Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia (SENSAS)

15 de novembro de 2026 atualizado por: University Hospital, Montpellier

SENSAS - Evaluation of the Reliability of a Wearable Device Combining Inertial Sensors and a Dynamometer for Analyzing Elbow Flexor Spasticity in Patients With Hemiplegia

SSpasticity is a common motor disorder following central nervous system injury, characterized by a velocity-dependent increase in muscle tone. It affects diverse populations, including patients with stroke, spinal cord injury, multiple sclerosis, or cerebral palsy, and leads to significant functional, pain-related, and medico-economic consequences.

Clinical assessment currently relies on the Modified Tardieu Scale (MTS) and the Modified Ashworth Scale (MAS), whose main limitations are low inter- and intra-rater reproducibility and subjective quantification. Coupling inertial measurement units (IMU) with a portable dynamometer represents a promising instrumental approach to overcome these limitations.

The hypothesis of this pilot clinical investigation is that instrumental measurement combining a portable dynamometer and inertial sensors would achieve satisfactory inter- and intra-rater reproducibility for the assessment of elbow flexor spasticity in patients with unilateral spastic paresis, thereby offering a reliable and objective alternative to current clinical scales.

This project proposes an instrumental assessment of upper-limb spasticity simultaneously combining kinematic measurement (angle, angular velocity) via two K-Move inertial sensors and measurement of resistance force via the K-Push dynamometer (Kinvent, Montpellier), which can easily be used during consultations or at the patient's bedside. This clinical investigation will secondarily assess the feasibility of a kinematic analysis based exclusively on videos acquired by tablet during the instrumental assessment.

Visão geral do estudo

Status

Ainda não está recrutando

Condições

Intervenção / Tratamento

Tipo de estudo

Intervencional

Inscrição (Estimado)

40

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Locais de estudo

      • Montpellier, França
        • University hospital Montpellier
        • Contato:
        • Investigador principal:
          • Eva Tamburro, MD

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

18 anos e mais velhos (Adulto, Adulto mais velho)

Aceita Voluntários Saudáveis

Não

Gêneros Elegíveis para o Estudo

Tudo

Descrição

Inclusion Criteria:

  • Adult subject
  • Unilateral spastic paresis of the elbow flexors (MTS ≥ 1/4), all etiologies combined
  • Subject hospitalized in the Physical Medicine and Rehabilitation department of Montpellier University Hospital (CHU), undergoing a rehabilitation stay, whether as an inpatient, in a week-hospital, or in a day-hospital setting.

Exclusion Criteria:

  • Subject lacking the cognitive capacity to perform and understand the experimental protocol
  • Subject unable to read and/or write
  • Subject with burns, wounds, abscesses, or any other skin lesion in the area where the devices are to be used
  • Subject with a non-united or healing fracture of the arm, forearm, or wrist of either upper limb
  • Subject with severe osteoporosis or bone fragility considered incompatible with repeated passive mobilizations
  • Pregnant or breastfeeding woman.
  • Absence of free and informed consent
  • Subject not affiliated with a social security scheme
  • Subject under a legal protection measure (safeguard of justice, guardianship, curatorship)
  • Subject deprived of liberty by judicial or administrative decision
  • Persons participating in another study with an exclusion period still ongoing.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Outro
  • Alocação: N / D
  • Modelo Intervencional: Atribuição de grupo único
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: Instrumented spasticity assessment

Adults with unilateral spastic paresis undergoing standardized clinical and instrumental assessment of elbow flexor spasticity using inertial sensors (K-Move), a handheld dynamometer (K-Push), and the Modified Tardieu Scale. Measurements are repeated to evaluate inter-rater and intra-rater reproducibility.

Instrumented assessment of elbow flexor spasticity using two wearable inertial sensors (K-Move) synchronized with a handheld dynamometer (K-Push). The system simultaneously records joint angle, angular velocity, and resistance force during passive elbow movements. Measurements are compared with the Modified Tardieu Scale and repeated to assess inter-rater and intra-rater reproducibility.

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Inter-rater and intra-rater reproducibility of instrumental measurement for elbow flexor spasticity assessment
Prazo: Day 1 and Day 2

Reproducibility of biomechanical parameters obtained from the instrumental measurement combining two inertial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast). Parameters include total elbow extension range of motion (in degrees), mean stretch velocity (in meters per second, m/s), angle of onset of spasticity (in degrees), referred to as the "catch", maximum resistance force measured by the dynamometer (in Newtons), mean resistance force measured by the dynamometer (in Newtons) over the entire movement, and extent of spastic resistance defined as the range of motion over which the resistance persists (in degrees). Inter-rater reproducibility will be assessed between two independent raters on Day 1, and intra-rater reproducibility will be assessed by repeated measurements performed by the same rater between Day 1 and Day 2.

Day 1 and Day 2

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Difference in total elbow range of motion between the affected and unaffected upper limbs
Prazo: Day 1

Comparison of the total elbow extension range of motion (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean stretch velocity between the affected and unaffected upper limbs
Prazo: Day 1

Comparison of the mean stretch velocity (in meters per second, m/s) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the maximum resistance force between the affected and unaffected upper limbs
Prazo: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the mean resistance force between the affected and unaffected upper limbs
Prazo: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Difference in the extend of spastic resistance between the affected and unaffected upper limbs
Prazo: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb and the unaffected controlateral limb, measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast), measured in real time

Day 1
Comparison of Instrumental and Clinical Measures of total elbow range of motion
Prazo: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental and Clinical Measures of stretch velocity
Prazo: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Instrumental Resistance Force and Clinical Spasticity Assessment
Prazo: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Instrumental extend of spastic resistance and Clinical Spasticity Assessment
Prazo: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured in real time.

Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Duration of instrumental measurement compared with Modified Tardieu Scale assessment
Prazo: Day 1

Comparison of the administration time required to perform the instrumental assessment using inertial sensors and a dynamometer versus the clinical assessment using the Modified Tardieu Scale. Time will be recorded in minutes and seconds.

Day 1
Feasibility of post hoc biomechanical parameter extraction from video recordings
Prazo: Day 1

Feasibility of extracting biomechanical parameters from synchronized video recordings acquired during instrumental assessment using predefined image quality and joint landmark visibility criteria. Feasibility will be determined by the proportion of participants for whom video-based biomechanical analysis can be successfully performed.

Day 1
Comparison of Video-Derived and Instrumental total range of motion
Prazo: Day 1

Comparison of the total elbow extension range of motion (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental mean stretch velocity
Prazo: Day 1

Comparison of mean stretch velocity (in meters per second, m/s) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental maximum resistance force
Prazo: Day 1

Comparison of the maximum resistance force measured by the dynamometer (in Newtons) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of mean resistance force
Prazo: Day 1

Comparison of the mean resistance force measured by the dynamometer (in Newtons) over the entire movement obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Instrumental total range of extend of spastic resistance
Prazo: Day 1

Comparison of the extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) obtained through real-time instrumental measurement using two initial sensors (K-Move) combined with a portable dynamometer (K-Push), and retrospective measurement through video analysis, during passive elbow extension of the affected limb at three stretch velocities (slow, intermediate, and fast).

Day 1
Comparison of Video-Derived and Clinical Measures of total range of motion
Prazo: Day 1

Total elbow extension range of motion (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Measured angle (in degrees) measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of stretch velocity
Prazo: Day 1

Mean stretch velocity (in meters per second, m/s) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Stretch velocity categorized as slow, medium or fast, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale

Day 1
Comparison of Video-Derived and Clinical Measures of extend of spastic resistance
Prazo: Day 1

Extend of spastic resistance defined as the range of motion over which the resistance persists (in degrees) of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1
Comparison of Video-Derived and Clinical Resistance Force Measurements
Prazo: Day 1

Mean resistance force measured by the dynamometer (in Newtons) over the entire movement of the affected limb measured with the instrumental measurement combining two initial sensors (K-Move) and a portable dynamometer (K-Push) during passive elbow extension at three stretch velocities (slow, intermediate, and fast) measured retrospectively through video analysis Modified Tardieu Scale Assessment, from 0 to 4, measured on the affected limb during the clinical assessment of spasticity using the Modified Tardieu Scale The Modified Tardieu Scale grade ranges from 0 to 4, where 0 indicates no resistance during passive movement and 4 indicates sustained clonus lasting more than 10 seconds. Higher scores indicate a more pronounced muscle reaction to fast passive stretch.

Day 1

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Investigadores

  • Investigador principal: Eva Tamburro, MD, University Hospital, Montpellier

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

1 de fevereiro de 2027

Conclusão Primária (Estimado)

1 de fevereiro de 2029

Conclusão do estudo (Estimado)

1 de fevereiro de 2029

Datas de inscrição no estudo

Enviado pela primeira vez

25 de agosto de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

15 de novembro de 2026

Primeira postagem (Real)

20 de novembro de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

20 de novembro de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

15 de novembro de 2026

Última verificação

1 de novembro de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • RECHMPL25_0311
  • 2026-A01370-51 (Outro identificador: IDRCB)

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

INDECISO

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .