elbow range of motion between the affected and unaffected upper limbs
Aikaikkuna: Day 1
|
Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the mean stretch velocity between the affected and unaffected upper limbs
Aikaikkuna: Day 1
|
Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the maximum resistance force between the affected and unaffected upper limbs
Aikaikkuna: Day 1
|
Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the mean resistance force between the affected and unaffected upper limbs
Aikaikkuna: Day 1
|
Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
Aikaikkuna: Day 1
|
Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time |
Day 1
|
|
Comparison of Instrumental and Clinical Measures of total elbow range of motion
Aikaikkuna: Day 1
|
Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Instrumental and Clinical Measures of stretch velocity
Aikaikkuna: Day 1
|
Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
Aikaikkuna: Day 1
|
Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
Aikaikkuna: Day 1
|
Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time. Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
Aikaikkuna: Day 1
|
Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds. |
Day 1
|
|
Feasibility of post hoc biomechanical parameter extraction from video recordings
Aikaikkuna: Day 1
|
Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed. |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of motion
Aikaikkuna: Day 1
|
Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental mean stretch velocity
Aikaikkuna: Day 1
|
Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental maximum resistance force
Aikaikkuna: Day 1
|
Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of mean resistance force
Aikaikkuna: Day 1
|
Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
Aikaikkuna: Day 1
|
Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast). |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of total range of motion
Aikaikkuna: Day 1
|
Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of stretch velocity
Aikaikkuna: Day 1
|
Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale |
Day 1
|
|
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
Aikaikkuna: Day 1
|
Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
|
Comparison of Video-Derived and Clinical Resistance Force Measurements
Aikaikkuna: Day 1
|
Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch. |
Day 1
|
Yhteistyökumppanit ja tutkijat
Sponsori
Sponsori
Tutkijat
Tutkijat
- Päätutkija: Eva Tamburro, MD, University Hospital, Montpellier
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Arvioitu)
Opiskelun aloitus
Ensisijainen valmistuminen (Arvioitu)
Ensisijainen valmistuminen
Opintojen valmistuminen (Arvioitu)
Opintojen valmistuminen
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Ensimmäinen Lähetetty
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin päivitys julkaistu
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
Muut tutkimustunnusnumerot
- RECHMPL25_0311
- 2026-A01370-51 (Muu tunniste: IDRCB)
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .